Page 1
Power-PRO™ TL Model 6550 Operations/Maintenance Manual For parts or technical assistance: USA: 1-800-327-0770 UK: +44 (0) 1635 262431 2010/06 6550-309-001 REV A www.stryker.com...
Refer to your local distributor for return and/or collection systems available in your country. BS EN 1789:2000 Certified Return To Table of Contents 6550-309-001 REV A www.stryker.com...
NOTE Provides special information to make maintenance easier or important instructions clearer. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
INTENDED USE OF PRODUCT The Stryker Model 6550 Power-PRO™ TL ambulance cot is a wheeled device which consists of a platform mounted on a wheeled frame that is designed to support patients in a horizontal position. The device has siderails and has the option available to support the temporary or permanent placement of I.V.
The Power-PRO™ TL is designed to be compatible with competitive cot fastener systems. DeWALT ® is a registered trademark of Black & Decker Inc. Patents pending. The yellow and black color scheme is a proprietary trademark of Stryker Corporation. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 10
Introduction SPECIFICATIONS (CONTINUED) Environmental Conditions Operation 130 F (54 C) Temperature -30 F (-34 C) 100% Relative Humidity Atmospheric Pressure Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Hambridge Rd. Newbury, Berkshire RG14 5EG, England Please have the serial number (see Figure 1) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication. SERIAL NUMBER LOCATION Serial Number HEAD END Figure 1: Cot Serial Number &...
Modifying the product also voids its warranty (see page 148). • It is the responsibility of the cot operator to ensure that the cot being used in the Stryker Cot Fastener Systems meets the installation specifications listed on page 22.
Page 14
• SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY. If the products described above are used to clean Stryker EMS equipment, measures must be taken to insure the cots are wiped with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the cots will leave a corrosive residue on the surface of the cots, possibly causing premature corrosion of critical components.
Page 15
• Always remove the battery before washing the cot. • A preventative maintenance program should be established for all Stryker EMS equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product. Close attention should be given to safety features including, but not limited to: • Hydraulic power mechanism...
Page 16
• These are general instructions for installation of the Pedi-Mate®. Safe and proper use of the Pedi-Mate® is solely at the discretion of the user. Stryker recommends that all users be trained on the proper use of the Pedi-Mate® before using it in an actual situation. Retain these instructions for future reference. Include them with the product in the event of transfer to new users.
Summary of Safety Precautions PINCH POINTS WARNING: Pinch Point WARNING: Pinch Point between litter interface and base when lowering litter. WARNING: Pinch Point Figure 3: Potential Pinch Points Return To Table of Contents www.stryker.com 6550-309-001 REV A...
• Stryker continually seeks advancements in product design and quality. Therefore, while this manual contains the most current product information available at the time of printing, there may be minor discrepancies between your cot and this manual. If you have any questions, please contact Stryker Customer Service or Technical Support at +44 (0) 1635 262431.
Verify siderails operate and latch properly Verify backrest cylinder operates properly through range of motion Verify the footrest operates properly Install body restraints. Restraints intact and operating properly. No rips or tears in mattress cover Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 20
Verify Pedi-Mate® restraint package (if equipped) is installed properly Verify pocketed backrest storage pouch (if equipped) is installed properly Verify oxygen bottle holder (if equipped) operates properly Verify 36 in restraint extender (if equipped) is included Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Caution must be used at all times during placement of the cot into the ambulance. WARNING It is the responsibility of the cot operator to ensure that the cot being used in the Stryker Cot Fastener Systems meets the installation specifications listed on page 22.
Page 22
Cot Fastener Installation Return To Table of Contents 6550-309-001 REV A www.stryker.com...
• To reduce the risk of damage to the cot or cot fastener, do not attempt to activate the cot height activation while it is engaged in the cot fastener (not included). Figure 7: Cot in Fastener (Head End) Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Press down firmly on the foot pedal until the locking mechanism disengages as shown in Figure 8. Roll the cot out of the patient compartment as shown in Figure 9. Figure 8: Press Foot End Figure 9: Remove Cot Return To Table of Contents 6550-309-001 REV A www.stryker.com...
