Keysight Technologies S9106AC-2IF Startup Manual page 56

Multi-port transceiver solutions
Hide thumbs Also See for S9106AC-2IF:
Table of Contents

Advertisement

6 Verify Operation
4. Verify IQ Analyzer Mode displays a signal at 10 GHz and –10 dBm.
a. Select Peak Search from the drop-down menu panel (top-right corner).
If there is a signal at 10 GHz of approximately –10 dBm,
RF Out ➃
RF In ➄
RF Transceiver 7 - 15 GHz,
and
of the S9106AC-2IF are working properly. (The power level is only
approximate because there is loss in the cable being used and this
loss is not being corrected during the measurement.)
52
Keysight S9106AC-2IF and S9106AC-4UP 5G FR1 Multi-port Transceiver Solutions, Startup Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

S9106ac-4up

Table of Contents