Page 1
Power-PRO™ XT Model 6500 Operations/Maintenance Manual For parts or technical assistance: USA: 1-800-327-0770 2010/10 6500-509-001 REV B www.stryker.com...
Refer to your local distributor for return and/or collection systems available in your country. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
NOTE This provides special information to make maintenance easier or important instructions clearer. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
The device has side rails, supports for fluid infusion equipment, and patient restraint straps. The frame is collapsible for use in an ambulance. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
The Power-PRO™ XT is designed to be compatible with competitive cot fastener systems. DeWALT ® is a registered trademark of Black & Decker Inc. Patents pending. The yellow and black color scheme is a proprietary trademark of Stryker Corporation. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 11
Introduction SPECIFICATIONS (CONTINUED) Environmental Conditions Operation 130 F (54 C) Temperature -30 F (-34 C) 100% Relative Humidity Atmospheric Pressure Return To Table of Contents www.stryker.com 6500-509-001 REV B...
3800 E. Centre Avenue Portage, MI 49002 Please have the serial number (A) of your Stryker product available (as shown in Figure 1) when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication. SERIAL NUMBER LOCATION Head End Figure 1: Cot Serial Number &...
Housing Hydraulic Unit Wheel Lock (Optional) Transport Wheels Cot Retaining Post Transport Wheels Short Safety Hook Stryker part number Long Safety Hook 6060-036-017 Stryker part number J Safety Hook 6060-036-018 Stryker part number 6092-036-018 Figure 2: Cot Components Return To Table of Contents www.stryker.com...
• Any emergency vehicle to be used with this cot must have the in-fastener shut-off system installed (see page 26). • It is the responsibility of the cot operator to ensure that the cot being used in the Stryker Cot Fastener Systems meets the installation specifications listed on page 25.
Page 15
• Stryker recommends that a two person operation is used when using the kickstand. • Make sure that the patient weight is centered on the cot before using the kickstand.
Page 16
• SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY. If the products described above are used to clean Stryker EMS equipment, measures must be taken to insure the cots are wiped with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the cots will leave a corrosive residue on the surface of the cots, possibly causing premature corrosion of critical components.
Page 17
• Always remove the battery before washing the cot. • A preventative maintenance program should be established for all Stryker EMS equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product. Close attention should be given to safety features including, but not limited to: • Hydraulic power mechanism...
Page 18
• These are general instructions for installation of the Pedi-Mate®. Safe and proper use of the Pedi-Mate® is solely at the discretion of the user. Stryker recommends that all users be trained on the proper use of the Pedi-Mate® before using it in an actual situation. Retain these instructions for future reference. Include them with the product in the event of transfer to new users.
Ensure proper hand placement on hand grips. Hands should be clear of red safety bar pivots while loading and unloading the cot or whenever changing height position of the cot with two or more operators. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
The cot can be set at any cot load height position. Establish the required cot load height before placing the cot into service. Figure 4.3: Adjusting Height Figure 4.4: Securing Cables Return To Table of Contents www.stryker.com 6500-509-001 REV B...
With the cot loaded with a minimum of 99 pounds (45 kg), check the raise/ lower function With the cot loaded with a minimum of 99 pounds (45 kg), check the load/ unload function Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 23
Verify head end storage flat (if equipped) is installed properly Verify pillow (if equipped) included Verify 36" restraint extender (if equipped) is included Verify the transfer flat (if equipped) is included Return To Table of Contents www.stryker.com 6500-509-001 REV B...
WARNING It is the responsibility of the cot operator to ensure that the cot being used in the Stryker Cot Fastener Systems meets the installation specifications listed on page 25. Injury may result if a non-compatible cot is used in the Stryker Fastener System.
• Any emergency vehicle to be used with this cot must have the in-fastener shut-off system installed. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Federal Regulation KKK-A-1822. Stryker offers three different types of safety hooks that are ordered and shipped with your cot. These safety hook types are designed to meet the needs of various emergency vehicle configurations, specifically the length and location of the floor structure support that is located in the rear of the vehicle.
* The length of the socket head cap screws depends on the thickness of the vehicle floor. Use screws that are long enough to go completely through the patient compartment floor, washer and nut by at least two full threads. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
• To avoid injury, verify that the safety bar has engaged the safety hook before removing the cot from the patient compartment. Note: Stryker recommends that, prior to installation, the certified mechanic plan the placement of the safety hook in the rear of the vehicle.
