A separate motor drive controls the height of the microscope and tip relative to the sample surface. This manual details facility requirements, installation requirements and procedures, maintenance requirements and procedures, and applications used with the Dimension 3100 Scanning Probe Microscope. This chapter discusses the following: •...
• Applications: Section 1.6 • Maintenance and Troubleshooting: Section 1.7 System Overview There are two typical configurations of the Dimension 3100 Scanning Probe Microscope (see Chapter 3 for detailed information). • VT-103-3K with ELCON • VT-102 The configurations provide options for acoustic and mechanical vibration isolation, as well as various options for positioning control station components.
The Dimension 3100 ships with a high quality tower-style Pentium PCI computer system. Control Station Overview The Dimension 3100 SPM control station consists of four components: input and display devices (keyboard, trackball, mouse and monitors), computer, NanoScope controller and Dimension 3100 controller.
System Overview Control Station Overview 1.2.2 Computer The computer is the main control unit of the Dimension 3100 SPM system; it supports a 100MB Zip® disc drive, CD-ROM, CDRW, or DVDRW drive and 1.44MB floppy disc drive (see Figure 1.2b). The computer receives data from the input devices, and controls external hardware via the standard ports and the input/output (I/O) bus.
25-pin D cable connection between the rear of the NanoScope IIIa and the computer (see Figure 1.2c). The NanoScope IIIa connects to the back of the Dimension 3100 via a 37-pin D cable connection from the front of the controller (see Figure 1.2c).
Control Station Overview 1.2.4 Dimension 3100 Controller The Dimension 3100 controller controls the vacuum and air supply and optics illumination. The Dimension 3100 controller is controlled via a serial cable connection between it and the computer (see Figure 1.2d). The Dimension 3100 controller features gauges on the front panel to indicate vacuum and positive pressures.
Illumination can be computer or manually controlled. The knob located on the front of the Dimension 3100 controller regulates manual illumination control. For computer control, there is an “illumination” parameter in the Other Controls panel to control brightness. The illuminator control knob must be in the “off”...
For more information, contact Veeco. The Dimension 3100 Microscope Electronics Box houses the following components: Main SPM Circuit Board This smaller circuit board mounts on a larger board and contains switches to route feedback signals from the head.
3.5mm clearance between the tip and sample. 1.3.3 Stage System The improved Dimension 3100 X-Y stage provides substantially better positioning repeatability (3µm unidirectional and 4-6µm bidirectional) than its predecessor, the Dimension 3000. The stage is more than twice as fast for moving from one location to another.
Preamp Board The preamp board is located inside the Dimension 3100 SPM microscope head. The preamp board contains a preamplifier circuit for both photodetector signals, a laser diode power supply circuit that regulates the output of the laser, and ±12 volt regulators for the preamp circuit. The voltage regulators include current limiting to protect the circuit.
Dimension SPM head. The electrodes are oriented as shown when viewing the Dimension 3100 SPM from the front. With the Scan angle parameter in the control panel set to 0.00, the fast-scan direction is parallel to the front or in the direction of the X-axis.
System Overview Dimension 3100 SPM Overview Figure 1.3d Dimension Head Scanner The end of the scanner allows easy removal of the imaging probe assembly. This permits loading the tip on a separate fixture such that head removal becomes unnecessary. Table 1.3a Scanner Piezo Tube Specifications Specification...
System Overview Dimension 3100 SPM Overview 1.3.5 Cantilever Holder The cantilever holder is a small printed circuit card or acrylic block that holds the cantilever firmly at the proper angle. The standard cantilever holder is used for operation in air, while a clear glass cantilever holder is used for fluid cell operations.
Focus tracking and automated engagement Sample Size & Handling Samples may be up to 8 inches in diameter and 0.5-inch thick using a multipurpose, 8-inch chuck. For samples thicker than 0.5", contact Veeco. Table 1.4a Sample Specifications Specifications Wafers and disk media: 8-inch dia. x 0.5"...
Dimension 3100 SPM. Applications Several applications can be applied using the Dimension 3100 SPM. For specific information regarding these applications, please refer to the appropriate chapters in this manual. The following is a list of common applications used with the Dimension 3100 SPM: •...
Chapter 2 Safety This chapter details the safety requirements involved in installation of the Dimension 3100 Scanning Probe Microscope. Specifically, these safety requirements include safety precautions, environmental conditions, and equipment safety applications. Training and compliance with all safety requirements is essential during installation and operation of the Dimension 3100 SPM.
Ce symbole indique un risque de la matière corrosive. La matière corrosive peut causer des dommages sérieux au matériel. Dieses Symbol bedeutet “Gefahr durch Korrosive Materialien”. Korrosive Materialien können erhebliche Schäden am Gerät oder anderen Gegenständen verursachen. Rev. D Dimension 3100 Manual...
Safety Precautions Safety Precautions Because the Dimension 3100 SPM features independently motorized components, it is crucial that operators become familiar with precautions to avoid injury to themselves and/or damage to samples. This section of the manual should be read by ALL persons working with or around the system.
Page 41
CAUTION: Please contact Veeco before attempting to move the Dimension 3100 SPM system. ATTENTION: Il est impératif de contacter Veeco avant de déplacer le Dimension 3100 SPM. VORSICHT: Bitte kontaktieren Sie Veeco bevor Sie das Dimension 3100 SPM System transportieren.
Page 42
Safety Safety Precautions WARNING: The Dimension 3100 SPM contains a diode laser with an output of less than 1.0mW at 670nm. AVERTISSEMENT:Le microscope “Dimension 3100 SPM” est équipé d’une diode laser dont la puissance de sortie est inférieure à 1mW à 670nm.
Safety Safety Precautions WARNING: The Dimension 3100 SPM uses a halogen lamp to illuminate samples. Exposure to non-ionizing radiation from this lamp is well within the current exposure guidelines published by the American Conference of Governmental Industrial Hygienists (ACGIH). Typical IR exposure to the user from the sample illuminator is less than 3 mW/cm .
Page 44
The movements of all axes are slow, but are capable of exerting high forces. A hand caught in the stage of a Dimension 3100 SPM could be injured severely. ATTENTION: La platine des microscopes est automatisée dans les directions X, Y et Z, et est programmable.
Page 45
VORSICHT: Vermeiden Sie es, Flüssigkeiten auf dem Probentisch oder über elektronische Bauteile, insbesondere den Mikroskopkopf, zu verschütten. Wenn es notwendig ist, Flüssigkeiten zu verwenden, benutzen Sie dem Bedarf entsprechend nur geringe Mengen. Rev. D Dimension 3100 Manual...
Sperrvorrichtungen zu umgehen. Ergonomics The Dimension 3100 SPM design promotes compatibility in the integration of user personnel and equipment within a semiconductor manufacturing environment. Specifically, the ergonomics of the Dimension 3100 SPM design prevent personal injury, equipment damage, and minimizes procedural errors.
Safety Environmental Conditions Environmental Conditions Environmental conditions that may affect the performance of the Dimension 3100 SPM are vibration and noise. See Chapter 3 for detailed condition specifications. Equipment Safety Applications 2.5.1 Dimension 3100 SPM Facility Requirements Figure 2.5a Dimension 3100 SPM Footprint 14.2"...
Verify that clean dry air is available at 60-100 psi to the vibration isolation table. Verify that vacuum is available at ≥24" Hg to the system (IS3K-2 acoustic hood only). A vacuum _______ 5. pump is integrated into the Dimension 3100 controller for use with enclosures other than the IS3K-2. Module Installation _______ 1.
Page 49
______ Install the control station by completing the following: _______ Set the table to be used as the control station next to the Dimension 3100 SPM. _______ Place the input and display devices on the user console (monitors [2], mouse, keyboard and trackball).
5. Turn on the NanoScope controller using the power switch located on the rear of the NanoScope controller. 6. Turn on the Dimension 3100 controller using the power switch on the rear of the Dimension 3100 controller. Dimension 3100 Manual...
5. Turn on the monitors (2) using the push-button switches located on the front of the monitors. 6. Turn on the NanoScope controller using the power switch located on the rear of the NanoScope controller. 7. Turn on the Dimension 3100 controller using the power switch on the rear of the Dimension 3100 controller. Rev. D...
1. Press to log into Windows NT (see Figure 2.8a). CTRL DELETE Note: The screen begins with the panel split across both screens. 2. Drag the panel to one screen for use (see Figure 2.8a). Dimension 3100 Manual Rev. D...
2. Password: (Leave blank) Note: If you are unable to log on, verify with the process engineer that the password has not been changed. The system is shipped from the factory with a blank password. Rev. D Dimension 3100 Manual...
This allows the stage controller and computer to locate home positions and reset the coordinate system of the stage. If the Focus Surface and Locate Tip icons grayed out, the stage needs to be initialized. Dimension 3100 Manual Rev. D...
Begin stage initialization if necessary. Hazard Labels The Dimension 3100 SPM hazard labeling system identifies possible hazard areas. Caution must be taken according to the label warnings when working with associated areas. The following labels appear on the Dimension 3100 SPM: Rev.
Figure 2.9c) indicates that the area to which the label is affixed is affected by a laser. The Noninterlocked Protective Housing Label is affixed to the manual access door. Figure 2.9c Noninterlocked Protective Housing Label Dimension 3100 Manual Rev. D...
Chapter 3 Facilities Requirements This chapter details facility site requirements, safety requirements, and configuration options for the Dimension 3100 Scanning Probe Microscope. Specifically, this chapter details environmental requirements and equipment facilities drawings. Compliance with the following requirements and specifications is essential before beginning installation.
3.1a). The VT-103-3K is comprised of an air table, on which the Dimension 3100 microscope rests (not shown) and an acoustic hood which can be raised to access the instrument, and lowered to seal the instrument during operation. The ELCON is a compact control station console used to house the computer and control electronics underneath, and the monitors, mouse and keyboard on top.
This configuration consists of two basic elements: the VT-102 which is an air table for vibration isolation, and a typical “table top” version of the control station (computer, control electronics, and accessories). The VT-102 is a compact air table on which the Dimension 3100 microscope rests and does not include an acoustic hood.
The keyboard and mouse (not shown) sit on the small ledge at the front of the instrument. Finally, the Dimension 3100 microscope (not shown) rests on top of the vibration isolation table inside the IS3K-2 acoustic hood. The entire acoustic hood can raise and lower during installation.
