2.6 Attaching a Strainer

Part
Number
Strainer
60-9004-367 Weighted Strainer-
stainless steel ends,
polypropylene center
69-2903-138 Low Flow Strainer-
Stainless Steel
60-3704-066 CPVC Strainer
Figure 2-10 Strainers
To select the right strainer for your application, see Table 2-2 and
Figure 2-10.
To attach the strainer to the suction line, carefully screw the
strainer's threaded connector into the suction line. If attaching
the strainer to an FEP-lined suction line, heat the end of the
suction line before threading the strainer into the line. Warming
FEP-lined tubing will make it more pliable to avoid damage.
Table 2-2 Strainers
dia.
Application
A
Routine
10 mm
applications
0.40 in
Low flow
10 mm
applications
0.40 in
Highly acidic
10 mm
liquids, metals
0.40 in
sampling
Weighted
60-9004-367
Isco 4700 Refrigerated Sampler
dia.
length
B
C
32 mm
267 mm
1.25 in
10.5 in
14 mm
178 mm
0.562 in
7.0 in
33 mm
292 mm
1.3125 in
11.5 in
Low Flow
69-2903-138
øB
Section 2 Installation
Maximum
Unanchored Depth
Vinyl
FEP-lined
6.7 m
4.5 m
22 ft
15 ft
6.7 m
4.5 m
22 ft
15 ft
1.2 m
N/A
4 ft
øA
CPVC
60-3704-066
C
2-13