1 SAFETY PRECAUTIONS
1.2 SAFETY PRECAUTIONS CONTINUED
• Do not apply the Ultrasound Applicator to the
patient during the Head Warming period.
Applicator must remain in Applicator Hook during
the Head Warming period.
se only degassed water in power meter for
U
testing Ultrasound Applicators. Use of other types
of water will cause false test results. see page 12
for degassed water recipes.
• Use of other brands or types of tools, equipment,
fixtures, materials, and supplies other than those
specifically listed on
page 12
and calibration results.
• If proper equipment is not available or cannot be
obtained, send the Ultrasound Applicators to the
factory for calibration.
o not aerate water when filling power meter
• D
• Unplug the unit from the power source and/or
remove Battery before attempting any removal or
replacement procedures to prevent electrical
shock.
• Dielectric Withstand Test must always be
performed after a Power Supply has been
replaced. Field technicians unequipped to preform
the Dielectric Withstand Test should contact the
manufacturer for Power Supply replacement.
• Handle Power Supply shield with caution.
Penetration of the shield may result in
damage to the unit and/or personal injury.
• The Power Supply will retain high voltage in
the Power Supply Capacitor C4. Discharge
Capacitor using a multimeter prior to any
service of the Power Supply
• Applicator must be calibrated after removal
and replacement of any part, using an
Ohmic UPM DT 10 or UPM DT 100.
Failure to calibrate a repaired Applicator,
using the proper equipment, may result in
serious injury or harm to the patient.
will give bad test
.
3
Intelect TranSport ® /Vectra Genisys ® Ultrasound Therapy Systems

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Intelect TRANSPORT Vectra Genisys 2782 and is the answer not in the manual?

Questions and answers

Jessica Finley
February 11, 2025

How do you electrical safety test this

1 comments:
Mr. Anderson
February 12, 2025

To perform electrical safety testing on the Chattanooga Group Intelect TRANSPORT Vectra Genisys 2782, follow these steps:

1. Power Requirements Check: Ensure the unit operates within the specified input range of 100-240VAC, 75 VA, 50/60 Hz.

2. Leakage Testing: Conduct all necessary leakage tests according to NFPA 99 (National Fire Protection Association) “Health Care Facility” standards.

3. Dielectric Withstand Test: If the unit fails this test or the leakage test, it may indicate serious internal issues. Do not attempt to repair it yourself; send it to the factory for repair.

4. Visual Inspection:
- Check for any visible signs of damage or abuse.
- Ensure the fan is functioning correctly.

5. Unit Startup and Fan Testing:
- The fan should function only when the ultrasound is being emitted.
- If the fan is not blowing outward, check if it is wired incorrectly and rewire or replace it if necessary.
- If the fan is not operating at all, it may be faulty and require replacement.

Following these steps ensures compliance with safety standards and proper functionality of the unit.