Information about Personal Safety
A.2-16
CAUTION
Uncontrolled system movements and radiation!
Injury or radiation damage.
A . 2
Always observe the patient during system movements.
✧
Press STOP if a system movement is performed that has not
✧
been initiated on a gantry operator panel, on the control
box or during a scan (esp. at autorange).
Press STOP if a key sticks or a movement does not stop
✧
immediately when the key is released.
Press STOP if the patient table does not stop as expected.
✧
Press STOP if the patient table position displayed differs by
✧
more than 10 mm from the actual position.
Press STOP if the patient is not correctly positioned during
✧
system movements.
Press STOP if the patient table moves in the wrong direc-
✧
tion.
Press STOP if the HOLD key does not respond during a scan.
✧
Press EMERGENCY OFF if the system does not respond to
✧
the STOP keys and there is danger.
Shut down the system and notify the Siemens Customer
✧
Service if system malfunctions are detected.