INSTALLATION

Installation
1. Remove the packing tape and clean up.
Remove all the tapes that are securing the doors and the inner attachments. Open the doors for ventilation.
If the outer panels are dirty, use a cloth to wipe them with a diluted neutral detergent (Undiluted detergent
can damage the plastic components. For the dilution, refer to the instructions on the detergent).
Wipe off the residual detergent with a wet cloth and then wipe off any moisture.
Note: Remove the cable tie around the power supply cord to prevent corrosion of the cord coating.
2. Set up the humidifying pan and humidifying pan cover (Fig. 1).
3. Install 4 trays (Fig. 2).
Humidifying pan cover
4 Trays
Humidifying pan
Fig. 1
Fig. 2
4. Adjust the levelling feet.
Adjust
the
levelling
feet
by
turning
them
anticlockwise until the incubator is level (Fig. 3).
Note: Incubating on a sloping tray may adversely
affect the cultivation.
Shrink
Stretch
Leveling feet
Fig. 3
5. Earth the incubator.
Earth the incubator during installation to prevent electric shock. If there is no earth wire at the location,
consult with qualified service personnel.