• Read all labels and instructions on the cot before using the cot. • Use a minimum of two (2) trained operators to operate the cot while a patient is on the cot. If additional assistance is needed, see “Using Additional Assistance” on page • Do not adjust, roll or load the cot into a vehicle without advising the patient. Stay with the patient and control the cot at all times. • The cot can be transported in any position. Stryker recommends that the operators transport the patient in the lowest comfortable position to maneuver the cot. • Only use the wheel lock(s) during patient transfer or without a patient on the cot. • Always use the restraint straps. • Use properly trained helpers when necessary to control the cot and patient. WARNING • Improper usage of the cot can cause injury to the patient or operator. Operate the cot only as described in this manual.
• In addition to the head and foot end, larger patients may require additional operators. The Power-PRO TL cot can transport a patient in any height position, however, lower heights may improve balance and stability, so Stryker recommends keeping the cot in the lowest position that is comfortable for the cot operators.
Grasping the cot improperly can cause injury. Keep hands, fingers and feet away from moving parts. To avoid injury, use extreme caution when placing your hands and feet near the base tubes while raising and lowering the cot. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Lower the tail lift to the ground and ensure that it is fully lowered and stopped before disengaging the tail lift safety gate and allowing the cot to be rolled off of the tail lift. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
• To reduce the risk of damage to the cot or cot fastener, do not attempt to activate the cot height activation while it is engaged in the cot fastener (not included). Return To Table of Contents www.stryker.com 6550-309-001 REV A...
50 stones (318 kg) (700 lb). • Proper inspection and maintenance of the optioanl oxygen bottle holder accessory should be followed as identified in the Preventative Maintenance Checklist (see page 57). Return To Table of Contents 6550-309-001 REV A www.stryker.com...
• Once the weight is off of the ground, the operator(s) must support the load of the patient, cot and any accessories. Failure to support the load properly may cause injury to the patient or operator. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
• Activation of the manual back-up release handle may cause the cot to drop slowly if less than 7 stones (45 kg) (100 lb) are on the cot. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Note: Batteries slowly lose charge when not on the charger. CAUTION Remove the battery if the cot is not going to be used for an extended period of time (more than 24 hours). Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Note: Batteries slowly lose charge when not on the charger. CAUTION Remove the battery if the cot is not going to be used for an extended period of time (more than 24 hours). Return To Table of Contents www.stryker.com 6550-309-001 REV A...
See the SMRT Power System Operations/Maintenance manual (6500-009-101) for additional SMRT Pak and SMRT Charger information. See the Battery System manual for battery and charger information. DeWALT ® Return To Table of Contents 6550-309-001 REV A www.stryker.com...
(HHH.H hours) that the cot has been in use under power. You can use the hour meter to determine the frequency for preventative maintenance procedures as listed on page Figure 19: Hour Meter Return To Table of Contents www.stryker.com 6550-309-001 REV A...
The arrows indicate alternate attachment areas. When attaching the restraint straps to the cot, the attachment points should provide both strong anchorage and proper restraint position while not interfering with equipment and accessories. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 39
Inspect the restraints at least once a month (more frequently if used heavily). Inspection should include checking for a bent or broken receiver or latch plate, torn or frayed webbing, etc. Any restraint showing wear or not operating properly must be replaced immediately. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
USING THE OPTIONAL RESTRAINT BELT EXTENSION Use the restraint belt extension (6082-160-050), as shown in Figure 22, for extra length when buckling the lap belt around large patients. Figure 22: Attaching the Restraint Belt Extension Return To Table of Contents 6550-309-001 REV A www.stryker.com...
WARNING Siderails are not intended to serve as a patient restraint device. See page 38 for proper restraint strap usage. Failure to utilize the siderails properly could result in patient injury. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Make sure that the head or foot section is securely locked into position. CAUTION Never operate the cot with the head section or foot section in an unlocked position. Figure 24.2: Head and Foot Section Retracted Return To Table of Contents 6550-309-001 REV A www.stryker.com...