475 mm Air and 13.5" Vacuum 343 mm 4005 Top View Front View Side View Weight: 150 lbs. for Dimension 3100 SPM assembly only. Note: Vibration and acoustic isolation is strongly recommended (see Section 3.5). Rev. D Dimension 3100 Manual...
An appropriate seat should be used which brings the user’s knees to within 3-4 inches (76-102mm) of the bottom of the keyboard tray. Figure 3.3a IS3K-2 - Front View 4009 62.0" 1575 mm 4" 101.6 mm 4.25" 108 mm External Vacuum Supply Module Dimension 3100 Manual Rev. D...
Open 127 mm 42.0" Closed 106 mm 4010 30.0" Electical 765 mm and Vacuum Hook-up Figure 3.3c IS3K-2 - Top View 30.5" 775 mm 42.0" 1066 mm 9.0" 228.6 mm External Vacuum Supply 4011 Module Rev. D Dimension 3100 Manual...
3.3.2 VT-103-3K Dimensions, Utilities and Clearance An isolation hood/table is required for acoustic and vibration isolation of the Dimension 3100. The table must be moved to its final location before Veeco personnel can install and train on the SPM. Figure 3.3f VT-103-3K - Front View 4006 50.0"...
The vibration isolation table may be supplied in lieu of the VT-103-3K or IS3K-2 for selected applications that do not require acoustic isolation for the desired performance level.The table must be moved to its final location before Veeco personnel can install and train on the SPM. Figure 3.3i VT-102 Vibration Isolation Table 24.0"...
60" min. 4004 1524 mm Top View Front View Weight: 170 lbs. nominal total; 140 typically on table top. Note: Dimensions shown for computer and controller equipment are approximate and subject to change without notice. Rev. D Dimension 3100 Manual...
914 mm Power: 1,800W; MOUSE 1 phase; 100V, KEYBOARD 120V or 240V TRACK BALL Min. Door width 37.0" required: 36" 940 mm Top View DISPLAY CONTROL MONITOR MONITOR 48.7" 1,237 mm 31.7" 805 mm Front View Dimension 3100 Manual Rev. D...
Facilities Requirements Summary If the Dimension 3100 system is used in a clean room with raised floors a pedestal must be provided to support leveling feet on the IS3K-2 or VT-103-3K. House vacuum is optional in all cases. If used, connections should accept a 1/8"...
BBN Criterion C - 1000X Microscopes. Aligners, Steppers to 1.5µm Geometries BBN Criterion D - Steppers, E-Beams to 0.3µm Geometries, CD Inspection Equipment. Most SEMs to 50,000X BBN Criterion E - Anticipated Adequate for Future Fabrication and Test Equipment for Low Submicron Geometries Dimension 3100 Manual Rev. D...
Facilities Requirements General Facilities Guidelines General Facilities Guidelines The following list contains general facilities recommendations for the Dimension 3100 system: • For the VT-103-3K and VT-102 vibration isolation systems, you must provide either clean dry air or nitrogen with a pressure capacity of 80psi in order to float the air tables.
Chapter 4 Installation This chapter details the installation procedure for the Dimension 3100 Scanning Probe Microscope (SPM) system from receiving to full installation. Specifically, the following topics are discussed in this chapter: • Shipping and Receiving: Section 4.1 • Equipment Requirements: Section 4.1.1...
Installation Shipping and Receiving Shipping and Receiving 4.1.1 Equipment Requirements The following equipment is necessary for successful installation of the Dimension 3100 SPM system. Verify the following is on-hand before beginning installation: Equipment Received • Dimension 3100 Manual • Dimension 3100 Controller •...
Page 75
Installation Shipping and Receiving Cables Received • Controller-to-Dimension 3100 Cable, 37-pin D • Dimension 3100 DC Power Cable • Fiber Optic Cable • Frame Grabber Video Cable (BNC to BNC) • Power Cords (3) • Serial Cable, 9-pin D, 12' •...
Uncrating the System Uncrating the System 4.2.1 Uncrate the Dimension 3100 SPM System 1. Using scissors, cut and remove the plastic shipping band encircling the Dimension 3100 shipping crate. 2. Lift the cardboard shipping box off of the Dimension 3100 microscope.
1. Place the vibration isolation table, or the microscope platform, at the desired location. 2. Place the vibration isolation pad on the isolation table. 3. Place the Dimension 3100 microscope in the operating location on the vibration isolation table or other platform.
Installation Installing the Dimension 3100 System 8. Align the two sets of three mounting holes on the long edges of the flexure with the three mounting holes of the chuck base and the three mounting holes on the X-Y stage. Align one...
Installing the Dimension 3100 System 4.3.2 Install the Control Station 1. Set up a table to be used as the control station next to the Dimension 3100 microscope. Note: Keep the control station in close proximity to the Dimension 3100 unit without touching the vibration isolation table.
Verify that the machine is powered-down and locked-out before attempting to make any connections. 4.4.1 Dimension 3100 Control Station Connections Connect the Display and Input Devices 1. Connect the monitor power cords (2) to the power strip, but do not initiate power.
Installation Connecting the Dimension 3100 System Figure 4.4a Computer (rear view) Computer USB port Video 2 Video 1 COM 4 COM 3 Rev. D Dimension 3100 Manual...
Page 82
Installation Connecting the Dimension 3100 System Note: The computer ships with the network board disabled to avoid error messages for computers not used on a network. 1. To use the computer on a network, select My Computer > Control > Panel > System >...
Installation Connecting the Dimension 3100 System Connect the NanoScope Controller Note: The NanoScope III and IIIa are addressed in this manual. For information regarding a NanoScope IV, please refer to your NanoScope IV Controller Manual (004-115-000). Connect extensions to the NanoScope III (IIIa) Controller. See Table 4.4b,...
Table 4.4c Figure 4.4d list of extensions, connection information and connection location. Note: The Dimension 3100 cables are bundled through the cable clamp provided if the system includes a vibration isolation table. (See section titled Route the VT103 Air Table Cabling at the end of the chapter).
Status 1 2 3 4 5 6 7 8 9 3908 2. To set illumination, toggle the manual control switch on the front of the Dimension 3100 controller to ON for manual control or OFF for computer control (see Figure 4.4e).
Dimension 3100 Microscope Electronics Box Connections 1. Connect the external microscope components to the Electronics Box. See Table 4.4d Figure 4.4f for list of extensions, connection information and connection location. Table 4.4d Dimension 3100 Microscope Electronics Box Connections Part Cable Function Number Vacuum Hose...
Remaining Dimension 3100 Cables 3907 1. Route all cables connected to the Dimension 3100 microscope through the slot located in the far right region of the VT103 vibration isolation table. 2. Position the Dimension 3100 microscope in the desired location.
3. Turn on the NanoScope IIIa controller using the power switch located on the rear of the NanoScope IIIa controller. 4. Turn on the Dimension 3100 controller, using the power switch on the rear of the Dimension 3100 controller. Dimension 3100 Manual...
Chapter 5 Stage System The Dimension 3100 Scanning Probe Microscope (SPM) features a large sample stage capable of positioning large samples such as silicon wafers and computer hard drive media, as well as small samples. The X-Y stage consists of a pair of stacked, perpendicular slides and uses an open loop (unencoded) architecture with stepper motors to drive the stage to user-specified coordinates.
5.1.2 Magnetic Pucks Samples may also be secured to the stage using magnetic pucks. This system allows for easy mounting and removal of small samples. For detailed mounting instructions, see Chapter Dimension 3100 Manual Rev. D...
Mounting of Samples 5.1.3 Axis Orientation—Motorized X-Y Stages When viewing the Dimension 3100 SPM from the front, stage movements are defined over two axes of motion: X (left-right), and Y (front-back). For X-axis movements, lesser (decreasing) coordinates are located to the left. For Y-axis movements, positive (increasing) coordinates are...
To load or unload a sample from the stage using the Load New Sample function, complete the following: 1. Select Load New Sample from the Stage pop-down menu. The screen displays a dialog box (see Figure 5.2a). Figure 5.2a Stage Load/Unload Prompt Stage Load/Unload? Cancel Dimension 3100 Manual Rev. D...
Moving to tip position Note: As the objective moves positions, the tip should begin to come into focus. When the focus position is attained, the screen displays trackball instructions for achieving a focus (see Figure 5.2c). Rev. D Dimension 3100 Manual...
When the surface is already partially in focus (or close to it), use the Autofocus feature. To use the Focus Surface feature complete the following: 1. Select Focus Surface from the Stage menu. The screen displays trackball instructions for achieving a focus (see Figure 5.2d). Dimension 3100 Manual Rev. D...
7. For samples which are difficult to bring into focus, move to an edge of the sample, which is easy to find in the optical image, and bring the top of the edge into focus. 8. Move the sample back to the desired X-Y position. 9. Verify that the surface remains in focus. Rev. D Dimension 3100 Manual...
3. Enter the X and Y coordinates of the desired move to position, then click Move. 4. The stage moves to the new position. Should it become necessary to halt movement, click on the Pause button (see Figure 5.2f). Figure 5.2f Abort Motion Prompt Stage Press <Pause> to abort motion Dimension 3100 Manual Rev. D...
1. Select the Focus Surface function under the Stage menu to focus on the sample and move the sample to the desired origin. 2. Select the Set Reference option under the Stage menu. The screen offers five options to the operator (see Figure 5.2g). Rev. D Dimension 3100 Manual...
(see Figure 5.2h). The left image depicts two closely located points to define the X-axis. The right image depicts a more accurate reference line defined by two, widely spaced points. Figure 5.2h Defining the X-Axis Dimension 3100 Manual Rev. D...
The name of the move program currently loaded is displayed in the Program name panel (in the above example, the program is called “move”). Moves are first programmed using the Teach function. Later they may be executed using the Run function. Rev. D Dimension 3100 Manual...
6. Use the controls to position the stage and sample to the desired location(s) to be programmed into the computer. Note: These controls are identical to the Focus Surface controls explained in Chapter 5 of this manual. Dimension 3100 Manual Rev. D...
4. Select the Remove Step option. Note: When individual program steps are removed, all subsequent steps are “moved up” by one count. 5. Click Save to save the edited version of program. Rev. D Dimension 3100 Manual...