Verify that the cot legs lock into the load position before disengaging the safety bar from the safety hook. Failure to properly lock the cot into position can cause injury to the patient or operator and/or damage to the cot. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
• Read all labels and instructions on the cot before using the cot. • Use a minimum of two (2) trained operators to operate the cot while a patient is on the cot. If additional assistance is needed, see “Using Additional Assistance” on page • Do not adjust, roll or load the cot into a vehicle without advising the patient. Stay with the patient and control the cot at all times. • The cot can be transported in any position. Stryker recommends that the operators transport the patient in the lowest comfortable position to maneuver the cot. • Only use the wheel lock(s) during patient transfer or without a patient on the cot. • Always use the restraint straps. • Use properly trained helpers, when necessary, to control the cot. WARNING • Improper usage of the cot can cause injury to the patient or operator. Operate the cot only as described in this manual.
• If the cot is equipped with the optional kickstand, make sure that the kickstand remains in the retracted position and does not engage during transport. • Transporting the cot in lower positions reduces the potential of a cot tip. If possible, obtain additional assistance or take an alternate route. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
CAUTION Do not “jog” the cot past the established cot load height of the product when the safety bar engages the vehicle safety hook or damage may occur to the product. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
• Once the weight is off of the ground, the operator(s) must support the load of the patient, cot and any accessories. Failure to support the load properly may cause injury to the patient or operator. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
• Do not “jog” the cot past the load height while the safety bar is engaged. Figure 22: Disengaging the Safety Bar Return To Table of Contents www.stryker.com 6500-509-001 REV B...
• Do not “jog” the cot past the load height while the safety bar is engaged. Figure 25: Roll the Cot out of the Vehicle Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Figure 29. Note: When operating the manual back-up release handle, avoid rapid lifting or lowering of the base or movement may appear sluggish; lift with a slow constant motion. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
When unloading the cot from the patient compartment, ensure that the caster wheels are safely set on the ground or damage to the product may occur. Figure 32: Disengaging the Safety Bar Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Remove the load wheels from the patient compartment of the vehicle. CAUTION When unloading the cot from the patient compartment, ensure that the caster wheels are safely set on the ground or damage to the product may occur. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Ensure proper hand placement on hand grips. Hands should be clear of red safety bar pivots while loading and unloading the cot or whenever changing height position of the cot with two or more operators. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Note: Batteries slowly lose power when not on the charger. CAUTION Remove the battery if the cot is not going to be used for an extended period of time (more than 24 hours). Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Note: Batteries slowly lose power when not on the charger. CAUTION Remove the battery if the cot is not going to be used for an extended period of time (more than 24 hours). Return To Table of Contents www.stryker.com 6500-509-001 REV B...
See the SMRT Power System Operations/Maintenance manual (6500-009-101) for additional SMRT Pak and SMRT Charger information. See the Battery System manual for battery and charger information. DeWALT ® Return To Table of Contents 6500-509-001 REV B www.stryker.com...
The hour meter, located on the foot end control enclosure as shown in Figure 38, indicates the amount of time (HHH.H hours) that the hydraulics have been activated. You can use the hour meter to determine the frequency for preventative maintenance procedures as listed on page Figure 38: Hour Meter Return To Table of Contents www.stryker.com 6500-509-001 REV B...
40.3. The arrows indicate alternate attachment areas. When attaching the restraint straps to the cot, the attachment points should provide both strong anchorage and proper restraint position while not interfering with equipment and accessories. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 49
Inspect the restraints at least once a month (more frequently if used heavily). Inspection should include checking for a bent or broken receiver or latch plate, torn or frayed webbing, etc. Any restraint showing wear or not operating properly must be replaced immediately. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
USING THE OPTIONAL RESTRAINT BELT EXTENSION Use the restraint belt extension, as shown in Figure 41, for extra length when buckling the lap belt around large patients. Figure 41: Attaching the Restraint Belt Extension Return To Table of Contents 6500-509-001 REV B www.stryker.com...