To lower the gatch, lift either of the red lifting loops (B) to relieve pressure on the locking mechanism and while holding the loop, lift up on the red release handle (C) until the bracket disengages. Lower the gatch to the flat position. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Operation Guide OPERATING THE WHEEL LOCK(S) The Power-PRO TL includes corner to corner wheel locks as a standard feature, with an option to specify locks for all four wheels. To secure the wheel lock(s), press down on the base of the locking pedal (A) until it rests firmly against the surface of the wheel as shown in Figure 26.
Note: When engaging the cot into the fastener, it is advised to disengage the caster steer lock feature to allow the head end casters to swivel freely and allow the head end of the cot to more easily align with the fastener. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
(A) as shown in Figure 28.3. Ensure that the pouch and its contents do not interfere with the cot operation before raising or lowering the cot or transporting the cot. Figure 28.3: Push Bar Storage Pouch Return To Table of Contents 6550-309-001 REV A www.stryker.com...
To avoid damage to the I.V. pole, the weight of the I.V. bags or equipment must not exceed 2.8 stones (18 kg) (40 lb). Figure 29.1: 2-Stage I.V. Pole Storage Position Figure 29.2: 2-Stage I.V. Pole Return To Table of Contents www.stryker.com 6550-309-001 REV A...
To avoid damage to the I.V. pole, the weight of the I.V. bags or equipment must not exceed 2.8 stones (18 kg) (40 lb). Figure 30.1: 3-Stage I.V. Pole Storage Position Figure 30.2: 3-Stage I.V. Pole Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Optional Accessories The accessories listed below can be purchased and installed on the Power-PRO TL cot. Operation Guide Accessory Part Number Page Number Equipment Hook 6500-147-000 Restraint Package 6091-300-010 Pedi-Mate® Backrest Storage Pouch 6500-130-000 Oxygen Bottle Holder 6550-002-020 Return To Table of Contents www.stryker.com...
• To avoid damage to the equipment hook, the weight of the accessories or equipment must not exceed 2.5 stones (16 kg) (35 lb). • To avoid damage to the equipment hook, remove all accessories or equipment from the hook while transporting the stretcher in the ambulance. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Figure 32.2: Fastening the Pedi-Mate® Buckle WARNING To avoid accidental release of the Pedi-Mate®, and possible injury to the infant, ensure that the buckle is located away from obstructions on the cot or accessories. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 52
Note: These are general instructions for installation of the Pedi-Mate®. Safe and proper use of the Pedi-Mate® is solely at the discretion of the user. Stryker recommends that all users be trained on the proper use of the Pedi-Mate® before using it in an actual situation. Retain these instructions for future reference. Include them with the product in the event of transfer to new users.
• The weight of the equipment in the pocketed backrest storage pouch (if equipped) must not exceed 1.4 stones (9 kg) (20 lb). • Ensure that the storage pouch does not interfere with the operation of the retractable head section. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
• To avoid damage to the oxygen bottle holder (if equipped), the weight of the equipment must not exceed 2.9 stones (18 kg) (40 lb). • Only use oxygen bottles which lie within the width of the cot and are large enough to be firmly held by the straps. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
• Always remove the battery! Never wash the cot with the battery installed. • Follow the cleaning solution manufacturer’s dilution recommendations exactly. • The preferred method Stryker Medical recommends for power washing the cot is with the standard hospital surgical cart washer or hand held wand unit.
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY. If the products described above are used to clean Stryker EMS equipment, measures must be taken to insure the cots are wiped with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the cots will leave a corrosive residue on the surface of the cots, possibly causing premature corrosion of critical components.
Preventative maintenance should be performed at a minimum of annually. A preventative maintenance program should be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product. The cot requires regular maintenance. Establish and follow a maintenance...
