The male Luer fittings can be inserted into the fluid ports on the front of the fluid cell, and the
opposite side of the fitting is connected to the silicone tubing. The female Luer fittings connect the
silicone tubing to the syringe.
Use of the O-rings is optional. Instructions for operating the fluid cell with and without the O-ring
are detailed in this chapter. The previous O-ring design was based on a circular cross-section. The
new O-ring design has an "S" shaped cross-section with thin, flexible walls. This new design makes
insertion of the O-ring easier, reducing the chances of leaking and minimizing imaging artifacts
caused by lateral friction.
Two additional fluid cell designs are available from Veeco:
These fluid cells have additional ports for the insertion of a reference electrode and a counter
electrode.

8.2.1 Clean Fluid Cell and O-ring

To reduce contamination problems and to obtain high-quality images, clean the fluid cell, and O-
ring if applicable, as follows:
1. While soaking the fluid cell and O-ring in warm, soapy water, place a few drops of liquid
dish soap on them. Use nonabrasive soap to prevent scratching the glass surfaces.
2. Gently rub the fluid cell and O-ring with a cotton swab or finger. Avoid scratching the glass
surface with abrasive material.
3. Using distilled water, rinse the fluid cell and O-ring of all soap.
4. Using 0.2µm-filtered, compressed air or dry nitrogen, blow dry the fluid cell until all
moisture evaporates.

8.2.2 Select the Probe

AFM probes featuring low stiffness cantilevers produce the best results for biological applications.
We recommend sharpened tips to start: 100µm ("short"), V-shaped cantilevers with oxide-
sharpened silicon nitride tips; for example, models NP-S (standard) or NP-STT (oriented twin tip).
Stiffer etched silicon probes (model FESP, single beam, 225µm long) used at very small oscillation
amplitudes also obtain good results. Experiment to find which probes work best for your sample.
Rev. B
Contact Mode electrochemical fluid cell (Model ECFC)
TappingMode electrochemical fluid cell (Model MMTMEC)
Note:
It is necessary to connect an external potentiostat to perform electrochemical
measurements. Contact Veeco for more information.
Note:
For additional information on selecting a probe, please refer to the applications
note "Choosing AFM Probes for Biological Applications."
MultiMode SPM Instruction Manual
Fluid Operation
General Fluid Operation
125