Siemens SOMATOM Spirit Operator's Manual page 406

Table of Contents

Advertisement

Constancy Test
Performing the pixel noise
test
F.3-22
The pixel noise is determined from two tomograms of the water
phantom in the same way as for the daily quality measurement.
F . 3
(→ Page F.2-1 Daily Quality Measurements)
Press the Start key on the control box.
Scanning is started.
F . 3
i
After the previous test has been completed, the Noise test is
F
. 3
started automatically.
Two images of the water phantom are recorded with identical
parameters. The difference between the first and the second
measurement is shown.
The following evaluations are displayed:
ROI
Mean value of the CT value
Sigma value (pixel noise)
Quality Assurance
SOMATOM Spirit
F . 3
F . 3
F . 3
F . 3
F . 3
F . 3
0 . 0
0 . 0
0 . 0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents