4 SPECIFICATIONS

INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY SYSTEMS

4.1
Figure 4.1 below provides physical details of the
Intelect Transport/Vectra Genisys Ultrasound Therapy
Systems. This section also provides ultrasound
specifications to aid in troubleshooting.
A. Intelect Transport/Vectra Genisys Ultrasound Therapy Systems Physical Specifications
DEPTH
DEPTH
Dimensions
Width.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 28.8 cm (11.3")
Height . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 16.3 cm (6.4")
Depth.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 32.8 cm (12.9")
Weight
Standard Weight (with base) .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .2.3 kg (5.07 lb)
Battery Pack . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .0.85 kg (1.87 lb )
Power
Input. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 100 - 240 VAC, 50/60 Hz 75 VA
Output . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . +24 V, 1.0 A
Electrical Class .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. CLASS I
Electrical Type
Ultrasound TYPE B.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .
Battery Type . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. Nickel Metal Hydride (NiMH)
WIDTH
WIDTH
FIGURE 4.1
(1.2 V x 20 size AA)
8
Intelect TranSport ® /Vectra Genisys ® Ultrasound Therapy Systems
Refer to this section when performing
troubleshooting, replacement, and repair of the
Intelect Transport/Vectra Genisys Ultrasound Therapy
Systems.
Operating Environment
Temperature .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..Between 59° F and 104° F
Relative Humidity .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 30%-60%
Atmospheric Pressure .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 950-1,050 h Pa
Complies with:
Listed by Intertek Testing Services NA Inc.
Conforms to UL Standard 60601-1
Certified to CAN/CSA Standard C22.2 No.
601.1-M90 w/A2
9700675
Compliant to IEC/EN 60601-1,
IEC 60601-1-2, 60601-2-5
HEIGHT
HEIGHT
(15° C and 40° C)