To raise the backrest, as shown in Figure 42, squeeze handle (A) for pneumatic assist in lifting the backrest to the desired height. To lower the backrest, squeeze handle (A) and push down on the backrest frame until the backrest has reached the desired height. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
To lower the footrest, lift the footrest frame (A) and while holding the frame, lift up on the release handle (B) until the bracket disengages. Lower the footrest until flat. Foot End Figure 44: Footrest Elevated Return To Table of Contents www.stryker.com 6500-509-001 REV B...
To lower the gatch, lift either of the red lifting loops (A) to relieve pressure on the locking mechanism and while holding the loop, lift up on the red release handle (B) until the bracket disengages. Lower the gatch to the flat position. Foot End Figure 45: Gatch Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Wheel lock(s) are only intended to help prevent the cot from rolling while unattended and to aid in patient transfer. A wheel lock may not provide sufficient resistance on all surfaces or under loads. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
To avoid damage to the I.V. pole, the weight of the I.V. bags or equipment must not exceed 40 pounds (18 kg). Figure 47.1: 2-Stage I.V. Pole Storage Position Figure 47.2: 2-Stage I.V. Pole Return To Table of Contents 6500-509-001 REV B www.stryker.com...
To avoid damage to the I.V. pole, the weight of the I.V. bags or equipment must not exceed 40 pounds (18 kg). Figure 48.1: 3-Stage I.V. Pole Storage Position Figure 48.2: 3-Stage I.V. Pole Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Part Number Page Number Equipment Hook 6500-147-000 Oxygen Bottle Holder, Removable 6080-140-000 Kickstand 6085-002-000 Restraint Package 6091-300-010 Pedi-Mate® Transfer Flat 6005-001-001 Pocketed Back Rest Pouch 6500-130-000 Head End Storage Flat 6500-128-000 Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Use the equipment hook (A) (see Figure 49) to hang additional accessories or equipment, such as defibrillators or monitors. Figure 49: Equipment Hook CAUTION To avoid damage to the equipment hook, the weight of the accessories or equipment must not exceed 35 pounds (15.9 kg). Return To Table of Contents www.stryker.com 6500-509-001 REV B...
• To avoid damage to the retractable head section oxygen bottle holder (if equipped), the weight of the equipment must not exceed 40 pounds (18 kg). • Do not use two head end oxygen bottle holders at the same time. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
• The kickstand (p/n 6085-002-000) is not compatible with the optional base storage net (p/n 6500-160-000). WARNING • Stryker recommends that a two person operation is used when using the kickstand. • Make sure that the patient weight is centered on the cot before using the kickstand.
Figure 52.2: Fastening the Pedi-Mate® Buckle WARNING To avoid accidental release of the Pedi-Mate®, and possible injury to the infant, ensure that the buckle is located away from obstructions on the cot or accessories. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 63
Note: These are general instructions for installation of the Pedi-Mate®. Safe and proper use of the Pedi-Mate® is solely at the discretion of the user. Stryker recommends that all users be trained on the proper use of the Pedi-Mate® before using it in an actual situation. Retain these instructions for future reference. Include them with the product in the event of transfer to new users.
• Do not store items under the cot mattress. Storing items under the mattress can interfere with the operation of the cot. • The weight of the equipment in the pocketed backrest storage pouch (if equipped) must not exceed 20 pounds (9 kg). Return To Table of Contents 6500-509-001 REV B www.stryker.com...
• Do not store items under the cot mattress. Storing items under the mattress can interfere with the operation of the cot. • The weight of the equipment in the head end storage flat (if equipped) must not exceed 40 pounds (18 kg). Return To Table of Contents www.stryker.com 6500-509-001 REV B...
• Be careful when retracting the base to avoid damaging items stored in the base storage net. Note: The kickstand (p/n 6085-002-000) is not compatible with the optional base storage net (p/n 6500-160-000). Return To Table of Contents 6500-509-001 REV B www.stryker.com...
• Always remove the battery! Never wash the cot with the battery installed. • Follow the cleaning solution manufacturer’s dilution recommendations exactly. • The preferred method Stryker Medical recommends for power washing the cot is with the standard hospital surgical cart washer or hand held wand unit.
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY. If the products described above are used to clean Stryker EMS equipment, measures must be taken to insure the cots are wiped with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the cots will leave a corrosive residue on the surface of the cots, possibly causing premature corrosion of critical components.
Preventative maintenance should be performed at a minimum of annually. A preventative maintenance program should be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product. The cot requires regular maintenance. Establish and follow a maintenance...
