5 TROUBLESHOOTING

5.8 POWER SUPPLY TEST

A. Test
1. Separate top from bottom. Refer to 6.1,
part C.
2. Disconnect the Power Supply Harness from
the Control Board. Lay the unit open.
3. Using a multimeter, hold the red wire of the
multimeter to the prong on the connector
for the red wire of the Power Supply Harness.
Hold the black wire of the multimeter to
the prong on the connector for the black
wire of the Power Supply Harness. Refer to
Figure 5.3.
B. Test Results
1. Power Supply reads approximately 24 VDC,
Power Supply is good.
2. Power Supply does not reads approximately
24 VDC, Power Supply failed test.
a) Replace Power Supply.
Intelect TranSport ® /Vectra Genisys ® Ultrasound Therapy Systems
FIGURE 5.3
16