Phcbi MCO-230AICL-PE Operating Instructions Manual page 52

Co2 incubator
Table of Contents

Advertisement

H
O
DECONTAMINATION (OPTIONAL)
2
2
11. Press "H
O
" for 3 seconds to open the Setting
2
2
Position window. Enter the password if prompted.
12.
Press
"Next"
Decontamination
Step1
something to worry about setting position, check
the inside again.
13. The system check starts automatically. If an
alarm or safety function is activated, refer to Table 8
on page 66.
14.
If
the
system
Decontamination Step2 screen appears. Press
"OK"
to
start
H
decontamination is performed automatically from
Step3 to Step8 (procedure 15).
Notes:
• The outer door is electrically locked for safety until
H
O
decontamination is completed.
2
2
• A buzzer sounds when H
completed (after about 100 minutes).
52
to
display
the
screen.
If
there
is
normal,
the
O
decontamination.
2
2
O
decontamination is
2
2
H
O
2
2
is
H
O
2
2
H
O
2
2