______________________________________
Optional connector for fieldbus
Inlet connection
Depending on the ordering details, fittings and/or connectors on your instrument might differ from the
images above.
1.5. Calibration
The µ-FLOW and LIQUI-FLOW are factory calibrated. Periodical inspection, recalibration or verification of the accuracy
may be subject to individual requirements of the end user.
Bronkhorst certifies that the instrument meets the rated accuracy. Calibration has been performed using measurement
standards traceable to the Dutch Metrology Institute (VSL).
1.6. Maintenance
The µ-FLOW and LIQUI-FLOW needs no regular maintenance if operated properly, with clean media, compatible with
the wetted materials, avoiding pressure and thermal shocks and vibrations. Units may be purged with a clean, dry and
inert gas.
In case of severe contamination, cleaning the inside of the device may be required. After cleaning, recalibration of the
instrument is recommended.
Inexpertly servicing instruments can lead to serious personal injury and/or damage to the instrument or
the system it is used in. Servicing must therefore be performed by trained and qualified personnel.
Contact your Bronkhorst representative for information about cleaning and calibration. Bronkhorst has
a trained staff available.
1.7. Documentation
The µ-FLOW and LIQUI-FLOW come with all necessary documentation for basic operation and maintenance. Some parts
of this manual refer to other documents, most of which can be downloaded from the Bronkhorst website. Calibration
certificates, test certificates and material certificates are included in the scope of delivery or can be provided on request.
The documentation listed in the following table is available on the µ-FLOW and LIQUI-FLOW product
pages under
Type
Document name
µ-FLOW and LIQUI-FLOW brochure
Brochure
Manuals
Instruction manual µ-FLOW and LIQUI-FLOW (this document)
Quick Start Guide µ-FLOW and LIQUI-FLOW
Hook-up diagram Analog/RS232
________________________________________________________________________________________________
9.17.161
µ-FLOW & LIQUI-FLOW Instruction Manual L01, L13(I), L23(I), L13V12, L23V12
Bronkhorst
www.bronkhorst.com/products
®
______________________________________
8DIN connector
LED status indicator and push button
Outlet connection
Document no.;
IP40 enclosure
9.60.021
9.17.161
n/a
9.16.119
Document no.;
IP65 enclosure
9.60.021
9.17.161
n/a
9.16.126
Page 9 of 50
Need help?
Do you have a question about the u-FLOW Series and is the answer not in the manual?