Basic Operation On Lcd Touch Panel; Operation From Menu Key - Phcbi MCO-230AICL-PE Operating Instructions Manual

Co2 incubator
Table of Contents

Advertisement

BASIC OPERATION ON LCD TOUCH PANEL

Operation from Menu key

Pressing the Menu key at the bottom right corner of the Top screen brings up the Menu screen.
From the Menu screen, the following options are available:
Press "Set" to bring up the Stand-by Setting Screen. Use this to set the temperature, CO
and high-limit temperature alarm (page 33).
Press "Log" to bring up the Log screen. Use this to display log graphs, export logs, set the log interval,
and set the unique ID (pages 37–44).
Press "Lock" to bring up the Lock screen. Use this to set the key lock (page 34), user-ID, password,
and optional Auto-lock (pages 55-58).
Press "Tools#1" to bring up the Tools#1 screen. Use this to set alarms (pages 35–36), set and operate
the optional UV lamp (pages 47–49), and calibrate CO
Press "Tools#2" to bring up the Tools#2 screen. Use this to set date, time, brightness, and sleep mode
(pages 45–46).
Notes:
• The Service key is available only to qualified engineers.
• Other functions, such as gas supply line changeover and H
from the Top screen.
To navigate away from any of the screens
accessed from the Menu screen, press "Top" to
return to the Top screen or "Back" to return to
the previous screen.
density (pages 62–63).
2
O
decontamination, are performed directly
2
2
density,
2
29