Training Record Training Date Training Method Trainee Name Basic Refresher Owner’s Manual, In-Service, Training Update Formal Class, Etc. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Cap-Side Hydraulic Hose Main Cable (See Detail B) Manual Release Cable Hydraulic Manifold Assembly (See Detail A) Electronics Housing Motor Mount (Control Board Inside) Switch Manual Release Hydraulic Cylinder Handle LCD Screen Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Manual Valve Cap-Side (Locking) Manual Valve Item Connection Connection Item DETAIL B Orange A-Valve Blue Wiring Schematics B-Valve Main Cable 8-Pin Connector White Green Pressure Switch Brown Black Wireless Expansion Port Yellow Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Pressure Switch Pressure Switch Open B-Valve Open? Open? Open B-Valve and Ramp in Motor Hold B-Valve open until button released or pressure switch closes Run motor at max speed Stop, Close B-Valve Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 65
Troubleshooting Guide ELECTRICAL SYSTEM BLOCK DIAGRAM Lift and Extend Functions Press Extend (+) Button Open A-Valve Run motor at max speed Stop Motor Return To Table of Contents www.stryker.com 6550-309-001 REV A...
If voltage is present, replace the wire harness. C. If the GREEN light turns on, but does not lower, try the other switch. If the other switch works, replace the bad switch. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 67
(H) (-) lead on black (+) lead on green while pressing the extend (+) button. If voltage is not present, replace the electronics assembly. If voltage is present, replace the main cable. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 68
Litter does not extend in the manual Check manual cable adjustment. mode. Change ‘non-locking’ manual valve. High Speed retract does not engage. 1. Check that weight is off of the casters. Change pressure switch. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Flip the gatch assembly (6550-001-019) toward the foot end of the cot until it rests on the telescoping foot end (6550-001-015). Reverse the above procedures to reassemble. Step 1 Step 3 Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Remove the port plug using a 3/16" Allen wrench. Fill the reservoir up to the bottom of the fill port. Replace the plug and run the cot up and down a few times. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Using the 5/32” Allen wrench and 7/16” combination wrench or socket, reinstall the socket screw. Test the pedal locking force and verify that the pedal holds properly before returning the cot to service. MINIMUM MAXIMUM Figure 35: Wheel Locking Force Adjustment Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Base Assembly 6550-001-012 Rev G (Reference Only) Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 86
Base Assembly Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 87
Base Assembly Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 88
Base Assembly Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 89
6500-001-227 Base Tube Pivot Post 6500-001-228 Inner Lift Tube Sleeve 6500-001-229 Base Tube, Foot 6550-001-203 Solid Spacer, Foot End 6500-001-308 Base Strap, Right 6500-001-309 Base Strap, Left 6500-001-310 Base Strap Clamp Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 90
Support Link 6550-001-106 Label, Steerlock 6500-001-327 Half Shell Bearing 6500-001-328 Lower Bearing Carrier 6500-001-329 Middle Bearing Carrier 6500-301-051 Base Pivot Lift Tube Weldment 6550-101-066 Inner Lift Tube Weldment, Base Pivot 2 Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Inner Lift Tube Assembly, Litter Pivot, Right - 6550-001-034 Rev C Item Part No. Part Name Qty. 0004-634-000 Button Head Cap Screw 0055-100-075 Riv Nut 6500-001-125 Base Dead Stop 6550-001-260 Litter Pivot Inner Lift Tube Weldment 1 Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Inner Lift Tube Assembly, Litter Pivot, Left - 6550-001-035 Rev C Item Part No. Part Name Qty. 0004-634-000 Button Head Cap Screw 0055-100-075 Riv Nut 6500-001-125 Base Dead Stop 6550-101-058 Litter Pivot Inner Lift Tube Weldment 1 Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Hitch Assembly, Head End 6550-001-021 Rev B (Reference Only) Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 95