Page 102
100 frames at 256 samples each. To run a series of programmed moves, complete the following: 1. Select Stage > Programmed Move. 2. Select the Program name field, and enter the name of the program to be run. Dimension 3100 Manual Rev. D...
To initialize the stage, complete the following: 1. Before beginning initialization, verify that the stage is clear of loose items and debris. 2. Select Stage > Initialize. The dialog box offers two options (see Figure 5.2p). Rev. D Dimension 3100 Manual...
Optics should move to end of travel 6. Next, the camera zoom optics assembly zooms. When the camera zooms in to the limit of its travel, click Ok to continue (see Figure 5.2t). Figure 5.2t Stage Zoom Prompt Stage Zoom in Dimension 3100 Manual Rev. D...
Stage Zoom out 5.2.8 SPM Parameters The SPM parameters menu lists the most important Z-axis parameter values for loading/unloading the stage and engaging samples. These values are explained in more detail in Chapter Rev. D Dimension 3100 Manual 85/(86 Blank)
Chapter 6 Cantilever Preparation The Dimension 3100 Scanning Probe Microscope comes furnished with etched silicon cantilever substrates for TappingMode AFM and silicon nitride cantilevers for Contact AFM modes. The cantilever probes should be inspected under the microscope when used for the first time to gain a better understanding of how the probes and substrates are connected and separated.
The cantilevers are stored tip-side-up, and the silicon is very brittle. The cantilever will break off of the substrate if contacted. 6.1.1 Wafer Tool Kit A wafer tool kit for working with silicon cantilever substrates is included with the Dimension 3100 SPM system. The kit contains the following: •...
It may be helpful to tip the substrate to one side to help grasp it in the tweezers. CAUTION: Silicon is extremely brittle.Be very careful to avoid any contact with the probe lever because it will immediately snap off. Rev. D Dimension 3100 Manual...
From the tip side, the cross-section of the tip near the lever is approximated by an inverted ‘kite’ shape. All of these subtleties arise from the etching process used to make the tip, which employs caustic solutions to perform wet anisotropic etching of the silicon. Dimension 3100 Manual Rev. D...
This method does not work well in small openings of less than 5 microns where, depending on the depth of the step, other tip edges may contact other faces of the small opening. Wall angle measurements are best measured in open areas for these reasons. Rev. D Dimension 3100 Manual...
70 -> 80 deg. in the image. Figure 6.1e depicts the resultant effect of the angled back ridge on the step angle measurement for a deeper trench depth. This is tip and topography dependent. Dimension 3100 Manual Rev. D...
4. Apply downward pressure with the tweezers until the strip breaks free from the silicon ring (see Figure 6.2a). Figure 6.2a Silicon Nitride Cantilevers in a Wafer Press here to break out strip Position glass slide underneath for support Rev. D Dimension 3100 Manual...
Page 114
7. While holding down the next substrate on the strip with the wooden end of a cotton swab, grip the overhanging piece with a pair of wide tweezers and rotate downward until the cantilever substrate breaks off. 8. Repeat this process until as many cantilever substrates as required are removed. Dimension 3100 Manual Rev. D...
For most applications use the 200µm cantilever with the wider legs. For atomic scale images, the 100µm triangular cantilever with the wider legs yields good results. Rev. D Dimension 3100 Manual...
(towards the cantilever holder). The steepest measurable angle from side to side (parallel to the edge of the probe's substrate) is approximately 55°. Both of these figures assume that the measurement does not have interference from other edges. Dimension 3100 Manual Rev. D...
The result is a noticeably sharper point at the end of the pyramid. Regrettably, along with the increased sharpness of the tip comes a slight increase in double tip effect experienced with the oxide sharpened process. Rev. D Dimension 3100 Manual 97/(96 Blank)
Chapter 7 Head, Probe, & Sample Preparation This chapter includes information regarding the Dimension 3100 Scanning Probe Microscope (SPM) setup and operation procedures for Contact Mode and TappingMode. Specifically, this chapter details removal and installation of the microscope head, mounting the cantilever, changing the tip, loading and positioning samples, focusing the optics, and general information regarding engaging and withdrawing the tip.
If during the engage sequence the sum signal changes to a value outside the interlocked limits, the downward motion stops, preventing damage to scanners had the Z-stage continued downward. Dimension 3100 Manual Rev. D...
Proben auf dem Probenteller liegen. Vermeiden Sie unter allen Umständen, in das reflektierte Laserlicht zu schauen. Alle Bediener des Mikroskops sollten größte Vorsicht walten lassen um zu vermeiden, in den von der Probenoberfläche reflektierten Laserstrahl zu schauen. Rev. D Dimension 3100 Manual...
Figure 7.2b). The cantilever holder stand is a 2.5-inch diameter black anodized aluminum cylinder that ships with the Dimension 3100 SPM for easy installation of the substrate onto the cantilever holder. The cantilever holder stand has three stations for different tip holders: 1) standard AFM, 2) fluid imaging AFM and 3) STM.
Single crystal silicon substrates are installed face-up so the tip points away from the AFM cantilever holder. This ensures that the cantilever and tip face toward the sample once the cantilever holder is mounted on the head. Dimension 3100 Manual Rev. D...
Dimension head. There is a diagram printed on top of the Dimension head illustrating which direction the laser moves when turning the laser control knobs clockwise. Dimension 3100 Manual Rev. D...
You can see the laser spot by removing the SPM head from the dovetail (without unplugging the cable between the SPM head and the Dimension base) and holding the SPM head over the granite or piece of paper. Rev. D Dimension 3100 Manual...
SPM head and the Dimension base. Dropping the head will likely result in damage and Veeco factory repair. 2. Verify the laser beam is visible on the surface below. If it is not, turn the rear-right laser control knob counter-clockwise until the laser spot appears on the surface below.
SPM head and the Dimension base. Dropping the head will likely result in damage and Veeco factory repair. 2. Verify the laser beam is visible on the surface below. If it is not, turn the rear-right laser control knob counter-clockwise until the laser spot appears on the surface below.
8. Move the laser onto the tip-end of the cantilever by reversing the direction of the rear-right laser control knob clockwise until the spot disappears from the surface below. 9. The laser is now on one of the two cantilevers. To determine which cantilever the laser is on: Dimension 3100 Manual Rev. D...
• Contact Mode AFM: 4 - 6V • TappingMode: 1.5 - 2.5V Note: If the laser sum signal is low, either the laser is not aligned or the photodetector knob needs adjusting (described in Step Rev. D Dimension 3100 Manual...
first center the laser on the screen on the head using the photodetector adjustment knobs, then use the detector schematic on the image monitor to finish positioning the laser. Dimension 3100 Manual Rev. D...
4. Focus on the cantilever using the trackball. You can zoom in using the Zoom button on the trackball. 5. Quit the Locate Tip sequence by pressing on the keyboard or clicking at the ENTER bottom of the Locate Tip dialog box. Rev. D Dimension 3100 Manual...
Place the calibration sample or other small sample on one of the 1.5cm diameter metal disks used for sample mounting. The Dimension 3100 SPM ships with several steel sample disks that you may attach to a magnetic sample holder located in the Dimension Chuck Spare Parts Kit. Also provided with the instrument are red and white colored double-sided adhesive patches, or “sticky tabs,”...
CAUTION: Since the command Focus Surface moves the scanner vertically, be careful when making this adjustment to ensure that the tip does not hit the sample surface. Rev. D Dimension 3100 Manual...
7.2.15 Set Initial Scan Parameters 1. In the Scan Controls panel, set the following: • Initial Scan Size is set to 1µm (or desired scan size). Veeco recommends you always initially engage with small scan sizes. • X and Y Offsets are set to 0.
1. To access the stage parameters panel, select SPM Parameters from the Stage pop-down menu. 2. For initial set-up, use the following values. When done, click to close the SPM Parameters window. Rev. D Dimension 3100 Manual...
This parameter defines how high the tip moves above the sample when Stage > Load New Sample is executed. After the Z-stage is raised, the sample stage moves as far forward in the Y direction (towards you) as possible. Dimension 3100 Manual Rev. D...
Chapter 8 Contact AFM This chapter covers procedures for operating the Dimension 3100 Scanning Probe Microscope (SPM) in Contact Mode AFM. It is assumed that the operator has previously prepared a Contact Mode tip and aligned the laser per instructions provided in Chapter 7 of this manual.
1. Align the laser using the laser control knobs. 2. Verify the laser beam is positioned on the back of the cantilever, with a spot visible in the Dimension head filter screen and a sum signal of 4-6V. Dimension 3100 Manual Rev. D...
3. Use the trackball with the bottom left button depressed to focus on the tip end of the cantilever. 8.1.7 Focus Surface 1. Select Stage > Focus Surface or click on the Focus Surface icon. 2. Focus on the sample surface using the trackball with the bottom-left button depressed. Rev. D Dimension 3100 Manual...
Without “X” Parameter will not display 3. Click the “minus box” in the upper left corner of the panel, and click Show all items. The panel will once again appear in normal Realtime mode. Dimension 3100 Manual Rev. D...
3. Set the Scan Angle to 0. 4. Set X and Y Offsets to 0. Figure 8.1c Suggested Scan Controls Settings Other Controls Panel Keep the Z limit at its maximum value. Figure 8.1d Suggested Other Controls Settings Rev. D Dimension 3100 Manual...
300nm initially. 3. Set Line direction to either Trace or Retrace. 4. On the Channel 2 panel, verify Data type is set to Off to disable the panel. Figure 8.1f Suggested Channel Controls Settings Dimension 3100 Manual Rev. D...
Veeco website (www.veeco.com) for more information. Silicon Nitride Cantilevers Silicon nitride cantilevers for the Dimension 3100 SPM are available in two process variations: standard and sharpened. Sharpened silicon nitride cantilevers (Model DNPS) are almost identical in appearance to the standard silicon nitride cantilevers, but have a slightly sharper end at the very tip.
These contaminants often form a capillary bridge between the tip and sample, generating large adhesive forces, also referred to as capillary forces (see Figure 8.2a). For more force curve information, see Chapter Figure 8.2a Force Curve Approach Jump to contact Contact Adhesion Pull-off Dimension 3100 Manual Rev. D...
To collect accurate topographical data, set the Data type parameter to Height. Topographical Deflection data is only reasonable on very smooth, flat samples. Topograhpical Deflection data is used infrequently and only for high resolution work. Rev. D Dimension 3100 Manual...
