4.10.2.2. Fieldbus operation
For information about parameter access and availability for Bronkhorst® instruments in a specific
fieldbus network, consult the according Documentation.
4.11. Hardware interface
The µ-FLOW and LIQUI-FLOW offer – depending on the installed digital interface type – the following controls:
LED indications
•
•
Multifunctional switch
Rotary switches
•
4.11.1.
On top of the housing, the instrument is equipped with two (or three) LED indicators and a multifunctional switch, which
can be used to monitor the instrument visually and start several functions manually.
The LEDs on top of the instrument indicate the operational state. The meaning of some indications depends on the
specific fieldbus interface of the instrument (if installed).
Position
Colour
•
Left
/
Middle
•
Right
•
*) The Interface status LED is only used by (Ethernet based) interface types EtherCAT®, POWERLINK and PROFINET.
**) Different interface types use specific names for the different indicator LEDs (indicated between brackets, also see
the specific Documentation)
The tables below list the different LED indications:
•
Green
Pattern
Time
off
continuous
on
continuous
0.1 sec on,
short flash
2 sec off
0.2 sec on,
blink
0.2 sec off
2 sec on,
long flash
0.1 sec off
•
Red
Pattern
Time
off
continuous
on
continuous
short flash
0.1 sec on,
2 sec off
________________________________________________________________________________________________
9.17.161
µ-FLOW & LIQUI-FLOW Instruction Manual L01, L13(I), L23(I), L13V12, L23V12
LED indications
Function
•
interface status*
(bi-colour)
operational mode (Mode/MOD/RUN)**
(green)
error/warning messages (Error/Err/NET)**
(red)
Indication
Power off or program not running
Normal operation mode
No communication, valves are in
Special function mode; the instrument is busy performing a special function
Configuration mode; the 8DIN connector is set for RS232 communication
(ProPar) at 38,400 Baud
Indication
No error
Critical error; the instrument needs servicing before it can be used
FLOW-BUS
Node occupied: re-install instrument
PROFIBUS DP
No data exchange between master and slave (automatic recovery)
Modbus
Data is being received or transmitted
DeviceNet™
Minor communication error
EtherCAT
®
Instrument is not in OP mode
PROFINET
No application relation established
Bronkhorst
®
______________________________________
safe/default state
Page 23 of 50