Information about Personal Safety
A.2-16
CAUTION
Uncontrolled system movements and radiation!
Injury or radiation damage.
A . 2
Always observe the patient during system movements.
Press STOP if a system movement is performed that has not
been initiated on a gantry operator panel, on the control
box or during a scan (esp. at autorange).
Press STOP if a key sticks or a movement does not stop
immediately when the key is released.
Press STOP if the patient table does not stop as expected.
Press STOP if the patient table position displayed differs by
more than 10 mm from the actual position.
Press STOP if the patient is not correctly positioned during
system movements.
Press STOP if the patient table moves in the wrong direc-
tion.
Press STOP if the HOLD key does not respond during a scan.
Press EMERGENCY OFF if the system does not respond to
the STOP keys and there is danger.
Shut down the system and notify the Siemens Customer
Service if system malfunctions are detected.