5. Set the LookAhead gain to 0.7 initially for samples with step-like features oriented perpendicular to the fast scan direction. Otherwise, it should be left at 0.00. Note: The LookAhead gain includes information from the previous scan line to determine the current gain setting. Dimension 3100 Manual Rev. D...
Highpass filter distorts the height information in the image. As a result, this filter must be Off when accurate height information is desired. The Highpass filter is typically used only for atomic images. Rev. D Dimension 3100 Manual...
Force Calibration Mode The Force Calibration command in the View > Force Mode > Calibration menu allows you to check the interaction between the cantilever and the sample surface. For detailed information regarding Force Calibration, see Chapter Dimension 3100 Manual Rev. D...
Chapter 9 TappingMode AFM This chapter details procedures for operating the Dimension 3100 SPM in TappingMode in air. For information regarding TappingMode in fluids, see Chapter 10. For information regarding loading a TappingMode tip and aligning the SPM laser see Chapter 7 of this manual.
“return signal,” deflects in a regular pattern over a photodiode array, generating a sinusoidal, electronic signal. Figure 9.1a Tapping Cantilever in Free Air Laser beam Return signal Cantilever Dimension 3100 Manual Rev. D...
The following is a general outline of basic operational procedures involved in TappingMode AFM. For more detailed instructions, refer to Chapter 7 of this manual. 9.2.1 Select Mode of Operation • Select Microscope > Profile. • Select TappingMode as the mode of operation. Rev. D Dimension 3100 Manual...
3. Use the trackball with the bottom left button depressed to focus on the tip end of the cantilever. Note: You may need to adjust the illumination and zoom to clearly see the probe. Dimension 3100 Manual Rev. D...
The initial Cantilever Tune panel appears with the Frequency Sweep (a plot of cantilever response as a function of applied oscillation frequency) on the display monitor. • Choose either the manual or automatic tuning method (see Automatic Tuning Manual Cantilever Tuning). Rev. D Dimension 3100 Manual...
7. Center the peak on the frequency sweep plot shown on the display monitor using the Zoom In and Offset commands after identifying the maximum amplitude peak with the lowest frequency in the frequency response plot. Dimension 3100 Manual Rev. D...
9.2c). Figure 9.2b Cantilever Tune Control Panels for Main Controls 10. Specify the RMS amplitude after tuning the cantilever to its resonant frequency. The desired operating amplitude depends on the sample and other scanning conditions. Rev. D Dimension 3100 Manual...
0.05 KHz/div 11. Click . The parameters set in the Cantilever Tune control panel appear in the Realtime control panel. 12. Click on to exit the Cantilever Tune command and leave the parameters CANCEL unchanged. Dimension 3100 Manual Rev. D...
Without “X” Parameter will not display 3. Click the “minus box” in the upper left corner of the panel, and click Show all items. The panel will once again appear in normal Realtime mode. Rev. D Dimension 3100 Manual...
3. Set the Scan Angle to 0. 4. Set X and Y Offsets to 0. Figure 9.2f Suggested Scan Controls Settings Other Controls Panel Verify the Z Limit is at the maximum value. Figure 9.2g Suggested Other Controls Settings Dimension 3100 Manual Rev. D...
5. After engaging, it may be necessary to recheck the Cantilever Tune parameters. Select View > Sweep > Cantilever Tune. In the Tip Offset box, enter 200nm and click OK. Adjust the Drive frequency if necessary. Rev. D Dimension 3100 Manual...
3. Use the Focus Surface command to move the SPM up if you desire more clearance between the tip and sample. CAUTION: Never withdraw samples without verifying that the tip has adequate clearance during the entire sample removal sequence. Dimension 3100 Manual Rev. D...
TappingMode AFM Advanced TappingMode AFM Operation Advanced TappingMode AFM Operation This section discusses the more subtle aspects involved in operating the Dimension 3100 in TappingMode. 9.4.1 Resonating Techniques Without a thorough understanding of principles associated with cantilever resonating techniques, you may generate distorted data. Understanding the Cantilever Tune process and the effects of real-time scan parameters is critical for effective operation of the microscope.
Integral gain and a twofold decrease in the Scan rate. The tip now tracks the surface when it descends into the pit as well as when it exits. The Trace and Retrace lines now coincide closely. Dimension 3100 Manual Rev. D...
Figure 9.4d Suggested Range of Drive Frequencies Drive Frequency should be within this range. Cantilever Response 12.54 nm/div Setpoint Center Frequency - 338.8723 KHz 5.0 KHz/div Rev. D Dimension 3100 Manual...
Scan rate as the Scan size is increased. Use Scan rates of 0.5-1.0 Hz for large scans on samples with tall features. High scan rates help reduce drift, but only use them on flat samples with small scan sizes. Dimension 3100 Manual Rev. D...
The drive frequency can be adjusted to compensate for such a shift by using the Offset command located above the Cantilever Tune plot. The phrase response curve can be adjusted as well. Rev. D Dimension 3100 Manual...
Contact Veeco for more information. This chapter details basic operation of the Dimension 3100 SPM in fluid, using both Contact Mode and TappingMode. Methods of preparing the sample for fluid operation are also included.
fluid. In addition, minor adjustments must be made to correct for refractive effects as the laser beam transits air-fluid boundaries. This chapter assumes familiarity with standard operation of the Dimension 3100 AFM in air. If you are not familiar with air operation of the AFM in Contact Mode, please follow the procedures...
find which cantilevers work best for their sample. The application note “Choosing AFM Probes for Biological Applications” discusses the appropriate probes for different fluid imaging applications. Contact Veeco for a copy. Note: For additional information on selecting a cantilever, please refer to the Veeco website (www.veeco.com). Removing Organic Contamination CAUTION: When cleaning the cantilever holder, take care to avoid scratching the glass surfaces in the center of the cantilever holder.
Note: For a list of articles describing biological applications of AFM, including sample preparation techniques, contact Veeco. CAUTION: Do not attempt to operate the standard air tip holder in a fluid environment.
1. Place the sample in a plastic petri dish. Note: If you wish to construct a custom sample holder, please contact Veeco for more information. 2. Secure the sample to the bottom of the petri dish lid or other sample holder with epoxy or other non water-soluble adhesive.
Throughout all procedures outlined in the following sections, you will work with fluids on and around the Dimension 3100 SPM. When handling fluids, keep a quantity of filter paper and/or paper towels nearby for wicking away any spilled fluid. The Dimension head is designed to be immersed in no more than 3mm of fluid when used with a protective skirt.
Page 176
En cas de fuite, nettoyer et sécher immédiatement les surfaces touchées. VORSICHT: Vermeiden Sie bitte, korrosive Flüssigkeiten auf freiliegende Oberflächen zu verspritzen; andernfalls wären Beschädigungen die Folge! Falls Sie Flüssigkeit verspritzt haben, säubern und trocknen Sie alle betroffenen Flächen sorgfältig. Dimension 3100 Manual Rev. D...
1. To mount a probe in the holder, turn the holder over so that the leaf spring and four sockets are facing down and plug the holder onto the appropriate docking station of the cantilever installation fixture. 2. Grip the tip holder by the edges and gently push down on the holder. Rev. D Dimension 3100 Manual...
6. Set the head back into the dovetail and lock into place by releasing the knurled head clamp screw, located at the upper-right of the Z-stage, until the thread is just loosened. Dimension 3100 Manual Rev. D...
Figure 10.5a Tip Holder Installed With Protective Skirt Maximum Fluid Depth protective skirt 7 mm sample fluid tip holder cantilever substrate 10.5.6 Align Laser Use the technique for aligning the laser onto the cantilever discussed in Chapter 7 of this manual. Rev. D Dimension 3100 Manual...
fluid. Note: Immersed samples may tend to dry out during imaging due to evaporation; therefore, keep a quantity of fluid nearby to periodically replenish lost fluid. Dimension 3100 Manual Rev. D...
Page 181
fluid and positioned one millimeter above the sample surface. All that remains is to tune the cantilever. Verify that the Microscope is Dry Verify that all Dimension 3100 surfaces are free of spilled fluid. Wick away moisture and droplets with filter paper. Rev. D...
Turn the detector mirror adjustment screws to center the laser spot on the laser detector as described in Chapter Set several parameters: • For Contact Mode: Vertical Deflection signal = roughly -1V, and Setpoint = 0V. (These are typical starting parameters). • For TappingMode: Vertical Deflection signal = 0V. Dimension 3100 Manual Rev. D...
When focusing on the sample surface in air before adding fluid to the sample area, the following procedure is necessary: 1. Align laser 2. Locate tip 3. Focus surface 4. Add fluid 5. Readjust laser alignment Rev. D Dimension 3100 Manual...
Page 184
When focusing on the sample surface directly in fluid, the following procedure is necessary: a. Align laser (in air) b. Locate tip (in fluid) c. Focus surface (hit engage several times) d. Focus surface (beyond surface by another 300µm) e. Realign laser f. Realign laser in photodetector Dimension 3100 Manual Rev. D...
The system tolerates some deviation in the shape of the peak. Manual Cantilever Tune is the counterpart to the step used in standard (air) TappingMode to find the resonant frequency of the cantilever. In liquid, however, the cantilever resonance is largely Rev. D Dimension 3100 Manual...
Page 186
3. In the Cantilever Tune panel, set the Drive frequency parameter to 10kHz and the Sweep width parameter to 20kHz. 4. Set the Drive amplitude to 200mV. 5. Zero the Amplitude setpoint. 6. Set the Amplitude limit to 2.5V. Dimension 3100 Manual Rev. D...
Operate at a frequency lower than the resonant frequency to avoid shifting the resonant frequency upon approach of the tip to the surface (see Figure 10.5e). Figure 10.5d Cantilever Tune Control Panels for Main Controls Rev. D Dimension 3100 Manual...
0.50 KHz/div 11. Click . The parameters set in the Cantilever Tune control panel appear in the Realtime control panel. 12. Click on to exit the Cantilever Tune command and leave the parameters CANCEL unchanged. Dimension 3100 Manual Rev. D...
Without “X” Parameter will not display 3. Click the “minus box” in the upper left corner of the panel, and click Show all items. The panel will once again appear in normal Realtime mode. Rev. D Dimension 3100 Manual...
