Introduction/Basics
This device has to be serviced every two years. The chapter
describes this routine inspection.
Please use the maintenance and checklist included in the back
of the service manual to document the bi-annual inspection.
1-4

Testing Tools and Resources

The following tools and resources are required to perform the
maintenance and service tasks described in this manual.
Please make sure that all tools and resources are calibrated
and in perfect working order before commencing any service or
maintenance tasks.
Manometer
Range 0-70 mm Hg,
Test Class 1.0
Manometer
Range 0-7 bar,
Test Class 1.6
Manometer
Range 0-100 bar,
Test Class 1.6
Silicone Tube
8x2 mm
PVC Tube
6x4 mm, 1 m
T Connector
8-8-8 mm
Flow Column
Range 0-20 l/min CO2
Max. Pressure Drop at 9 l/min < 10
mbar
Flow Column
Range 0-40 l/min CO2
Max. Pressure Drop at 9 l/min < 10
mbar
Veress Hollow Needle
Length 100 mm,
Opening Diameter 1.4 mm,
Inner Diameter 1.6 mm
Heater
Calibration Box
T100/T101
Universal Multimeter
1
Introduction/Basics
1-5

Front of the Device

Start/Stop Key
1
Display
2
Gas Heater Connection
3
Insufflation Tube Connection
4
Nominal Gas Flow +
5
Nominal Gas Flow -
6
Gas Consumption Reset Key
7
Menu Key
8
Nominal Pressure -
9
Nominal Pressure +
bl
Line Voltage LED
bm
ON/OFF Key
bn