Precautions When Handling H Reagent - Phcbi MCO-230AICL-PE Operating Instructions Manual

Co2 incubator
Table of Contents

Advertisement

H
O
DECONTAMINATION (OPTIONAL)
2
2
16. Press "OK" to return to the Top screen.
Note: The OK key does not work when the H
generator is attached to the connector by the cable.
17. Dilute the remaining H
Rinse and wash the H
outside the chamber.
Note: Do not wash either the inside or outside of the H
18. After H
O
decontamination, surplus H
2
2
bottom of the chamber and in the bottom part of the H
duct. This solution contains H
glasses and rubber gloves and wipe it up with a non-woven cloth.
19. Ventilate the chamber sufficiently and place all the attachments
back inside the chamber.
Note: After H
O
decontamination, replace the connector cap firmly
2
2
over the connector on the side of the chamber (Fig. 4).
Precautions when handling H
Observe the following precautions when handling optional H
 Handling precautions
• Wear protective equipment, such as protective glasses and rubber gloves.
• Do not start a fire in the area where the reagent is being handled.
• Do not leave any reagent in the container after it has been used or while it is being used.
• Do not place flammable or combustible materials near the area where the reagent is handled.
 Precautions for storage
• Store in a cool, dark place.
• Always close the container cover securely to prevent impurities from becoming mixed into the reagent.
• Check the container to make sure that there is no damage, corrosion, or cracking.
• Store the container with the inlet facing upwards, and make sure that the container will not tip or be
knocked over.
 Precautions for disposal
• Dispose in accordance with local laws and regulations.
54
O
reagent in the H
O
2
2
2
O
generator with distilled water. Keep the H
2
2
O
liquid will remain at the
2
2
O
at a low density. Wear protective
2
2
O
2
2
generator with a large volume of water and dispose of it.
2
O
generator in a clean environment
2
2
O
generator with alcohol.
2
2
O
generator
2
2
O
reagent
2
2
O
reagent MCO-H2O2:
2
2
Connector cap
Fig. 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mco-230aicuvl-pe

Table of Contents