3. For Contact Mode, in the Feedback Controls panel set the Deflection Setpoint to 0V. It is not necessary to enter a value for Amplitude Setpoint for TappingMode because this value is automatically determined during the engage process. Dimension 3100 Manual Rev. D...
Chapter Note: The cantilever will typically adhere to the sample surface much less in fluid; therefore, it is often possible to image at much smaller contact forces in liquid than in air. Rev. D Dimension 3100 Manual...
10.5.19 Clean Cell and Protective Skirt When you are finished scanning, remove and clean the fluid probe holder and protective skirt, following the procedure described in Section 10.5.1. In the Feedback Panel, set Z Modulation to Disabled. Dimension 3100 Manual Rev. D...
If the sample is short-lived and must be imaged quickly, mask the tip against contamination by bringing the tip into gentle contact with an uncontaminated substrate surface. Set the Dimension 3100 in Contact Mode by switching the AFM mode parameter on the Other Controls panel to Contact. Engage the substrate surface using a zero scan size.
If you suspect this problem, simply shift the imaging site to a location outboard of the tip and cantilever. You should find more individual samples there. Dimension 3100 Manual Rev. D...
0.02nm. The STM tip is held using a dedicated STM tip holder which plugs into the end of the Dimension head’s tube (see Figure 11.1a). Dimension 3100 Manual Rev. D...
3. Load the calibration standard (sample) onto the stage using a magnetic puck. Ensure that the standard is in electrical contact with the stage surface. This usually necessitates using an electrically conductive epoxy (silver-based) to affix the sample to the surface of the puck (e.g., Dynaloy 325). Rev. D Dimension 3100 Manual...
Page 198
Dimension-series microscopes due to their much larger stages and frames. Customers desiring to image atomic features should use Veeco’s dedicated STM microscope, which is equipped with a much smaller, more rigid frame.
Page 199
48nA for a very brief period of time. • If none of the above procedures improves the quality of the signal, replace the tip and try again. If that doesn't work, call Veeco for technical assistance. Rev. D Dimension 3100 Manual...
Special low-noise preamps have been developed for spectroscopic functions. The preamps which are included on scan heads designated as either AI or Veeco, have no voltage drop due to the tunneling current.
This constant height provides a reference from which to measure and record the fluctuations in the tunneling current. The current mode is most useful for imaging atoms with relatively small scan sizes. Dimension 3100 Manual Rev. D...
First, increase the Integral gain until oscillations start to appear, and then, back off a little. Next, adjust the Proportional and LookAhead gains. High frequency fuzz will appear on the signal when the Proportional gain is set too high. Setting the LookAhead gain too Rev. D Dimension 3100 Manual...
Page 204
The Filter parameter (if available) provides the option of filtering the tunneling current signal. A lowpass filter with a cut-off frequency of 25KHz can be applied to the analog tunneling signal in hardware. Typically, the filter is selected for atomic-scale images; otherwise, no filtering should be selected. Dimension 3100 Manual Rev. D...
Page 205
Although the microscope will accept any 0.010" diameter tip, tips made of platinum iridium (PtIr) and tungsten are used most often. PtIr tips are supplied by Veeco while tungsten tips can be electrochemically etched from tungsten wire by following instructions at the end of this chapter. In general, most of the discussion in this manual involving tips and noise reduction applies to both types of tips, but there are some applications which are tip specific.
Derating parameters in the Microscope > Calibrate > Scanner dialog box. As described in the following paragraphs, the Head Autocalibrate command in the Offline > Utility menu can be used to calculate the sensitivity and derating parameters of STM heads. Dimension 3100 Manual Rev. D...
The spectroscopic plots should aid in comparative studies between samples or between different regions on a sample, but they will not reveal the precise make-up of that sample. Rev. D Dimension 3100 Manual...
This section deals with problems related to the scan heads or the microscope. If a problem exists with a scan head, try a second one under the same conditions, if possible. Otherwise, the following list of symptoms and cures may be helpful: Dimension 3100 Manual Rev. D...
Samples with resistances 1 Megohm or greater will be difficult to image even with high bias voltages. Measuring the bulk conductivity of the sample with probes may not tell the whole story. Probes may easily penetrate oxide or contamination layers on the sample surface yielding reasonable Rev. D Dimension 3100 Manual...
Page 210
If the Realtime image appears to be flat, but captured images reveal detail, then try rotating the sample 90° in the microscope. The Realtime image leveling software tends to hide features that are parallel to the X-scan direction. Dimension 3100 Manual Rev. D...
Page 211
Setpoint Current—Raising the Setpoint current will effectively raise the gain of the feedback loop which can be quite helpful for large scans. It will also bring the tip closer to the surface but only by a small amount (i ≈ e -s ). Rev. D Dimension 3100 Manual...
Scanning Tunneling Microscopy (STM) Etching Tungsten Tips 11.6 Etching Tungsten Tips You can purchase tungsten tips from Veeco or make them yourself. This section describes the process of etching tungsten tips for use with the NanoScope. Materials Required • Variac auto transformer •...
Page 213
As your experience in etching grows, you will get better at throwing out the bad ones. 14. Repeat the etching procedure. Replace the etching solution when a fairly large amount of residue is present. Typically, you can etch 60 to 80 tips in a 40ml solution. Rev. D Dimension 3100 Manual...
Section 12.3.7 12.1 Introduction The Dimension 3100 SPM is capable of measuring frictional forces on the surfaces of samples using a special measurement known as lateral force microscopy (LFM). The name derives from the fact that cantilevers scanning laterally (perpendicular to their lengths) are torqued more as they transit high-friction sites;...
The trace and retrace LFM signals may not be the same magnitude. The tip will never be exactly perpendicular to the sample. In most cases, internal stress in the material of the cantilever will cause the cantilever to have a natural bias in some direction. Dimension 3100 Manual Rev. D...
Figure 12.3b below illustrates the movement of the tip as it encounters friction during a scan. Figure 12.3b Example of Frictional Forces on Tip During Scan (Front View) Trace Horizontal Deflection Retrace = Higher Frictional Areas Dimension 3100 Manual Rev. D...
Darker colors mean less friction. The sign of the color is inverted if Retrace is monitored (see Figure 12.3c). Figure 12.3c Friction Data High Friction Trace Low Friction Retrace High Friction Rev. D Dimension 3100 Manual...
The X-Y motion of an open loop scanner in trace and retrace does not overlap exactly. You can still use image subtraction, but some erroneous height data might be added to the LFM data in the final image. This is not an issue for closed loop scanners. Dimension 3100 Manual Rev. D...
Height artifacts will be the same direction in the scope display (see Figure 12.3e). Figure 12.3e Height Artifacts in LFM Data High Friction Trace Low Friction Retrace High Friction Rev. D Dimension 3100 Manual 201/(198 Blank)
The pull-off point of each magnet gives an additional index of its holding power. representation of this setup during a lowering cycle of one magnet is illustrated in Figure 13.1a. Dimension 3100 Manual Rev. D...
9 10 11 Height above steel plate (cm) 1. For the sake of simplicity, forces are represented here using a common scale in kilogram units; however, force is properly measured in Newtons (1 kg = 9.8 N). Rev. D Dimension 3100 Manual...
Ramp size parameter defines the total travel distance of the piezo. Therefore, you can obtain the maximum travel distance by setting the Z scan start to +220V, with the Ramp size set to 440V. Dimension 3100 Manual Rev. D...
AFM performance, and determination of the sensitivity of the cantilever deflection voltage in terms of voltage applied to the piezo. If used correctly, Force Calibration mode provides a variety of useful information. Rev. D Dimension 3100 Manual...
(Figure 13.2a). The Z scan rate parameter in the Main Controls panel defines the rate at which the piezo completes an extension-retraction cycle (and therefore the rate at which new curves are displayed). Rev. D Dimension 3100 Manual...
NanoScope display monitor. Cantilever deflection plots on the vertical axis of the graph. When the cantilever deflects downward, it plots on the graph’s downward vertical; when the cantilever deflects upward, it plots on the graph’s upward vertical. Dimension 3100 Manual Rev. D...
Page 231
It is possible to obtain quantitative measurements of sample elasticity. (See Radmacher, et al. 1994. Science, Vol. 265:1577-1579). Two imaging techniques measure and display elasticity at multiple points on a sample surface: force modulation and force volume imaging. Rev. D Dimension 3100 Manual...
View > Force Mode > Step enters force mode with the parameter start mode = step and the piezo fully retracted. • View > Force Mode > Volume enters Force Volume Mode. Items in the Force Calibration Control Window are discussed individually below. Dimension 3100 Manual Rev. D...
Reverse Velocity (Advanced Mode Only) This parameter defines the piezo speed to separate the tip and sample. X Offset, Y Offset These parameters define the X and Y positions where the force plot will be collected. Rev. D Dimension 3100 Manual...
Without X Rotate control, the tip moves in the direction of the cantilever during indentation. Normally, it is set to about 22.0 degrees. This parameter typically ranges between 12 and 25 degrees. Dimension 3100 Manual Rev. D...
The NanoScope software automatically calculates and enters the value from the graph after you use the mouse to draw a line to the graph. You must properly calibrate the Amplitude Sensitivity before amplitude data can be correctly displayed in nm. Rev. D Dimension 3100 Manual...
This parameter defines the amplitude of the voltage applied to the piezo system which drives the cantilever vibration. Changing the value of this parameter in Force Plot mode also changes the Drive amplitude parameter in the Image mode. Dimension 3100 Manual Rev. D...
An Absolute trigger permits the total force to be set using Setpoint values. Trigger Channel This parameter defines the channel to watch to determine when to reverse the ramp. Rev. D Dimension 3100 Manual...
Page 238
This parameter defines the size of Z step for Step or Motor Step modes. For Step, it the amount to move the piezo; for Motor Step it is the amount to move the motor. Auto Start When Enabled this parameter starts taking force curves upon entering Force Calibrate Mode. Dimension 3100 Manual Rev. D...
Run Single: Lowers and raises tip once by a distance equal to the Z scan size, then halts. • Stop: Halts all tip movement. Note: View additional menu subcommands by accessing the advanced Force Mode command. Rev. D Dimension 3100 Manual...
To minimize or calculate the contact force between the tip and sample, obtain a good force curve which shows the typical features displayed in the example curve in Figure 13.4b. However, the force curve rarely looks “typical” right after invoking Force Calibration mode. This section Dimension 3100 Manual Rev. D...
