Device Description; Technical Specifications - Stryker X8000 Service Manual

Light source
Hide thumbs Also See for X8000:
Table of Contents

Advertisement

Device Description

e Stryker X8000 Light Source is a light-generating unit
designed to illuminate surgical sites during endoscopic
applications.
e X8000 uses a 300-watt xenon bulb to
generate light, which it delivers to the surgical site via a
beroptic light cable.
e X8000 is compatible with all Stryker light cables, and,
with the proper light cable and adapters, can connec t to any
exible or rigid endoscope.
e X8000 is equipped with Electronic Sc ope Sensing
Technology (ESST), a special safety feature that helps
prevent accidental burns caused by a light cable that is
not connected to the scope. When operated with an ESST
light cable, the X8000 senses when the scope and the light
cable are separated and places the light source in Standby
mode. In Standby mode, the X8000 will reduce light output
to a minimum, preventing the light cable from generating
excessive heat.
6
Technical Specifications
Electrical
Primary:
100 – 120 VAC, 50/60 Hz, 450 W
220 – 240VAC, 50/60 Hz, 450 W
Fuses (2): 5.0A, 250V
Dimensions
Height:
4.75˝ (12.1 cm)
Width:
12.5˝ (31.8 cm)
Depth:
16.8˝ (42.7 cm)
Weight:
16.0 lbs (7.3 kg)
Fiberoptic Cable Range
2 – 6.5 mm diameter
Bulb
Type:
300 Watt Xenon (Elliptical)
Life:
Approximately 500 hours
Operating Conditions
10 – 40°C
30 – 75% Relative Humidity 
Transportation and Storage
-20 – 50°C
10 – 75% Relative Humidity 
700 – 1060 hPa
Classi
cations and Approvals
Complies with medical safety standards:
• IEC 60601-1:2005
• CAN/CSA C22.2 No.601.1-M90
• UL 60601-1: 2003
Complies with medical EMC standard:
• IEC 60601-1-2:2001
Class 1 Equipment
Type CF applied parts
Water Ingress Protection, IPX0 — Ordinary
Equipment
Continuous Operation
Patent Protection
U.S. #5,850,496; 6,110,107; and 6,689,050.
Other patents pending.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

220-200-000

Table of Contents