7. Remove the connector cap from the connector
on the bottom right of the far side of the chamber.
Connect the H
O
generator and connector by the
2
2
cable provided (Fig. 2).
Note: Be sure to keep the connector cap.
8. Insert four trays in the 2nd, 3rd, 4th, and 6th tray catches from the top of the chamber side.
Note: The trays included as accessories are designed to be appropriate for decontamination. If half trays
(MCO-35ST, optional) or trays for previous models are used, decontamination may not be sufficiently
effective.
9. Insert the humidifying pan cover and humidifying pan removed in procedure 1 (Fig. 3).
Front
10. Make sure that the duct, the fan, and the H
Then close the inner and outer doors.
Front
Humidifying
pan cover
Humidifying pan
H
O
2
2
generator
O
generator are securely attached (refer to page 24).
2
2
Connector
Fig. 2
Fig. 3
51