8 Tips and assistance
8.3.
Updating instrument software for testo 330-2
LL V2010 (0632 3307)
At www.testo.com/download-center you can download the current
instrument software (firmware) for the testo 330-2LL (0632 3307)
(registration required).
> Unplug the mains unit and switch off the testo 330.
1. Hold down [▲].
2. Plug in the mains unit, continue holding down [▲].
-
The display shows
3. Release
4. Insert the connecting cable (0449 0047) into the USB-port on
the instrument, then connect it to the PC.
-
Your PC recognises the testo 330 as a removable medium.
5. Copy the new file (ap330rel.bin) to the detected removable
medium.
-
In the display the status bar progresses from left to right. This
process may take a few minutes.
6. Disconnect the connecting cable from the testo 330.
-
Once the instrument software (firmware) has been updated, the
system will automatically reboot and is ready for use.
8.4.
Updating instrument software for testo 330-
2 LL/F (0632 3305)
At www.testo.com/download-center you can download the current
instrument software (firmware) for the testo 330-2 LL/F (0632 3305)
(registration required).
8.5.
Updating instrument software for testo 380
At www.testo.com/download-center you can download the current
instrument software (firmware) for the testo 380 (registration
required).
1. Connect particulate matter probe to the testo 380.
2. Switch on the testo 380.
3. Insert the connecting cable (0449 0047) into the USB port on
the testo 380, then connect it to the PC.
4. Open the executable file USBFlashUpdate.exe.
-
The window
5. Select testo 380 or particulate matter probe.
6. Select connected testo 380.
82
Firmware update
[▲]
.
testo 380 Update (USB)
along the bottom edge.
is displayed.