6 REMOVAL & REPLACEMENT
6.1 SEPARATING TOP FROM BOTTOM
Unplug the unit from the power source and/or
remove battery before attempting any removal or
replacement procedures to prevent electrical shock.
A. Part Numbers
Intelect TranSport Top .............................. 27399
Vectra Genysis Top ........................................ 27244
Base ......................................................................... 27252
B. Tools & Equipment Required
•
#1 Phillips Screwdriver
C. Separating Top from Bottom
1. Remove the Plynth from the bottom of the
system. Refer to Figure 6.1.
2. Remove the Mains Power cord from the
system. Refer to Figure 6.1.
3. Remove the rear Fan Grill from the system to
gain access to the two rear mounting screws.
Refer to Figure 6.2.
4. Remove the two front feet from the system
to gain access to the two front screws.
Refer to Figure 6.3.
5. Remove the four mounting screws from the
bottom housing using the #1 Phillips
screwdriver. Refer to Figure 6.3.
6. Carefully separate the upper and lower
housings of the system.
7. Disconnect the Power Supply and Fan
Harnesses from the Control Board of the
system. Refer to Figure 6.3.
8. Reassemble the unit by reversing the steps
above.
NOTE:
Do not overtighten the screws.
Overtightening will damage the threads of
the brass inserts.
Intelect TranSport ® /Vectra Genisys ® Ultrasound Therapy Systems
REMOVE
PLYNTH
REMOVE
FAN GRILL
REMOVE
FOUR SCREWS
21
REMOVE MAINS
POWER CORD
FIGURE 6.1
REMOVE
FRONT FEET
FIGURE 6.2
DISCONNECT POWER
SUPPLY AND FAN
HARNESSES
FIGURE 6.3