Constancy Test; Performing The Constancy Test - Siemens SOMATOM Spirit Operator's Manual

Table of Contents

Advertisement

CHAPTER
F.3
04/2012
F . 2

Constancy Test

Together with other characteristic values, the daily test values
are checked for constancy monthly. These tests are performed
on the complete set of phantoms.
In addition, the CTDI (CT Dose Index) must be measured at least
once every six months. The CTDI must also be measured after
any maintenance work that might have altered the test results.
i
We recommend having the constancy test and the CTDI mea-
F
. 3
surement performed by proper personnel.
If you want to conduct the constancy test and the CTDI measure-
ment yourself, you must perform all steps of the constancy test
or the CTDI measurement under your own responsibility.

Performing the constancy test

The monthly quality test is performed via the Quality Con-
stancy dialog box of the Local Service window.
F . 3
F . 3
F . 3
F . 3
F . 3
F . 3
0 . 0
F.3-1
0 . 0
0 . 0

Hide quick links: