<When the optional RS232C is in
use>
This code appears when an unknown
command is received.
<When the optional RS232C is in
use>
This code appears when an unknown
header is received.
<When the optional RS232C is in
use>
This code appears when received
data such as tare weight exceed the
rated operating range.
<When the optional RS232C is in
use>
This code appears when a received
header and the unit of numeric values
are incorrect.
<When the optional RS232C is in
use>
This code appears when the BCC
values are not matched.
<When the optional RS232C is in
use>
This code appears when the
connected equipment transmits too
many data.
This code appears when the
instrument has received an addition
command during pre-stable condition.
<When the optional RS232C is in
use>
This code appears when a received
zero adjustment command specifies
an invalid zero adjustment range.
<When the optional RS232C is in
use>
This code appears when the
instrument has received a tare
subtraction command, but is not ready
for tare subtraction.
24
Check the data transmitted by
the connected equipment.
Press the
key to reset the
error and then restart
communication.
Check the data transmitted by
the connected equipment.
Press the
key to reset the
error and then restart
communication.
Check the data transmitted by
the connected equipment.
Press the
key to reset the
error and then restart
communication.
Check the data transmitted by
the connected equipment.
Press the
key to reset the
error and then restart
communication.
Check the results of BCC
calculated by the connected
equipment. Press the
to reset the error and then
restart communication.
Divide the transmission data of
the connected equipment into
two parts and then transmit
each part separately.
Alternatively, reduce the
volume of the transmission
data.
Press the
key to reset the
error and then restart
communication.
Press the
key to reset the
error. After the instrument has
become stable, transmit the
addition command again.
Transmit a command within the
valid zero adjustment range
(within the range of ± 1.9% of
the weighing capacity)
Press the
key to reset the
error and then restart
communication.
Check the specifications/status
of the instrument. Make the
instrument ready for tare
subtraction and then transmit
the command.
Press the
key to reset the
error and then restart
communication.
key