6 REMOVAL & REPLACEMENT

6.2 FAN

Unplug the unit from the power source and/or
remove battery before attempting any removal or
replacement procedures to prevent electrical shock.
A. Part Numbers ....................................... 27158
B. Tools & Equipment Required
#1 Phillips Screwdriver
C. Fan
1. Separate top from bottom. Refer to 6.1,
part C.
2. With #1 Phillips screwdriver, remove the 2
screws securing the Fan. Refer to
Figure 6.4.
3. Remove Fan from the unit.
4. Remove Fan Baffle from Fan. Refer to
Figure 6.4.
5. Install the replacement by reversing the
steps above.
NOTE:
Refer to Figure 6.5 to route the Fan
Harness when installing. Make certain that
the Fan Baffle sits onto the system housing
baffle tabs. Refer to Figure 6.6.
Make sure the Fan mounting screws are
tightly secured. Refer to Figure 6.7.
Intelect TranSport ® /Vectra Genisys ® Ultrasound Therapy Systems
REMOVE SCREWS
22
FIGURE 6.4
ROUTE
HARNESS AS
SHOWN
FIGURE 6.5
BAFFLE
TABS
FIGURE 6.6
SECURELY TIGHTEN
MOUNTING SCREWS
FIGURE 6.7
FAN
BAFFLE