4.2.3

Analysing the Quality Check with FCS Express™

FCS Express™ opens automatically, once the QC procedure has finished. The QC results
will be analysed automatically by FCS Express™.
The first 4 digits of the LOT number need to be entered (e.g. "AK01") in the first part of the
LOT N° field. Make sure that the field contains no further characters, such as spaces.
(before Enter Quality Check material information)
#
Description
1
Input field Quality Check material information:
LOT Number (example "AK01")
2
Input field Quality Check material information:
Expiry Date
3
Input field Quality Check material information:
Reference concentration
4
QC report summary: Device status: Valid or
Invalid
CyStain™ VitalCount QC Procedure | QC Manual | June 2021
1
2
3
4
5
6
7
Figure 13: Exemplary QC report (Page 1 / 2)
#
5
6
7
(after Enter Quality Check material information)
Description
Summary of performed tests Test: Pass or Fail
General information Device serial number,
CyView Operator, Sample ID and Date
FCS Express™ template version