Error Code - Yamato DP-5601 Operation Manual

Table of Contents

Advertisement

Error code

Panel display
Conditions that trigger the error code
When there are any objects placed on the
platform of the instrument, pressing the
key to turn ON the display may
generate this code.
When there are any objects placed on the
platform of the instrument, pressing the
key to perform the zero reset may
generate this code.
When foreign material is caught between
the platform of the instrument and the
container or when the platform is
removed, pressing the
ON the display may generate this code.
When the platform of the instrument is
removed, pressing the
perform the zero reset may generate this
code.
Placing any objects that are heavier than
the weighing capacity of the instrument
displays this sign during operation.
Removing the platform of the instrument
displays this sign during operation.
When the number of measurement times
added has reached the maximum number
(9999), further addition of the number
generates this code.
This code appears when the added value
exceeds the maximum value
(999,999,999).
This code appears in the event of
abnormal A/D conversion., or when the
connected optional journal printer does
not function properly due to a paper jam,
etc.
<When the optional RS232C is in use>
This code appears when connected
equipment is not ready for receiving a
signal.
23
Corrective action
Remove the object from the
platform and press the
key.
The panel shows "0.00";
and the instrument is
operational.
Correctly install the
platform and press the
key to turn
key.
The panel shows "0.00";
and the instrument is
key to
operational.
Removing the object
returns to the weighing
display. Use the instrument
within its rated operating
range.
Correctly install the
platform.
The panel shows "0.00";
and the instrument is
operational.
Press the
clear the number of
measurement times added
by referring to page 15.
Press the
the error. Then clear the
added value by referring to
page 15.
Check the printing section
for the paper jam, etc.
Power OFF and then ON
by pressing the
and then the
Check the connector
connection. Press the
key to reset the error
then restart
communication.
key. Then
key to reset
key
key.
and