3 Overview
3.4

QC material LOT dependency

The CyView™ script as well as the FCS Express™ templates are designed for a specific
Count Check Beads green (05-4011_R) LOT.
Check before QC procedures
LOT of used Count Check Beads green and revision of the script / templates are matching
(see 4.2 Quality Check procedure).
"LOT N° mismatch" will be displayed in the FCS Express™ templates, if a template not
matching the Count Check Beads green LOT was used.
The Quality Check is designed solely for QC material stated in chapter 3.3 Quality Check
material. Similar products, such as Count Check Beads green (REF 05-4011) are not
suitable.
Matching LOT and template
The LOT number (LOT N
o
) of the Count Check Beads green is composed of
"NN00YYMMDD". They are compatible with the script / template version, identical to the
first 4 digits of the LOT N
o
"NN00".
To check if the LOT matches, start the FCS Express™ template manually and type in the
LOT N
o
of the Count Check beads green. If the LOT N
o
does not match "LOT N° mismatch"
will be displayed.
Find the template at: "C:\ProgramData\PartecGmbH\Cube_18\templates\Quality Control"