Training Record Training Date Training Method Trainee Name Basic Refresher Owner’s Manual, In-Service, Training Update Formal Class, Etc. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Rod-Side (Non-Locking) Manual Valve Cap-Side (Locking) Manual Valve Item Connection Connection Item DETAIL B Orange A-Valve Blue Wiring Schematics B-Valve Main Cable 8-Pin Connector White Green Pressure Switch Brown Black Wireless Expansion Port Yellow Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Top Hall Effects Stop until button is pressed Jog Up Sensor Detected? Ramp in motor (Up) Run motor at max speed Top Hall Effects Ramp out motor and stop Sensor Detected? Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 78
Open B-Valve and Ramp in Motor Hold B-Valve open until button released or pressure switch closes Run motor at max speed Bottom Hall Effects Sensor Detected? Ramp out motor and stop, Close B-Valve Return To Table of Contents 6500-509-001 REV B www.stryker.com...
If voltage is present, replace the wire harness. C. If the GREEN light turns on, but does not lower, try the other switch. If the other switch works, replace the bad switch. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 80
(H) (-) lead on black (+) lead on green while pressing the extend (+) button. If voltage is not present, replace the electronics assembly. If voltage is present, replace the main cable. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 81
Check manual cable adjustment. mode. Change ‘non-locking’ manual valve. High Speed retract does not engage. 1. Check that weight is off of the casters. Change pressure switch. Change hall effect cable. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
The parts and accessories listed on these pages are all currently available for purchase. Some of the parts identi- fied on the assembly drawing parts in this manual may not be individually available for purchase. Please call Stryker Customer Service USA: 1-800-327-0770 (Option 2) for availability and pricing.
Page 85
Restraint, Shoulder Harness 6060-260-045 Restraint Strap Plastic Cap (Short) 6082-160-051 Restraint Strap Plastic Cap (Tall) 6082-160-055 Safety Hook, Long 6060-036-017 Safety Hook, Short 6092-036-018 Siderail Assembly 6060-026-010 Storage Flat Option 6500-128-000 Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Remove the port plug using a 3/16" Allen wrench. Fill the reservoir up to the bottom of the fill port. Replace the plug and run the cot up and down a few times. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Note: If the arrow on the bottom bracket of the retaining post points toward the head end of the cot, the retaining post is set for an X-frame style cot. If the arrow points toward the foot end of the cot the post is set for an H-frame style cot. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Cot Assembly 6500-301-010 Rev A (Reference Only) Torque nut to 6.0 ± 0.5 Nm Torque nut to 55 ± 5 ft-lb Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 90
Cot Assembly Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 91
Cot Assembly Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 92
Cot Assembly Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 93
Cot Assembly Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 94
6082-026-010 Fold Down Siderail Assembly 6060-090-002 Serial Number Tag 6252-001-139 Label, Do Not Lubricate 0025-133-000 Dome Head Rivet 6082-032-045 Fowler Skin 6082-090-043 Label, Stryker 6060-090-004 Label 6100-003-125 Straight “T” Pivot 6550-001-201 Label, Oxygen Caution 6500-001-015 Foot End Assembly (page 116)
Page 95
Plastic Extrusion Spacer 0056-028-000 Bumper 0011-013-000 Flat Washer 6500-001-233 Label, Sensor Housing 6500-001-234 Label, Power-PRO™ XT Spec 6085-001-155 Label, Weight Capacity 0059-211-000 Cable Tie 0011-004-000 Flat Washer 0004-614-000 Button Head Cap Screw Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Base Assembly 6500-101-012 Rev A (Reference Only) Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 97
Base Assembly Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 98
Base Assembly Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 99
Button Head Cap Screw 0081-244-000 Flange Bearing 6060-002-010 Molded Wheel Assembly 0016-049-000 Nylock Hex Cap 6090-001-009 Caster Nut 6500-301-021 Outer Lift Tube Assembly (page 102) 2 6500-301-022 Inner Lift Tube Assembly (page 103) Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 100