Hitch Assembly, Head End Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 96
Label, Hitch Pinch Point, Head End 6550-001-087 Hitch Wear Strip, Head End 6550-001-134 Hitch Bar, Head End 6550-001-140 Hitch Bar Sleeve, Head End 6550-001-174 Hitch Cover, Head End 6550-001-175 Hitch Guide, Head End Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Mounted Hydraulics Assembly 6550-001-031 Rev A (Reference Only) Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 100
6500-001-164 Pivot, Cylinder Mount, Top 6500-001-169 Rod End, Cylinder 6500-001-194 Motor Mount 6500-001-212 Cross Bar, Motor Mount 6500-001-249 Plastic Extrusion, Spacer 6500-001-250 Plastic Extrusion, Spacer 6500-001-294 Motor Mount 6550-001-030 Hydraulics Assembly Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Hitch Assembly, Foot End 6550-001-022 Rev A (Reference Only) Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 104
Hitch Assembly, Foot End Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 105
Hitch Assembly, Foot End Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 106
6550-001-141 Hitch Bar Sleeve, Foot End 6550-001-146 Hitch Bar, Foot End 6550-001-167 Hitch Wear Guard, Foot End 6550-001-172 Cable Assembly 6550-001-173 Housing Cover, Foot End 6550-001-178 Bottom Plate Enclosure, Foot End Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Telescoping Foot End - 6550-001-015 Rev B Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 108
6085-001-170 Internal Bearing 6550-001-025 Bonded Frame, Foot End 6550-001-156 Section Release Trigger 6550-001-162 Link Release, Foot End 6550-001-163 Link Guide 6550-001-169 Bumper Pivot 6085-001-169 Headsection Nut 0004-168-000 Button Head Cap Screw Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Telescoping Head Section - 6550-001-020 Rev B Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 110
6085-001-170 Internal Bearing 6550-001-024 Bonded Frame, Head End 6550-001-156 Section Release Trigger 6550-001-161 Link Release, Head End 6550-001-163 Link Guide 6550-001-169 Bumper Pivot 6085-001-169 Headsection Nut 0004-168-000 Button Head Cap Screw Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Button Head Cap Screw 0015-050-000 Hex Nut 0016-028-000 Fiberlock Hex Nut 0021-180-000 Set Screw 0021-138-000 Set Screw 0028-076-000 External Retaining Ring 0946-035-025 Liner 6060-032-038 Gas Spring Yoke 6060-032-040 Fowler Lift Pivot 6550-001-051 Fowler Weldment Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Gatch Assembly 6550-001-019 Rev E (Reference Only) Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 113
Gatch Assembly Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 114
Telescoping Foot Section U-Tube 6550-001-124 Front Gatch Release 6550-001-125 Back Gatch Release 6550-001-126 Lever Front Gatch Release 6550-001-129 Gatch Pivot Cross Tube 6550-001-131 Gatch Bearing End Cap 6550-001-193 Gatch Handle Guard 0025-133-000 Rivet Return To Table of Contents 6550-309-001 REV A www.stryker.com...
No I.V. Pole Assembly Option 6550-218-000 Rev A (Reference Only) 0004-589-000 0004-589-000 Note: 0004-589-000 shown in cot assembly Item Part No. Part Name Qty. 6550-001-117 I.V. Pole Spacer Plate 6550-001-118 I.V. Clip Spacer Plate Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Note: 0004-589-000 shown in cot assembly Item Part No. Part Name Qty. 6100-115-060 I.V. Pole Clip 6500-001-041 2-Stage I.V. Pole Assembly, Right (page 118) 6550-001-117 I.V. Pole Spacer Plate 6550-001-118 I.V. Clip Spacer Plate Return To Table of Contents www.stryker.com 6550-309-001 REV A...
2-Stage I.V. Pole Assembly, Right - 6500-001-041 Rev C Item Part No. Part Name Qty. 6070-090-105 Label, No Push/Pull 6070-210-070 2-Stage I.V. Pole Assembly 6100-115-051 Socket Weldment 6500-001-253 Label 0025-079-000 Dome Head Rivet 6070-110-037 I.V. Pivot Pin Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Note: 0004-589-000 shown in cot assembly Item Part No. Part Name Qty. 6100-115-060 I.V. Pole Clip 6500-001-042 2-Stage I.V. Pole Assembly, Left (page 120) 6550-001-117 I.V. Pole Spacer Plate 6550-001-118 I.V. Clip Spacer Plate Return To Table of Contents www.stryker.com 6550-309-001 REV A...
