Siemens SOMATOM Spirit Operator's Manual page 264

Table of Contents

Advertisement

Function Tests
Checking radiation block
and warning lamps
D.2-4
You must check the radiation warning lamps on the control box,
on the gantry and, if present, next to the doors of the examina-
D . 2
tion room.
Check the proper function of the radiation warning lamps
during the daily image quality tests.
The radiation warning lamps must light up when radiation is
generated.
Make sure that radiation stops immediately when a door of
the examination room is opened
i
When the door is closed again you must restore system readi-
D . 2
ness with Setup > Continue.