SYSMEX CyStain VitalCount Manual page 15

Table of Contents

Advertisement

Fault
Unusal large CV values
and/or
failling
Alignment test within the
Quality Check report
Count Check Beads green
do
not
appear
predefined regions of the
configuration script and /or
failing Gate Position test
within the Quality Check
report
Count Check Beads green
measurement is not within
± 20 % range of the LOT-
specific concentration stated
on the bottle and /or failing
Counting test within the
Quality Check report
"LOT
mismatch"
displayed in the "Device
Status" after entering the QC
Material LOT number
CyStain™ VitalCount QC Procedure | QC Manual | June 2021
Remedy
Perform a priming procedure.
Optical
Perform a cleaning procedure.
Check the date of expiry of the QC material.
Shake the bottle of the QC material vigorously (e.g. by
vortexing) and repeat the measurement.
Check and adjust gain values for all parameters.
Check within the software if you have loaded the correct
in
the
configuration file for your measurement.
Shake the bottle vigorously (e.g. by vortexing) and repeat
the measurement.
Check the date of expiry of the QC material.
Perform a priming procedure.
Perform a cleaning procedure.
Adjust gain values for all parameters
Shake the bottle vigorously (e.g. by vortexing) and repeat
the measurement.
Check the date of expiry of the Count Check Beads green.
Perform a priming procedure.
Perform a cleaning procedure.
Refill the Sheath Fluid (ideally daily).
Renew the Inline Filter. The Inline Filter should be
replaced every 3 months.
Check and adjust Gain values for all parameters.
is
Make sure to use a matching FCS Express™ template
and Count Check Beads green LOT.
The LOT number of the Count Check Beads green is
composed of "NN00YYMMDD". They are compatible with
script / template version identical to the first 4 digits of the
LOT number "NN00".
5 Troubleshooting
15