2-Stage I.V. Pole Assembly, Left - 6500-001-042 Rev C Item Part No. Part Name 6070-090-105 Label, No Push/Pull 6070-210-070 2-Stage I.V. Pole Assembly 6100-115-051 Socket Weldment 6500-001-254 Label 0025-079-000 Dome Head Rivet 6070-110-037 I.V. Pivot Pin Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Note: 0004-589-000 shown in cot assembly Item Part No. Part Name Qty. 6100-115-060 I.V. Pole Clip 6500-001-041 2-Stage I.V. Pole Assembly, Right (page 118) 6500-001-042 2-Stage I.V. Pole Assembly, Left (page 120) Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Note: 0004-589-000 shown in cot assembly Item Part No. Part Name Qty. 6100-115-060 I.V. Pole Clip 6500-001-043 3-Stage I.V. Pole Assembly, Right (page 123) 6550-001-117 I.V. Pole Spacer Plate 6550-001-118 I.V. Clip Spacer Plate Return To Table of Contents 6550-309-001 REV A www.stryker.com...
3-Stage I.V. Pole Assembly, Right - 6500-001-043 Rev D Item Part No. Part Name Qty. 6070-090-105 Label, No Push/Pull 6070-215-070 3-Stage I.V. Pole Assembly 6100-115-051 Socket Weldment 6500-001-255 Label 6070-110-037 I.V. Pivot Pin 0025-079-000 Dome Head Rivet Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Note: 0004-589-000 shown in cot assembly Item Part No. Part Name Qty. 6100-115-060 I.V. Pole Clip 6500-001-044 3-Stage I.V. Pole Assembly, Left (page 125) 6550-001-117 I.V. Pole Spacer Plate 6550-001-118 I.V. Clip Spacer Plate Return To Table of Contents 6550-309-001 REV A www.stryker.com...
3-Stage I.V. Pole Assembly, Left - 6500-001-044 Rev D Item Part No. Part Name Qty. 6070-090-105 Label, No Push/Pull 6070-215-070 3-Stage I.V. Pole Assembly 6100-115-051 Socket Weldment 6500-001-256 Label 6070-110-037 I.V. Pivot 0025-079-000 Dome Head Rivet Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Note: 0004-589-000 shown in cot assembly Item Part No. Part Name Qty. 6100-115-060 I.V. Pole Clip 6500-001-043 3-Stage I.V. Pole Assembly, Right (page 123) 6550-001-044 3-Stage I.V. Pole Assembly, Left (page 125) Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Optional Push Bar - 6550-040-000 Rev A Item Part No. Part Name Qty. 0021-179-000 Set Screw 6550-001-026 Corner Handle Assy (page 128) 6550-001-108 Push Bar Sleeve 6550-001-199 Push Bar Storage Pouch Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Optional Corner Handle Assembly 6550-001-026 Rev B (Reference Only) Item Part No. Part Name Qty. 0026-387-000 Slotted Spring Pin 0038-589-000 Compression Spring 6510-001-119 Handle Ball 6550-001-067 Handle Weldment 6550-001-100 Push Bar Lock Button Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Label, SMRT Power 6500-030-000 Rev A (Reference Only) 6500-031-000 Rev A (Reference Only) 6500-032-000 Rev A (Reference Only) Item Part No. Part Name Qty. 6500-001-356 Label, SMRT Power Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Label, DeWALT® 6500-271-000 Rev A (Reference Only) 6500-275-000 Rev A (Reference Only) Item Part No. Part Name Qty. 6500-001-258 Label, DeWALT® Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Optional Accessories The accessories listed below can be purchased and installed on the Power-PRO TL cot. Assembly Page Accessory Part Number Number Equipment Hook 6500-147-000 Oxygen Bottle Holder 6550-002-020 Oxygen Bottle Holder, Removable 6080-140-000 Defibrillator Platform 6100-170-000 Pocketed Back Rest Pouch...