Page 241
Note: Sometimes the entire ramp will occur in the contact portion of the curve (see Figure 13.4a, between points 3 and 5). If this occurs, increase the Ramp size and/or decrease the Z scan start. Rev. D Dimension 3100 Manual...
Because the Z-axis leadscrew has some backlash, it may be necessary to rotate the leadscrew several turns by clicking on the Tip Up or Tip Down buttons before obtaining movement. Dimension 3100 Manual Rev. D...
5. The second click on the mouse causes the system to calculate the slope of the line segment and enter the -1/slope as the Sensitivity in the panel. 6. A click of the right mouse button will remove the line segment from the screen. Rev. D Dimension 3100 Manual...
Page 244
If a high initial contact force adversely affects the sample, engage the cantilever with a very small scan size. Then, minimize the force while the tip is confined to a small area of the sample where it experiences the relatively high initial engagement force. Once the force is minimized, increase the Dimension 3100 Manual Rev. D...
(point). The distance is then multiplied times the deflection sensitivity to obtain . In this example: d = (4.5 div) (1V/div) (70nm/v) = 315nm Therefore, the contact force is calculated as: F = (0.6N/m) (315nm) = 189nN Rev. D Dimension 3100 Manual...
Page 246
A simple alternative to calculating force is to follow these five steps: 1. Set sensitivity using the mouse. 2. Change units to metric. 3. Count vertical units from Setpoint to VCS min. 4. Multiply by Deflection Scale (adjacent to plot). 5. Multiply by the spring-constant, k. Dimension 3100 Manual Rev. D...
The examples in Figure 13.4f represent some of the general variations in force curves. For more information regarding force imaging, refer to Veeco’s application note Probing Nano-Scale Forces with the Atomic Force Microscope. Figure 13.4f Force Curve Examples...
SPM performance, calibrating the RMS amplitude or TM deflection, and configuring Lift Mode. For example, as the oscillating tip is brought closer to the surface, tip motion is dampened, which shows as an immediate drop in amplitude. When plotted, the graph resembles Figure 13.5a. Dimension 3100 Manual Rev. D...
(point 2). For TappingMode in air on a hard sample, each nanometer decrease in the cantilever position decreases the peak-to-peak vibration of the cantilever by two nanometers. Once the tip encounters Rev. D Dimension 3100 Manual...
You are now in Image mode. 4. To switch to Force Mode, click on the Realtime > View > Force Mode > Calibration or Advanced option. At least two panels should be visible: • Main Controls • Channel 1 Panel Dimension 3100 Manual Rev. D...
Page 251
Figure 13.5b TappingMode Force Plot Parameter Settings (Force Calibrate) 6. Set the Data type parameter to Amplitude under the Channel 1 panel. 7. Adjust the Z scan start parameter to obtain a satisfactory force plot using the left-right arrow keys. Rev. D Dimension 3100 Manual...
Sharper tips are more likely to plastically deform the sample, and are less sensitive to adhesive forces, than tips that are more blunt. Experiment to determine the tip that best meets your needs. Dimension 3100 Manual Rev. D...
This cantilever holder uses a piezoelectric bimorph to oscillate the cantilever against a sample surface. The force modulation cantilever holder is similar to the standard tapping cantilever holder; however, its piezo stack is much larger, allowing larger driving amplitudes. Rev. D Dimension 3100 Manual...
It may take experimentation to find a cantilever that matches the sample's requirements. For rubber and plastic samples Veeco recommends using 225µm long force modulation (Model # FESP) silicon cantilevers. For more delicate, samples, use 450µm long silicon cantilevers or silicon nitride cantilevers.
Note: Refer to the section on Tip Selection earlier in this chapter to help you choose an appropriate cantilever. Rev. D Dimension 3100 Manual...
Page 256
The display monitor plots the Frequency Sweep, showing cantilever oscillation amplitude versus frequency. To set the parameters controlling the Frequency Sweep plot, use the parameters in the Sweep Controls panel and the commands on the top menu bar of the display monitor. Dimension 3100 Manual Rev. D...
TappingMode operation. This is due to the cantilever not at resonance; therefore, its motion is mostly vertical. Vertical motion is not amplified by the beam deflection detection technique which is sensitive primarily to changes in cantilever angle. Rev. D Dimension 3100 Manual...
(see Figure 13.6e). n. Adjust the Drive amplitude so the maximum response amplitude is about 1V. Recenter the peak if necessary. Figure 13.6e Correctly Tuned Force Modulation Frequency Dimension 3100 Manual Rev. D...
Page 259
However, in this condition imaging may be more unstable, as the cantilever may pull off the surface unexpectedly. The setpoint value affects the force modulation contrast. Examine the contrast of the force modulation image. Rev. D Dimension 3100 Manual...
Page 260
Sometimes an oscillation that appears in the data will be due to “aliasing” as described in the next section. If you cannot adjust gains to eliminate unwanted oscillations without compromising the height image’s quality, see Section 13.6.5. Dimension 3100 Manual Rev. D...
To reduce the influence of frictional effects, use a smaller Setpoint or Drive amplitude. Or, if you are using “Negative LiftMode” (see Section 13.7), use a more positive Lift Height. Rev. D Dimension 3100 Manual...
Not all samples lend themselves to standard force modulation imaging. Even samples with excellent elasticity contrast may not show up in force modulation if their absolute elasticity is out of instrumental range. “Negative LiftMode” may prove useful for imaging otherwise difficult materials (see Section 13.7.3). Dimension 3100 Manual Rev. D...
AFM force modulation. This method is especially suited for softer materials, yielding higher resolution. Best results using negative LiftMode are obtained on relatively smooth samples (< 500nm vertical features); however, Veeco encourages experimenting with this technique on rougher surfaces as well. A general procedure for Force Modulation with “Negative LiftMode” is...
The contrast between light and dark reveals areas of high and low elasticity, with the dark area indicating harder material and the lighter areas indicating softer material. Dimension 3100 Manual Rev. D...
Z-positions and at thousands of XY positions during a single image scan, correlation of surface topography to interaction force, better quantization of the interaction force, and new methods of analysis. For detailed information regarding force volume imaging, contact Veeco for a copy of Support Note 240A, Force Volume.
Interleave scanning and LiftMode for obtaining magnetic and electric force data. For purposes of learning to use Interleave scanning, this section may be useful even to those users whose end application is not magnetic force microscopy. Dimension 3100 Manual Rev. D...
If there are any features taller than the lift height you entered, the system will lift the tip over those features rather than crashing the tip. In Retrace Lift the tip is on the surface during the main scan, and lifts during the retrace scan line. Rev. D Dimension 3100 Manual...
Interleave parameters to an “off” (grayed) condition. The values can be changed once engaged. If using Lift, set the Lift start height and Lift scan height. If using Interleave Mode, set the Gains and Setpoint. Dimension 3100 Manual Rev. D...
Lift scan for magnetic force imaging, the Drive frequency should be set to the side of the resonance; see Chapter 15. (However, certain parameters can be set independently for the interleave scan; see below.) Rev. D Dimension 3100 Manual...
Main and Interleave scanning. If using scan rates above a few hertz, it may be advantageous to remove or disable the filter. For more information on how to disable the filter, contact Veeco technical support.
Small amplitudes must be used to sense force gradients, such as Van der Waals forces, which occur over short distances (typically a few nm). As much of the cantilever travel as possible should be within the range of the force gradient. Rev. D Dimension 3100 Manual 253/(246 Blank)
2. Cantilever retraces surface topography on first retrace. 3. Cantilever ascends to Lift scan height. 4. Lifted cantilever profiles topography while responding to magnetic influences (second trace). 5. Lifted cantilever reprofiles topography while responding to magnetic influences (second retrace). Dimension 3100 Manual Rev. D...
By adding the Phase Extender or Quadrex Extender Electronics Modules to a NanoScope III or IIIa controller, or by using a NanoScope IV controller, the Dimension 3100 may be used for frequency modulation or phase detection with improved results (see Figure 15.1b...
(metal-evaporated video tape). Standard tape samples are provided with purchase of MFM probes, and can be obtained free of charge from Veeco. Other samples can also be used; however, you will not have the benefit of comparing your results with the images shown here. Obtaining a good image of the tape sample will familiarize you with Interleave and MFM techniques and provide a check that the system is correctly tuned to image magnetic samples.
(perpendicular to the sample surface). The MFM then senses force gradients due to the perpendicular component of the samples’s stray field. Tip magnetizers can be purchased from Veeco. 2. Mount a NanoProbe magnetic probe on the scanner or tip holder.
As the tip oscillates above the sample, a gradient in the magnetic force will shift the resonance frequency f (see Figure 15.2c). Tracking the variations in oscillation amplitude while in LiftMode yields an image of the magnetic force gradients. Dimension 3100 Manual Rev. D...
The lift height can be optimized later. 7. Set the remaining Interleave parameters (Setpoint, Drive Amplitude, Drive frequency, Gains) to the Main Controls values by setting the flags left of the Interleave Control column to Off. Rev. D Dimension 3100 Manual...
Page 282
Figure 15.2e Topographic (left) and Magnetic Force Gradient Image (right) Note: The MFM data displays in Channel 2; however, the parameter settings are different depending on whether Phase Detection or Amplitude Detection is being used Dimension 3100 Manual Rev. D...
Page 283
2. Switch the Channel 2 image Data type to Frequency. Try a Data scale (frequency shift) of approximately 10 Hz. 3. Select Other Controls, then adjust the frequency modulation gains: Rev. D Dimension 3100 Manual...
The ultimate lateral resolution of MFM is near 20nm. Resolution is affected by properties of the tip, including mechanical sharpness and magnetic structure. When in good condition, magnetically- coated tips routinely give 50nm resolution, and many achieve 30nm or better. Dimension 3100 Manual Rev. D...
Drive Amplitude is not increased to the extent that the tip strikes the surface on the low point of its swing. The signatures of tip-sample contact are white and black spots in the image, or, in extreme cases, noisy, high-contrast streaks across the whole Rev. D Dimension 3100 Manual...
Page 286
Drive Amplitude between main and Interleave scanning. This can interfere with very fast rates (> a few Hz). The filter can be disabled easily; contact Veeco technical support for more information. Dimension 3100 Manual...
