Cleaning The Rotation And Stator Disc In The Rotation Diluter - TESTO 380 Instruction Manual

Fine particle analyzer
Table of Contents

Advertisement

7 Maintaining the product
8. Insert connector into the filter chamber.
9. Fill the filter chamber evenly with 5 g new cotton wool. Do not
compress cotton wool unnecessarily.
10. Place the primary filter back on the cotton filter and press into
position. Pay attention to the notch on the primary filter.
11. Insert cotton filter into the condensate container and close
(clockwise).
12. Insert the condensate container into the testo 380 and click into
place. Push the tube onto the connector.
7.7.
Cleaning the rotation and stator disc in the
rotation diluter
1. Take the particulate matter probe out of the holder in the
transport case.
2. Open the rotation diluter cover (turn anti-clockwise).
58
If is advisable to clean the rotation diluter after every
measurement.
The rotation diluter is aligned with the measurement
system.
> In order to prevent incorrect measurements, rotation
discs must not be exchanged between measurement
systems.

Advertisement

Table of Contents
loading

This manual is also suitable for:

0632 3800

Table of Contents