Equipment Hook - 6500-147-000 Rev B Item Part No. Part Name Qty. 0025-079-000 Dome Head Rivet 6500-001-237 Equipment Hook Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Rev A 6080-140-010 Rev A (Reference Only) Item Part No. Part Name Qty. 6080-140-011 Oxygen Bottle Holder Hanger 6080-140-012 Oxygen Bottle Holder Note: Max Load = 1.78 stones (11.4 kg) (25 lb) Return To Table of Contents 6550-309-001 REV A www.stryker.com...
The weight of the monitoring equipment must not exceed 5.4 stones (34 kg) (75 lb) on the defibrillator platform (if equipped). Monitoring equipment should not be stored on the cot or on the defibrillator platform when in the ambulance vehicle. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Backrest Pouch - 6500-130-000 Rev A Item Part No. Part Name Qty. 6500-001-241 Pocketed Backrest Storage Pouch Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Recycling Passport Assembly Part Number: 6500-101-010 Rev D (Reference Only) View of battery Item Recycling/Material Code Important Information SMRT Pak (6500-101-010) NiCd Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 138
Recycling Passport Assembly Part Number: 6500-700-006 Rev B (Reference Only) View of battery Item Recycling/Material Code Important Information Battery Pack NiCd Note: See the DeWALT Manual for battery recycling information. ® Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 139
Recycling Passport Assembly Part Number: 6550-101-036 (Reference Only) Item Recycling/Material Code Important Information Printed Circuit Board Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 140
Recycling Passport Assembly Part Number: 6550-001-022 Rev A (Reference Only) Item Recycling/Material Code Important Information Printed Circuit Board Contains Liquid Crystal Display Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 141
Recycling Passport Assembly Part Number: 6550-001-014 (Reference Only) Item Recycling/Material Code Important Information External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 142
Assembly Part Number: 6550-001-030 Rev J (Reference Only) Item Recycling/Material Code Important Information Motor Contains Automatic Transmission Fluid* External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable * Mobil Mercon® V Synthetic Blend or equivalent Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 143
Recycling Passport Assembly Part Number: 6550-001-172 Rev C (Reference Only) Item Recycling/Material Code Important Information External Electrical Cable Return To Table of Contents www.stryker.com 6550-309-001 REV A...
• SMRT Charger (6500-201-010), AC/DC Charger (6500-070-000), • DeWALT ® DC/DC Charger (6500-072-000). • DeWALT ® Note: U is the alternating current mains voltage prior to application of the test level. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 145
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Page 146
If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model 6550 ower-PRO™ TL cot. Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m. Return To Table of Contents 6550-309-001 REV A www.stryker.com...
Page 147
(6500-070-000): Class A DC/DC Charger DeWALT ® (6500-072-000): N/A Cot: N/A SMRT Charger (6500-201-010): Complies Voltage Fluctuations Flicker Emissions AC/DC Charger DeWALT ® IEC 61000-3-3 (6500-070-000): Complies DC/DC Charger DeWALT ® (6500-072-000): N/A Return To Table of Contents www.stryker.com 6550-309-001 REV A...
Stryker warrants to the original purchaser that the welds on the Power-PRO TL will be free from structural defects for the expected 7 year life of the product as long as the original purchaser owns the product. Original purchasers will also obtain a three (3) year limited parts warranty for the X-frame components of the Power-PRO cot and a three (3) year limited power train warranty covering the motor pump assembly and hydraulic cylinder assembly.
Claim will be limited in amount to the actual replacement cost. In the event that this information is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full.
Page 151
United States Stryker Medical 3800 E. Centre Ave., Portage, Michigan USA 49002 European Representative Stryker France ZAC Satolas Green Pusignan Av. De Satolas Green 69881 MEYZIEU Cedex France 2010/06 6550-309-001 REV A www.stryker.com...