[Determining the unit weight through the weight measurement of
samples.]
First, enter the number of samples to be placed for weight measurement.
With the numeric keypad, enter '1' '0' as the number of items to be
placed, and then press
* Pressing
selection screen.
Here, samples are weighed to determine the unit weight.
Place the specified number of samples (10 samples) on the instrument.
Then, after the measured value becomes stable, press
* For information, measured weight appears on the Sub-display.
* Zero point reset and One-touch tare can be used during sample weight
measurement.
* Long-pressing
sample and returns to the sample size entry screen.
* When you are editing the program data already registered, press
without anything put on the instrument. Then the system keeps the
previous setting of sample weight measurement and proceeds to the
next step.
[Entering a preset tare.]
* If there is no need to enter this setting, leave this as zero and press
.
With the numeric keypad, enter '2' '0' as the preset tare, and then press
.
* Pressing
screen.
The unit weight is calculated using the weight of the placed items and the
number of the samples. The setting is now complete.
The data has been registered in the program number 8. You can call the
registered data even if the power is turned off.
Now, you can proceed with the work using the Counting function.
Counting function - How to weigh items
Press
to call the program number selection screen. With the numeric
keypad, enter '8' as the program number, and then press
the registered program is called, you can start work.
* You cannot call the settings of an unregistered program number through
this procedure.
Alternative calling method (Easy program call)
When the Main-display shows 0, enter '8' with the numeric keypad. Then
press
when the indication is blinking. You can also call program
data in this way.
When you place the items on the instrument, the Main-display shows the
number of the items.
When the upper/lower limits are set, the blue lamp comes on depending on
the number of items placed.
* While the unit weight is displayed during the selection and call of a program, the indicated unit weight is
rounded off to one tenth of the scale interval. Because–during actual calculation–the function uses a value
that is not yet rounded off, the measured weight is not always equal to [Indicated unit weight x Quantity].
(To return to the screen on which no program is called)
Remove all items from the instrument. When the Zero-point sign comes on,
press
.
* If the Zero-point sign does not come on, you cannot switch the
indications.
.
on the right screen returns to the program number
discards the weight measurement of the current
on the right screen returns to the sample weighing
Operation
Chapter 6
.
. Now that
42
How to use the Multi-functions