7.6 Adjustment with external weight

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each chan-
ge of location as well as in case of fluctuating environment temperature. To receive
accurate measuring values it is also recommended to adjust the balance periodically
in weighing operation.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is re-
quired for stabilization. Ensure that there are no objects on the weighing plate.
Ensure that there are no objects on the weighing plate.
Press the
key and keep it pressed down whilst simultaneously operating the
key
Weight value of adjustment weight is flashing on display. However, you may enter
a value of your choice via the number keys. *
a
PRE-
TARE
Carefully place adjusting weight in the centre of the weighing plate. Adjustment will
be carried out automatically after dead stop control.
Remove adjustment weight during count down to zero.
The balance returns automatically into weighing mode. An error message will ap-
pear on the display should a adjustment error occur or should the adjustment
weight be incorrect. Turn balance off, then restart it and repeat the adjustment
process.
* The adjustment should be made with the recommended adjustment weight (see
chap. 1 "Technical data"). Weights of different nominal values may be used for ad-
justment but are not optimal for technical measuring.
Info about adjustment weights can be found on the Internet at: http://www.kern-
sohn.com
Note
Use the
key to exit adjustment mode. The balance returns to weighing mode.