Base Tube Pivot Post 6500-001-227 Base Tube Pivot Bearing 6500-001-228 Base Tube Pivot Post 6500-001-229 Foot Base Tube 6500-001-230 Plastic Extrusion Spacer 0014-002-000 Flat Washer 0014-040-000 Flat Washer 0016-002-000 Fiberlock Nut Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Retaining Post Cap 6060-004-044 Post Tube 6080-090-108 Lift Here Label 6500-001-189 Top Pin Bracket 6500-001-190 Bottom Pin Bracket 6500-001-302 Base Tube Protector 6085-001-056 Outer Base Tube Weldment 6085-001-057 Outer Base Tube Weldment Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Inner Lift Tube, Litter Pivot - 6500-001-034 Patient Right Assembly Rev F Item Part No. Part Name Qty. 0004-634-000 Button Head Cap Screw 0055-100-075 6500-001-355 Inner Lift Tube Weldment 6500-001-125 Dead Stop Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Inner Lift Tube, Litter Pivot - 6500-001-035 Patient Left Assembly Rev F Item Part No. Part Name Qty. 0004-634-000 Button Head Cap Screw 0055-100-075 6500-301-053 Inner Lift Tube Weldment 6500-001-125 Dead Stop Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 106
Notes Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Hall Sensor Assembly 6500-001-028 Rev A (Reference Only) Item Part No. Part Name Qty. 0004-596-000 Button Head Cap Screw 6500-001-124 Sensor Housing 6500-001-160 Hall Effects Sensor 6500-001-199 Housing Cover Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Sensor Housing Assembly 6500-001-029 Rev A (Reference Only) Item Part No. Part Name Qty. 0004-596-000 Button Head Cap Screw 6500-001-124 Sensor Housing 6500-001-199 Housing Cover Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Powerplant Assembly 6500-001-031 Rev B (Reference Only) Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 114
Hydraulics Assembly (page 115) 6500-001-105 Litter Support Cross Tube 6500-001-169 Cylinder Rod End 6500-001-194 Motor Mount 6500-001-212 Motor Mount Cross Bar 6500-001-249 Plastic Extrusion Spacer 6500-001-250 Plastic Extrusion Spacer 6500-001-294 Motor Mount Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Foot End Assembly 6500-001-015 Rev E (Reference Only) Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 117
Foot End Assembly Route Hall Sensor Cable around Item (AP) and through this area Route Switch Cables around Item (AP) and through this area Route Manual Release Cable through this area Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 118
Foot End Assembly Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 119
6500-001-147 Single Spring 6500-001-153 Light Panel ORB 6500-001-154 Pull Handle 6500-001-156 Manual Release Cable 6500-001-159 Cable Assembly 6500-001-161 Hall Effects Cable 6500-001-275 Wire Route Clip 0028-116-000 Push Nut 6080-090-101 Label, Warning Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Nylock Hex Nut 0025-079-000 Dome Head Rivet 0025-132-000 Dome Head Rivet 6060-090-004 Label, Small 6082-001-085 2” Adhesive Loop Pole 6500-001-019 Trend Assembly (page 121) 6500-001-197 Foot Section Tube 6500-001-198 Foot Section Skin Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Part Name Qty. 0004-592-000 Button Head Cap Screw 0004-596-000 Button Head Cap Screw 0016-102-000 Nylock Hex Nut 6500-001-048 Gatch Assembly (page 123) 6085-001-031 Gatch Support Assembly (page 126) 1 6550-001-197 Velcro® Strap Return To Table of Contents 6500-509-001 REV B www.stryker.com...
6085-001-031 Rev C (Reference Only) Item Part No. Part Name Qty. 0004-614-000 Button Head Cap Screw 0025-079-000 Dome Head Rivet 0946-001-155 Bumper 6085-001-135 Gatch Support, Foot End 6500-001-346 Gatch Crosstube 0056-028-000 Black TPR Bumper Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Button Assembly - 6500-101-016 Rev B Item Part No. Part Name Qty. 6500-101-130 Switch 6500-001-359 Button Upper Housing, Foot End Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Button Head Cap Screw 0015-050-000 0016-028-000 0021-119-000 Set Screw 0021-138-000 Set Screw 0025-131-000 Rivet 0028-076-000 External Retaining Ring 0946-035-025 Liner 6060-032-038 Yoke 6060-032-040 Pivot 6082-032-050 Fowler Weldment 6082-032-052 Release Handle Weldment Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Head Section Assembly - 6085-001-037 Rev B SEE DETAIL Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 130
Sliding Head Section Safety Bar 6500-001-323 Safety Hook Pivot, Right 6500-001-324 Safety Hook Pivot, Left 6500-001-325 Safety Bar Torsion Spring, Left 6500-001-326 Safety Bar Torsion Spring, Right 6500-001-360 Head Section Sleeve Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Telescoping Tube Assembly - 6085-001-036 Rev C Item Part No. Part Name Qty. 0025-079-000 Dome Head Rivet 6085-001-144 Head Section Telescoping Tube 6500-001-080 Load Wheel Casting 6500-001-081 Load Wheel Casting 6500-001-084 Front Lifting Bar Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Head Section Lock Assembly - 6500-001-026 Rev C Item Part No. Part Name Qty. 0038-570-000 Compression Spring 0038-134-000 Compression Spring 6500-001-091 Top Latch Housing 6500-001-092 Bottom Latch Housing 6500-001-025 Latch Assembly 6500-001-095 Actuation Slide Return To Table of Contents 6500-509-001 REV B www.stryker.com...
