6 Using the product
1. Place cover cap (1) on the fresh air inlet of the condensate
container.
2. Place cover cap (2) on the probe shaft of the particulate matter
probe.
-
The tightness of the crude gas and measurement gas path is
displayed (crude gas path value < 0.02 l/min, measurement gas
path value <0.4 mbar).
-
If leaks in the crude gas path and/or the measurement gas path
were exposed during the tightness test, see Questions and
answers, page 75.
3. Once the tightness test has been successfully conducted,
remove the cover caps from the fresh air inlet and probe shaft:
[OK].
4. Finish the tightness test: [Finished].
6.3.2.
Particle: test types – compliance measurement,
adjustment help and approval test
Call up function:
>
[
40
To prevent incorrect measurements, the position of the
testo 380 and the particulate matter probe must not be
changed during a measurement.
For flue gas loss measurement, Testo recommends using
a combustion air temperature probe with connecting cable.
Self-heating of the measurement system during operation
may influence the combustion air temperature
measurement with a mini ambient air probe.
]
Measurement options
[OK]
Particle
→ [OK].