Install
April 2022
Figure 2-7: Rosemount Oxymitter DR Bracing Installation
Bracing is for vertical and horizontal transmitter installation. External bracing is required
for 9 ft. (2.74 m), 12 ft. (3.66 m), 15 ft. (4.57 m), and 18 ft. (5.49 m) probes as shown
below.
Note
Dimensions are in inches with millimeters in parentheses.
A. Brace bars (not by Rosemount)
B. Brace clamp assembly by Rosemount. (Both brace clamp assemblies are the same. You
C. Two holes: 0.625 (16) diameter for 0.50 (12) diameter bolt
D. Abrasive shield
Procedure
1. Weld or bolt adapter plate onto the duct.
2. If using the optional ceramic diffusion element, make sure to correctly orient the
28
must install and locate clamp assemblies and brace bars in the field).
vee deflector. Before inserting the transmitter, check the direction of the flow of
Reference Manual
00809-0100-4884
Emerson.com/Rosemount