ALARM PARAMETERS
3. Enter the required settings for each alarm on the
Alarm Setting #2 screen. Press "Apply" to save the
settings.
 Settings
• Door Delay:
When the conditions for the door alarm are reached, the alarm buzzer will sound after the time set here
has elapsed. Settable range: 1 minute
30 minutes; factory setting: 2 minutes.
Note: When the conditions return to normal within the door alarm delay time, the buzzer will not sound.
• Ring Back:
If the alarm buzzer is stopped by pressing the "Buzzer" key, the buzzer will sound again when the
conditions that activated the alarm continue after the time set here has elapsed.
Settable range: 1 minute
99 minutes; factory setting: 30 minutes.
Note: At Err01 (CO
gas cylinder empty), Err11 and Err12 (CO
sensor error), Err18 (UV lamp failure), and
2
2
Door alarm, the alarm is deactivated by pressing the "Buzzer" key and so the buzzer will not sound again
(Refer to pages 64–65).
• Remote Alarm:
The remote alarm continues even if the buzzer is stopped by pressing the "Buzzer" key.
Factory setting: ON.
When the alarm sounds, the buzzer can be
silenced by pressing "Buzzer". Refer to Tables 5–7
on page 65 for the way the incubator behaves at
the time of pressing "Buzzer" and the re-activation
of the alarm under each condition.
Resolve the cause of the alarm, referring to pages
64–66. The alarm itself is not deactivated by
pressing
"Buzzer"
(except
for
those
cases
mentioned in the previous note).
Note: The buzzer for the high-limit temperature
alarm cannot be silenced.