20-25nm/V. The exact value depends on the position of the laser spot on the cantilever, the cantilever geometry, and the particular instrument. A 1V Setpoint typically corresponds to h ~14-18nm. Rev. D Dimension 3100 Manual...
A hardware change is also required; the main electronics backplane board at the rear of the microscope must be swapped out (this may be done by either the customer or a factory representative). For more information, contact Veeco. The phase box is equipped with a slider switch for switching internal electronics between Dimension-series and MultiMode SPM signals.
1. Extender-compatible microscope electronics are required to permit operation of the phase detection extender option. Standard electronics on these microscopes require hardware upgrades. Consult Veeco for details. 2. Turn off the power to the NanoScope controller whenever connecting or disconnecting the Extender.
Be careful not to move the beam off the cantilever or feedback may be lost. Note: Optical interference is essentially eliminated by using phase detection or frequency modulation. Dimension 3100 Manual Rev. D...
Section 16.3.1 • Verify Electric Field at Surface: Section 16.3.2 • Fine Tune Lift Scan Height: Section 16.3.3 • Fine Tune Interleave Drive Amplitude: Section 16.3.4 • Optimize Tune in Vicinity of Surface: Section 16.3.5 Rev. D Dimension 3100 Manual...
Electric Techniques Overview 16.1 Electric Techniques Overview There are two types of electric techniques used with the Dimension 3100: Electric Force Microscopy (EFM) and Surface Potential Detection. Electric techniques are similar to magnetic force microscopy (MFM) and share many of the same procedural techniques.
Quantitative voltage measurements can be made of the relative voltages within a single image. This method requires the Extender Electronics Module and version 3.1 or later of the NanoScope software. Dimension 3100 Manual Rev. D...
For other samples having rough surface topography or small voltage variations, this technique may be undesirable because topographic features appear in the LiftMode image. Rev. D Dimension 3100 Manual...
The backplane board used with Dimension Series SPMs is shown below in Figure 16.2c. There is a header supplied with jumpers at the center of the board. For non-EFM applications and Surface Potential operation, leave or return jumpers to their original positions. Dimension 3100 Manual Rev. D...
4. Reconfigure jumpers on the backplane header using the appropriate jumper configuration. 5. After correctly configuring the backplane jumpers, replace the cover on the electronics box. 6. Plug the power cable back into the microscope. 7. Turn on the NanoScope controller. Rev. D Dimension 3100 Manual...
Page 298
1. Select Di > Microscope Select > Edit > Advanced, and set Analog 2 to Atten switch. 2. Click to exit both dialog boxes. Note: Remember to set Allow in attenuation to Allow when finished. For all other configurations it should be left on Allow. Dimension 3100 Manual Rev. D...
Table 16.2a Extender Electronics Module Toggle Switch Settings Mode Tip or Sample Voltage GND/ Surface FM/Phase Surface Analog 2 Potential Potential EFM with Analog 2 biasing tip or sample EFM in all other configu- rations Standard Operation Rev. D Dimension 3100 Manual...
Figure 16.2e Normal Jumper Configuration Sample Chuck Ground/Bias Sample Chuck Analog 2 Gain Select Sample Unused STM Signal (from Dimension head) Auxiliary D (to NanoScope III controller) Ground Ground Indicates Jumpers Dimension 3100 Manual Rev. D...
Page 301
Figure 16.2f Jumper Configuration for Application of Analog 2 Voltage to Tip Sample Chuck Ground/Bias Sample Chuck Analog 2 Analog 2 Sample Gain Select Unused STM Signal (from Dimension head) Auxiliary D (to NanoScope III controller) Ground Ground Indicates Jumpers Rev. D Dimension 3100 Manual...
Figure 16.2g Jumper Configuration (Application of Analog 2 Voltage to Sample) Sample Chuck Ground/Bias Sample Chuck Analog 2 Sample Gain Select Analog 2 Unused STM Signal (from Dimension head) Auxiliary D (to NanoScope III controller) Ground Ground Indicates jumpers Dimension 3100 Manual Rev. D...
You may also use current-limited power supplies. Connect voltage leads to pins on the header using soldered, push-on connectors. Do not solder leads directly to the header pins, as the heat may cause damage and/or make jumpering the pins difficult. Rev. D Dimension 3100 Manual...
Page 304
Voltage leads should be connected to pins on the header using soldered, push-on connectors. Do not solder leads directly to the header pins, as the heat may cause damage and/or make jumpering the pins difficult. Dimension 3100 Manual Rev. D...
Page 305
Figure 16.2j Normal Jumper Configuration with Extender Electronics Module ChuckGround/Bias Chuck Analog 2 or Gnd/OSC + DC signal Unused Sample Unused STM Signal Auxiliary 2 Output Ground Ground Indicates Jumpers Rev. D Dimension 3100 Manual...
Figure 16.2k Jumper Configuration for Application of Voltage to Tip Chuck Ground/Bias Sample Chuck Analog 2 or Analog 2 Gnd/OSC + DC signal Sample Unused Unused STM Signal Auxiliary 2 Output Ground Ground Indicates Jumpers Dimension 3100 Manual Rev. D...
Figure 16.2l Jumper Configuration for Application of Voltage to Sample Chuck Ground/Bias Sample Chuck Analog 2 or Gnd/OSC + DC signal Sample Unused Analog 2 Unused STM Signal Auxiliary 2 Output Ground Ground Indicates Jumpers Rev. D Dimension 3100 Manual...
You can also use current-limited power supplies. Connect voltage leads to pins on the header using soldered, push-on connectors. Do not solder leads directly to the header pins as the heat may cause damage or make jumpering the pins difficult. Dimension 3100 Manual Rev. D...
Page 309
Connect voltage leads to pins on the header using soldered, push-on connectors. Do not solder leads directly to the header pins, as the heat may cause damage and make jumpering the pins difficult. Rev. D Dimension 3100 Manual...
5. Set up the AFM as usual for TappingMode operation. 6. Select View > Sweep > Cantilever Tune. 7. Follow the procedure below for the type of electric force imaging desired, Phase Detection, Frequency Modulation or Amplitude Detection (see sections 16.2.4 through 16.2.6). Dimension 3100 Manual Rev. D...
Resonance shifts also give rise to phase shifts ∆φ used to generate an image of the electric force gradients; see Figure 16.2p. Figure 16.2p Shift In Phase at Fixed Drive Frequency ∆F ∆φ Drive Frequency Rev. D Dimension 3100 Manual...
Page 312
Feedback Controls panel. For more details, refer to EFM Troubleshooting/Pointers: Section 16.3. 11. Optimize the Lift scan height. For high-resolution, make the Lift scan height as small as possible without crashing the tip into the surface. Dimension 3100 Manual Rev. D...
Figure 16.2q. Figure 16.2q Shift In Amplitude at Fixed Drive Frequency ∆F ∆ amplitude Drive Frequency 1. Set the Drive frequency to the left side of the cantilever resonance curve, as shown in Figure 16.2r. Rev. D Dimension 3100 Manual...
Note: Optical interference is essentially eliminated by using phase detection or Frequency Modulation, available only with the Extender Electronics Module. Dimension 3100 Manual Rev. D...
Adjust drive frequency to center of peak or just to the left. If phase data is to be collected during the main line, zero the phase at the drive frequency. Rev. D Dimension 3100 Manual...
On the second pass, the piezo that normally vibrates the cantilever is turned off. Instead, to measure the surface ωt potential, an oscillating voltage is applied directly to the cantilever tip. This creates an Dimension 3100 Manual Rev. D...
At this point the tip voltage will be the same as the unknown surface potential. The voltage applied to the cantilever tip V tip is recorded by the NanoScope III to construct a voltage map of the surface. Rev. D Dimension 3100 Manual...
(e.g., on integrated circuits connect electrical leads directly to pads). For normal operation, the sample chuck is held at ground. For Application Module-ready systems, the bias is normally applied to the chuck. Be certain to carefully any electrical connections from the sample chuck. Dimension 3100 Manual Rev. D...
Figure 16.5b Jumper Configuration for Application of Voltage to Tip Chuck Ground/Bias Sample Chuck Analog 2 or Analog 2 Gnd/OSC + DC signal Sample Unused Unused STM Signal Auxiliary 2 Output Ground Ground Indicates Jumpers Rev. D Dimension 3100 Manual...
Do not solder leads directly to the header pins. Heat may cause damage and/or make jumpering the pins difficult. The sample should be electrically connected directly to the chuck or a standard sample puck using conductive epoxy or silver paint as shown below: Conductive Epoxy or Paint Sample Sample Chuck Dimension 3100 Manual Rev. D...
3. Mount a metal-coated NanoProbe cantilever into the standard cantilever holder. MFM-style cantilevers (225µm long, with resonant frequencies around 70kHz) usually work well. It is also possible to deposit custom coatings on model FESP silicon TappingMode cantilevers. Rev. D Dimension 3100 Manual...
Page 322
Drive amplitude that has been selected. Also, when Potential is selected as the Data type for the Channel 2 image, a feedback circuit is enabled in the Extender box which adjusts the dc voltage on the tip to maintain the cantilever oscillation Dimension 3100 Manual Rev. D...
Page 323
(The lock-in amplifier produces a voltage that is proportional to the cantilever amplitude.) Qualitative surface potential images can be collected using this lock-in signal. Also, if the sample has a surface potential that exceeds ±10V (greater than the range of the Rev. D Dimension 3100 Manual...
Interleave mode to Disabled, then back to Enabled. Also try reducing any external voltage that is being applied to the sample to stabilize the feedback loop, then turn the voltage back up. Dimension 3100 Manual Rev. D...
Veeco employs a software-guided calibration procedure for all microscopes. The procedural details of how calibration is executed using NanoScope software are beyond the scope of this document and include proprietary methods exclusive to Veeco. A brief overview of the theory is available in Section 17.1.1.
Critical Height Measurements All Years Monthly Note: Small Scan Size Calibration: If using scan sizes of 5µm or smaller, Veeco recommends calibrating the scanner for small scan sizes. Contact Veeco for further instructions. 17.1.1 Theory Behind Calibration Scanners typically consist of a hollow tube made of piezoelectric material such as PZT (lead zirconium titanate).
By precisely determining points along the scanner’s sensitivity curve, then applying a rigorous mathematical model, full-range measuring capabilities can be achieved with better than 1 percent accuracy. Consider the sensitivity curve represented in Figure 17.1b. Rev. D Dimension 3100 Manual...
