4 Operation

4.1 Software

CyView™ is used for control of the device, data acquisition, data analysis and storage. For
more information, please refer to the CyFlow™ Cube 6 V2m manual.
FCS Express™ is used for reporting and analysis of data. The software can be accessed
directly from within CyView™. Refer to FCS Express™ manuals and tutorials for detailed
information about the software.
4.2

Quality Check procedure

4.2.1

Setting up the Quality Check

Requirements
Logged in to CyView™ with privileges of Main User level or above (Administrator)
LOT and FCS Express™ template match, see chapter 3.4 QC material LOT
dependency
CyStain™ VitalCount configuration is active
Priming has been performed
Sample tube with 850 μl Count Check Beads green, see chapter 3.3 Quality Check
material
Sample tube with 1500 μl Sheath Fluid
Procedure
1. Select [Work Mode].
2. Switch Measure Mode Sample Source to "Sample Port".
3. Connect the Sample tube with 850 µl Count Check Beads green.
4. Click [Start].
5. Adjust gain values in a way that the signals of the Count Check Beads green appear
exactly in the center of the quality gate "QC (CCB)" within the main Plot.
1
E.g. VitalCount_A_AK01_Robby.cvc85
CyStain™ VitalCount QC Procedure | QC Manual | June 2021
1
.
Figure 6:Selecting Work Mode
DQC Plot 1
DQC Plot 2
Figure 7:Adjust gain values
4 Operation
7