OPERATION/ALARM LOG

Displaying alarm log

The incubator can save alarm log data (maximum 256 logs). Alarm log data can be displayed graphically
on the LCD touch panel.
Note: When number of saved alarm logs exceeds 256, the oldest log is deleted and overwritten.
1. Go to the Log screen (Menu  Log). Press
"Alarm" to display the Alarm screen.
2. Alarm logs from the last 7 days (including current
day) are displayed.
Note: When the number of alarm logs exceed 6,
pressing the top (▲) or the bottom (▼) log, scrolls
the log table and allows hidden alarm logs to be
seen.
3. Change the number in the "Last X Days" input
box to display alarm logs for specified days
(including current day).
Settable range: 1 day
45 days.
Note: Accurate to about 1 minute per month.
.
42