CHAPTER
F.5
Image quality of the camera
Test film
Camera test report
04/2012
F . 4

Camera Test

In the camera test, the image quality and the camera settings
are checked.
You can also film test images of the quality measurement in
order to meet requirements for documentation.
The camera was set for optimum image quality during installa-
F . 5
tion and should not be changed.
Please note the following points:
Use the correct film type (emulsion number).
Keep the developer bath at a constant temperature.
Use the correct chemicals.
Follow the development instructions supplied by the film
manufacturer.
After installation, a test film is recorded and evaluated. The
F . 5
results are summarized in a table. These values are used as ref-
erence values for the daily quality check of the camera.
B
If the equipment or the emulsion number of the film is changed,
F
. 5
the camera must be readjusted by a customer service technician
who is authorized to do so. After this, another camera test is
required.
You will find a form for the test report in (→ System Owner
F . 5
Manual)
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5