5.4.3 Control Panel Module
60-4704-035
Isco 4700 Refrigerated Sampler
Section
5
Maintenance
&
it
WARNING
Risk of injury or equipment damage. Familiarize yourself
with the general instructions in section 5.4.1 before
proceeding.
1. Open the top cover.
2. Remove the mounting screws (see Figure 5-7).
3. Pull the module away from the refrigerator body to expose
the wiring connectors.
4. Disconnect the wiring connectors identified in Figure 5-8.
5. Clean the control panel module's mounting surface on the
face of the refrigerated compartment. This will help ensure
that the gasket on the new module will seal the enclosure.
6. Attach the wiring connectors to the new module.
7. Align the module over the mounting holes and secure it
with the mounting screws. Torque screws 1.8 to 2.0 Nm (16
to 18 in/lbs).