INSTALLATION

Installation site

For correct operation, install the incubator in a location with the following conditions (referring also to the
precautions for safe operation on pages 5–9):
 Install in an environment with normal air.
 Do not expose to direct sunlight.
 Do not install near significant heat sources, such as heaters, boilers, ovens, or autoclaves.
 Ensure ambient temperature is at least 5 C lower than set temperature.
 Do not allow the ambient temperature to become too high.
 Install in a site with a relative humidity of 80 %R.H. or lower.
 Do not expose to flammable or corrosive gas.
 Do not install where objects may fall on the incubator from above.