Function Tests
Checking radiation block
and warning lamps
D.2-4
You must check the radiation warning lamps on the control box,
on the gantry and, if present, next to the doors of the examina-
D . 2
tion room.
Check the proper function of the radiation warning lamps
✧
during the daily image quality tests.
The radiation warning lamps must light up when radiation is
generated.
Make sure that radiation stops immediately when a door of
✧
the examination room is opened
i
When the door is closed again you must restore system readi-
D . 2
ness with Setup > Continue.
Operation
D . 2
D . 2
D . 2