2.4

Interlock Operation

X-ray generation stops automatically whenever the Interlock system is interrupted. The
Interlock system may be interrupted due to user action, i.e., opening the chamber door,
or a fault in the system, such as a failed switch or a fault in the Interlock system wiring.
An error message may be generated and displayed on the monitor, depending on the
cause of the interruption.
System messages will be displayed when the chamber door is opened during an x-ray
exposure. The messages will vary depending on whether the system is in 2D or 3D (CT)
imaging mode.
5081-9544 Revision 004
Faxitron CT Specimen Radiography System User Guide
In 2D imaging mode "Motion Error" and "Radiography Scan Canceled" error
messages will be displayed. The error messages will be superimposed one over the
other. To clear the error messages, click the Okay button on the top message and
then the remaining message. See figures below.
In 3D imaging mode "Motion Error" and "CT Scan Canceled" error messages will be
displayed. The error messages will be superimposed one over the other. To clear the
error messages, click the Okay button on the top message and then the remaining
message. When a CT scan is canceled for any reason, i.e., door is opened, Stop button
is pressed, or loss of detector communication, the system will remain in STANDBY
for about 1 minute before it is ready for another scan. See figures below.

Figure 2: "Motion Error" and "Radiography Scan Cancelled" Error Messages

Chapter 2: General Information
Page 13