As scan size is decreased, the piezo exhibits more linearity. These effects are further complicated by X-Y-Z coupling effects (the tendency for one axis to affect movement in other axes). Dimension 3100 Manual Rev. D...
Movement Time Through rigorous quality control of its scanner piezos, Veeco has achieved excellent modeling of scanner characteristics. Two calibration points are typically used for fine-tuning: at 150 and 440 volts. A third point is assumed at 0 nm/volts. These three points yield a second-order sensitivity curve to ensure accurate measurements throughout a broad range of scanner movements.
CD shipped with every system. Individually purchased scanners are shipped with a head/scanner disk containing backup files or a hard copy of the scanner parameters. 2. In the event that files are not found, fax or call Veeco for scanner calibration records. Dimension 3100 Manual...
Number of samples Slow scan axis Enabled Z limit 440 V Other Controls Units Volts Channel 1 Data type Height Z range ~ 20 V a. Adjust the Z range parameter to obtain the best contrast. Rev. D Dimension 3100 Manual...
2. On the display monitor, select the Angle command and use the mouse to draw a cursor between the edges (or centers) of widely spaced pits (see Figure 17.3a). Figure 17.3a Non-Orthogonal and Corrected, Orthogonal Image Dimension 3100 Manual Rev. D...
5. Repeat correction of Orthogonality until the scanned image shows less than 0.5º of error. Note: After a major change to the orthogonality parameter, you may need to physically realign the calibration standard to the image frame. Rev. D Dimension 3100 Manual...
The beginning third of the scan is the standard for judging almost all of the linearity values. Ignore the set of features near the edges of the scan since these may be distorted slightly. 8. Move the zoom box to the end third of the scan. Dimension 3100 Manual Rev. D...
4. After an adjustment in Fast arg is made, Fast mag0 may need readjusting. Repeat Fast mag0 adjustment procedure until the rulings are evenly spaced across the Fast-axis. Figure 17.4b X Scan Linearization Center Compressed Center Expanded Rev. D Dimension 3100 Manual...
5. If the features of the end third are too large for the box, decrease the parameter. If the features are too small, increase the parameter. Note: Compare only parts of the current scan, not the previous scan. Figure 17.4c Y Scan Linearization Center Compressed Center Expanded Dimension 3100 Manual Rev. D...
The yellow retrace line will shift in the same direction as the arrow. 9. When done, click to close the Scanner Calibration window. 10. Set Slow scan axis back to Enabled. 11. Select View > Image Mode. Rev. D Dimension 3100 Manual...
6. Wait one complete frame with the new value before readjusting the Slow mag1 value. 7. Check the final result by capturing an image and checking it with the Offline > Modify > Zoom window. The scanner is now ready for calibration of the X and Y parameters. Dimension 3100 Manual Rev. D...
5. As each routine is executed, adjust the scan slightly to optimize the calibration image using the Capture Control dialog box displayed on the control monitor throughout the calibration routines (see Figure 17.5b). Rev. D Dimension 3100 Manual...
After the first four images with the diagonal stripe pattern are captured, you can leave the system unattended while the program continues to completion. Some of the following images appear stretched in one dimension; however, this is normal. Dimension 3100 Manual Rev. D...
2. Select the Offline > Utility > Autocalibration command. The control monitor will display the X-Y Piezo Calibration dialog box (see Figure 17.6a). Figure 17.6a X-Y Piezo Calibration Prompt 3. Verify that the file name prefix assigned to the captured files from the Capture Calibration routine is correct. Rev. D Dimension 3100 Manual...
10µm apart (e.g., between bottom edges, left sides, etc.). Note: Measure features without regard to how they appear in calibration images. Features may be represented with stretched, distorted, or angled appearances Dimension 3100 Manual Rev. D...
Page 345
1-2 percent accuracy over most of the scanner’s measuring range. To obtain still better accuracy, the SPM can be fine-tuned to obtain maximum measuring accuracy. This is accomplished through the use of calibration parameters discussed in Section 17.7. Rev. D Dimension 3100 Manual...
Known distance between features ---------------------------------------------------------------------------------------------------- - SPM-calculated distance between features 7. Select the Realtime > Microscope > Calibrate > Scanner option. The Scanner Calibration dialog box displays (see Figure 17.7b). Figure 17.7b Scanner Calibration Panel Rev. D Dimension 3100 Manual...
8. Multiply the quotient obtained earlier by the Y slow sens value shown on the Scanner Calibration panel. 9. Enter the new value to adjust the scanner’s slow axis to more closely match calculated distances with actual feature distances. 10. To save the new parameter value, click Dimension 3100 Manual Rev. D...
X fast derate or Y slow derate value you recorded in the above step; m is the measured distance; and, v is the Scan size in volts: ⋅ – --- - s d 440 v – – ------------------------------------------------------------- - 440 v – Rev. D Dimension 3100 Manual...
Repeat steps above for the following parameters: Y fast sens, X slow sens, Y fast der, and X slow der to ensure the scanner is calibrated properly along the X- and Y-axis for scanning at 90º. Dimension 3100 Manual Rev. D...
X- and Y-axes do during scanning. Furthermore, offsets affect the piezo over a period of minutes. The silicon calibration references distributed by Veeco have 200nm vertical features accurate to within ± 3 percent. The calibration reference is referred to throughout the examples provided in this section.
4. Go to the display screen and draw a stopband over the pit as shown in Figure 17.8b. Figure 17.8b Draw a Stopband 5. Click Execute to complete the flattening procedure. 6. Quit the dialog box. Dimension 3100 Manual Rev. D...
Height data within the drawn cursor box displays on the monitor, showing two, prominent peaks. These peaks correspond to two elevations on the surface: the bottom of the pit and the top surface. There should be a line cursor on each peak. Rev. D Dimension 3100 Manual...
6. In the Data result dialog box located under the image on the display screen, check the Peak to Peak value and record it for future reference. 7. Click Quit to exit the Depth dialog box. Figure 17.8f Z Calibration Depth Dialog Box Dimension 3100 Manual Rev. D...
Z sensitivity value. Note: The numerator value above (200nm) is for Veeco 10µm silicon reference. For other calibration references, set the numerator equal to the depth of features measured by Depth analysis. Ideally, calibration references should have features with heights comparable to those being imaged and measured on samples.
Page 356
Step 5 Step Note: The measured depth should read 200nm on a Veeco 10µm silicon calibration reference. 5. Select Realtime > Microscope > Calibrate > Z to display the Z Calibration panel 6. Click on the Extended offset der parameter.
Maintenance, Troubleshooting and Warranty Maintenance 18.1 Maintenance Generally, the Dimension 3100 system requires very little maintenance. A simple schedule of cleaning and calibration check is usually sufficient. Every three months: 1. Wipe down the major components of the instrument. 2. Clean the SPM stage.
Page 359
5. Lubricate the screw and the leading edge of the threaded insert with a small amount of silicon vacuum grease. 6. Thread the screw back into the scanner head. 7. Repeat the cleaning procedure with the other three adjustment screws. Rev. D Dimension 3100 Manual...
Because the Dimension 3100 is sensitive to vibration, it is often a convenient tool for checking the performance of the vibration isolation system. The following test can be performed under ambient conditions or while an external noise source is applied to the floor.
Dimension control box. Dimension 3100 systems ship with a spare light bulb mounted within the Dimension control box back panel. Order a new illuminator bulb from Veeco when the spare bulb burns out. Please note availability and shipping time may vary. Figure 18.1a Change the Illuminator Bulb...
Page 362
7. Plug in the light bulb cable. 8. Slide the bulb mount panel back and replace the holding screws. 9. Order another spare bulb from Veeco. CAUTION: When you receive the new spare bulb, remove the power from the...
120V 281-000-029 0.2A 5 x 20mm 250V 240V 281-000-028 0.125A 5 x 20mm 250V Fuse Replacement Procedure 1. Remove all power from the system. 2. Remove the lid of the appropriate control box (4 screws). Rev. D Dimension 3100 Manual...
6. Using a screwdriver, press the slot in the fuse holder and rotate the fuse a 1/4 turn clockwise until the fuse is secure. 7. Order another spare fuse from Veeco. Be sure to turn off all system power before placing the spare fuse back in the holder location.
Do not tighten the nut more than 1/8-1/4 turn. This is typically enough additional tension to fix the problem. 7. Verify that the zoom optics work correctly and that the clutch still slips at the end of the motors travel. Rev. D Dimension 3100 Manual...
Troubleshooting 18.2 Troubleshooting The Dimension 3100 system is very reliable and generally will not need adjustment once properly calibrated, nor is it likely to develop a problem due to use. In most cases any problems that occur can be linked to some external event.
Page 367
The message will also appear when switching between AFM and STM probes. The intent of this message is a safety check to warn you that you have moved the objective unusually long distance. Select OK if you’ve changed to a different style of probe. Rev. D Dimension 3100 Manual...
Cables – The wires inside the cable connector on the Dimension scanner can be broken from the 21-pin plug. Disconnecting or connecting the scanner by pulling on the cable can cause this. Dimension 3100 Manual Rev. D...
Veeco’s obligation under these warranties is limited to repairing or replacing at Veeco’s option defective non-expendable parts. Veeco’s obligation shall not extend to defects that do not impair service. No claim will be allowed for any defect unless Veeco has received notice of the defect within thirty days following its discovery by purchaser.
Page 370
To be eligible for the above warranties, purchaser must perform preventative maintenance in accordance with the schedule set forth in the manual provided. Veeco assumes no liability under the above warranties for equipment or system failures resulting from improper operation, improper preventative maintenance, abuse or modifications of the equipment or system from the original...
Page 371
(installation and service only) — Symbols software power-up 35 Chucks . 248 vacuum 115 Circuitry Dimension 3100 microscope 8 Dimension 3100 controller 7 Color contrast 180 Aliasing 241 Color offset 180 Amplitude 146 Color table 180 Atomic Force Microscope (AFM)
Page 372
240 thermal hazard 6 — operating procedure 235 vacuum pump 7 — principles of 235 Dimension 3100 Microscope Electronics Box 7 Frequency Modulation larger SPM mother board 8 with MFM 263 main SPM circuit board 8 Frequency Sweep 135 —...