In-Fastener Shut-Off Assembly - 6500-001-027 Rev C Item Part No. Part Name Qty. 0004-376-000 Button Head Cap Screw 0015-016-000 Square Nut 0025-079-000 Rivet 6500-001-271 Magnet 6500-001-272 Holder Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Fiberlock Hex Nut 0025-079-000 Dome Head Rivet 6085-001-171 Oxygen Strap, Head End 6085-001-172 Oxygen Fowler Guard 6085-001-173 Oxygen Bottle Holder, Top 6085-001-174 Oxygen Bottle Holder, Bottom Note: Max Load = 40 pounds Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Label, Fowler Oxygen Bottle Holder 6500-001-260 Fowler Oxygen Bottle Holder Cover 6500-001-261 Fowler Oxygen Bottle Holder Strap 6500-001-262 Neoprene Pad 0025-079-000 Dome Head Pop Rivet Note: Max Load = 40 pounds Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Qty. 6100-115-060 Clip, I.V. Pole 6500-001-041 I.V. Pole Assembly, 2-Stage, Right (page 138) Item Part No. Part Name Qty. 6100-115-060 Clip, I.V. Pole 6500-001-043 I.V. Pole Assembly, 3-Stage, Right (page 139) Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Two-Stage I.V. Pole Assembly Patient Right - 6500-001-041 Rev C Item Part No. Part Name Qty. 6070-090-105 Label, Caution 6070-210-070 Pole Assembly 6100-115-051 Socket Weldment 6500-001-253 Label 0025-079-000 Dome Head Rivet 6070-110-037 I.V. Pivot Pin Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Three-Stage I.V. Pole Assembly Patient Right - 6500-001-043 Rev D Item Part No. Part Name Qty. 6070-090-105 Label, Caution 6070-215-070 Pole Assembly 6100-115-051 Socket Weldment 6500-001-255 Label 6070-110-037 I.V. Pivot Pin 0025-079-000 Dome Head Rivet Return To Table of Contents www.stryker.com 6500-509-001 REV B...
I.V. Pole Assembly, 2-Stage, Left (page 141) Optional 3-Stage I.V Pole Assembly, Left - 6500-216-000 Item Part No. Part Name Qty. 6100-115-060 Clip, I.V. Pole 6500-001-044 I.V. Pole Assembly, 3-Stage, Left (page 142) Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Two-Stage I.V. Pole Assembly Patient Left - 6500-001-042 Rev C Item Part No. Part Name Qty. 6070-090-105 Label, Caution 6070-210-070 Pole Assembly 6100-115-051 Socket Weldment 6500-001-254 Label 0025-079-000 Dome Head Rivet 6070-110-037 I.V. Pivot Pin Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Three-Stage I.V. Pole Assembly Patient Left - 6500-001-044 Rev D Item Part No. Part Name Qty. 6070-090-105 Label, Caution 6070-215-070 Pole Assembly 6100-115-051 Socket Weldment 6500-001-256 Label 6070-110-037 I.V. Pivot Pin 0025-079-000 Dome Head Rivet Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Optional Three-Stage I.V. Pole Assembly, Dual - 6500-217-000 Item Part No. Part Name Qty. 6100-115-060 Clip, I.V. Pole 6500-001-043 I.V. Pole Assembly, 3-Stage, Right (page 139) 6500-001-044 I.V. Pole Assembly, 3-Stage, Left (page 142) Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Label, SMRT Power 6500-027-000 Rev B (Reference Only) 6500-028-000 Rev B (Reference Only) Item Part No. Part Name Qty. 6500-001-356 Label, SMRT Power Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Label, DeWALT® 6500-270-000 Rev B (Reference Only) 6500-272-000 Rev B (Reference Only) Item Part No. Part Name Qty. 6500-001-258 Label, DeWALT® Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Oxygen Bottle Holder, Foot End 6500-140-000 Oxygen Bottle Holder, Removable 6080-140-000 Kickstand 6085-002-000 Head Extension with Pillow 6100-044-000 Transfer Flat 6005-001-001 Defibrillator Platform 6500-170-000 Pocketed Back Rest Pouch 6500-130-000 Head End Storage Flat 6500-128-000 Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Equipment Hook - 6500-147-000 Rev B Item Part No. Part Name Qty. 0025-079-000 Rivet Dome Head 6500-001-237 Equipment Hook Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Part Name Qty. 0004-508-000 Button Head Cap Screw 0011-001-000 Washer 0016-003-000 Nylock Nut 0025-079-000 Dome Head Rivet 6060-140-013 Strap Assembly 6500-001-040 Oxygen Bottle Holder (page 150) Note: Max Load = 40 pounds Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Head Extension - 6100-044-000 Rev B Item Part No. Part Name Qty. 6100-041-030 Pillow 6100-044-012 Head Extension Assembly (page 144) Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Backrest Pouch - 6500-130-000 Rev A Item Part No. Part Name Qty. 6500-001-241 Pocketed Backrest Storage Pouch Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Head End Storage Flat - 6500-128-000 Rev A Item Part No. Part Name Qty. 6500-001-232 Head End Storage Flat Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Batteries must be delivered to a battery collection site. Visit the RBRC website (www.rbrc.org) to find a nearby collection site or call the phone number shown on the recycling symbol. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 159
Batteries must be delivered to a battery collection site. Visit the RBRC website (www.rbrc.org) to find a nearby collection site or call the phone number shown on the recycling symbol. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 160
Recycling Passport Assembly Part Number: 6500-101-016 (Reference Only) Exploded view of switch assembly Item Recycling/Material Code Important Information Printed Circuit Board Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 161
Recycling Passport Assembly Part Number: 6500-001-028 (Reference Only) Exploded view of hall sensor assembly Item Recycling/Material Code Important Information Printed Circuit Board Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 162
Recycling/Material Code Important Information Printed Circuit Board Contains Liquid Crystal Display External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 163
Exploded view of hydraulics assembly Item Recycling/Material Code Important Information Motor Contains Automatic Transmission Fluid* External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable * Mobil Mercon® V Synthetic Blend or equivalent Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 164
Recycling Passport Assembly Part Number: 6500-001-159 (Reference Only) Item Recycling/Material Code Important Information External Electrical Cable Return To Table of Contents 6500-509-001 REV B www.stryker.com...
• SMRT Charger (6500-201-010), AC/DC Charger (6500-070-000), • DeWALT ® DC/DC Charger (6500-072-000). • DeWALT ® Note: U is the alternating current mains voltage prior to application of the test level. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 166
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Page 167
If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model 6500 ower-PRO™ XT cot. Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m. Return To Table of Contents www.stryker.com 6500-509-001 REV B...
Page 168
(6500-070-000): Class A DC/DC Charger DeWALT ® (6500-072-000): N/A Cot: N/A SMRT Charger (6500-201-010): Complies Voltage Fluctuations Flicker Emissions AC/DC Charger DeWALT ® IEC 61000-3-3 (6500-070-000): Complies DC/DC Charger DeWALT ® (6500-072-000): N/A Return To Table of Contents 6500-509-001 REV B www.stryker.com...
Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing, at its option, any product that is, in the sole discretion of Stryker, found to be defective. Expendable components, i.e. mattresses, restraints, I.V. poles, storage nets, storage pouches, oxygen straps, and other soft goods, have a one (1) year limited warranty.
Claim will be limited in amount to the actual replacement cost. In the event that this information is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full.
Page 172
United States Stryker Medical 3800 E. Centre Ave., Portage, Michigan USA 49002 European Representative Stryker France ZAC Satolas Green Pusignan Av. De Satolas Green 69881 MEYZIEU Cedex France 2010/10 6500-509-001 REV B www.stryker.com...