Do you have a question about the Power-PRO XT 6500 and is the answer not in the manual?
Questions and answers
Iván Garcia
January 24, 2025
Como inyectarle aceite hidráulico a la camilla?
Mi camilla Stryker 6500, al parecer necesita cambio de aceite. Últimamente no sube ni baja presionando + o - . Lo hace muy esporádicamente. De ahí deduZco UE le falta aceite hidráulico.
1 comments:
Mr. Anderson
February 10, 2025
To inject hydraulic oil into the Stryker Power-PRO XT 6500 stretcher, follow these steps:
1. Raise the cot to the full up position. 2. Ensure the fill port is horizontal and aligned with the hole in the motor mount. 3. Remove the port plug using a 3/16" Allen wrench. 4. Fill the reservoir up to the bottom of the fill port using Mobil Mercon® V Synthetic Blend Oil (6500-001-293). 5. Replace the plug and run the cot up and down a few times to circulate the oil.
Page 1
Power-PRO™ XT Model 6500 Operations/Maintenance Manual For Parts or Technical Assistance: USA: 1-800-327-0770 (option 2) Canada: 1-888-233-6888 2008/11 6500-309-001 REV A www.stryker.com...
The Power-PRO™ XT is designed to be compatible with competitive cot fastener systems. WALT ® is a registered trademark of Black & Decker Inc. Patents pending. The yellow and black color scheme is a proprietary trademark of Stryker Corporation. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 7
Support at: (800) 327-0770 or (269) 324-6500. Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 Please have the serial number of your Stryker product when calling Stryker Customer Service or Technical Support. HEAD END Include the serial number in all written communication.
NOTe This provides special information to make maintenance easier or important instructions clearer. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Refer to your local distributor for return and/or collection systems available in your country. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Stryker, found to be defective. Expendable components, i.e. mattresses, restraints, I.V. poles, storage nets, storage pouches, O2 straps, and other soft goods, have a one (1) year limited warranty.
Claim will be limited in amount to the actual replacement cost. In the event that this information is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full.
Failure to install the in-fastener shut-off module may cause injury to the patient or operator and/or damage to the vehicle. • It is the responsibility of the cot operator to ensure the ambulance cot being used in the Stryker cot fastener system meets the installation specifications listed on page 23.
Page 13
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY. If the products described above are used to clean Stryker patient care equipment, measures must be taken to insure the cots are wiped with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the cots will leave a corrosive residue on the surface of the cots, possibly causing premature corrosion of critical components.
• A preventative maintenance program should be established for all Stryker EMS equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product. Close attention should be given to safety features including, but not limited to: •...
These are general instructions for installation of the Pedi-Mate™. Safe and proper use of the Pedi-Mate™ is solely at the discretion of the user. Stryker recommends all users be trained on the proper use of the Pedi-Mate™ before using it in an actual situation.
Page 16
Summary of Safety Precautions WARNING: Pinch Points figure 2 - Potential Pinch Points Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Hydraulic Unit Wheel Lock (Optional) Transport Wheels Cot Retaining Post Transport Wheels Short Safety Hook Long Safety Hook Stryker part number Stryker part number 6060-036-017 6060-036-018 J Safety Hook Stryker part number 6092-036-018 figure 3 - Cot Components Return To Table of Contents www.stryker.com...
With the cot empty, check the raise/lower function. With the cot loaded with a minimum of 45 kg, check the raise/lower function. With the cot loaded with a minimum of 45 kg, check the load/unload function. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 19
Verify pocketed backrest storage pouch (if equipped) installed properly. Head-end storage flat (if equipped) installed properly. Verify pillow (if equipped) included. Verify 36" restraint extender (if equipped) is included. Verify the Bariatric transfer flat (if equipped) is included. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
(Figure 4e). Replace the sensor housing cover using the screws removed in step 1. Following completion of the sensor height adjustment, verify the cot properly engages the safety hook. figure 4e - Securing Cables Return To Table of Contents www.stryker.com 6500-309-001 REV A...
WaRNING It is the responsibility of the cot operator to ensure that the cot being used in the Stryker Cot Fastener System meets the installation specifications listed on page 20. Injury may result if a non-compatible cot is used in the Stryker Fastener System.
Damage to the product or injury to the patient or operator may occur. In-fastener shut-off must be installed in all ambulances in which the cot will be used. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
* The length of the socket head cap screws depends on the thickness of the vehicle floor. Use screws long enough to go completely through the patient compartment floor, washer and nut by at least two full threads. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
To avoid injury, verify the safety bar has engaged the safety hook before removing the cot from the patient compartment. Note: Stryker EMS recommends that, prior to installation, the certified mechanic plan the placement of the safety hook in the rear of the vehicle. Procedure: Remove the cot from the fastener and unload it from the vehicle.
Page 27
Verify the cot legs lock into the load position before disengaging the safety bar from the safety hook. Failure to properly lock the cot into position can cause injury to the patient or operator and/or damage to the cot. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
The arrows indicate alternate attachment areas. When attaching the restraint straps to the cot, remember the attachment points should provide both strong anchorage and proper restraint position while not interfering with equipment and accessories. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 29
Inspection of the restraints should be done at least once a month (more frequently if used heavily). Inspection should include checking for a bent or broken receiver or tang, torn or frayed webbing, etc. Any restraint showing wear or not operating properly must be replaced immediately. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Cot features USING The ReSTRaINT beLT eXTeNSION (OPTIONaL) figure 20 - attaching the Restraint belt extension Use the restraint belt extension for extra length when buckling the lap belt around large patients. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
22 - fastening the Pedi-Mate™ buckle WaRNING To avoid accidental release of the Pedi-Mate™, and possible injury to the infant, ensure the buckle is located away from obstructions on the cot or accessories. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Page 32
These are general instructions for installation of the Pedi-Mate™. Safe and proper use of the Pedi-Mate™ is solely at the discretion of the user. Stryker recommends all users be trained on the proper use of the Pedi-Mate™ before using it in an actual situation.
CaUTION Wheel lock(s) are only intended to help prevent the cot from rolling while unattended. A wheel lock may not provide sufficient resistance on all surfaces or under loads. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
If the arrow on the bottom bracket of the retaining post points toward the head end of the cot, the retaining post is set for an X-frame style cot. If the arrow points toward the foot end of the cot the post is set for an H-frame style cot. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
To lower the footrest, lift the foot rest frame (A) and, while holding the frame, lift up on the release handle (B) until the bracket disengages. Lower the footrest until flat. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Siderails are not intended to serve as a patient restraint device. Refer to pages 24 and 25 for proper restraint strap usage. Failure to utilize the siderails properly could result in patient injury. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
An uncharged or depleted battery may cause poor ambulance cot performance. • er to the D WALT ® manual (Stryker part number 6500-001-206) for battery and charger information. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
CaUTION • A preventative maintenance program should be established for all Stryker EMS equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product. Close attention should be given to safety features including, but not limited to: •...
Do not store items under the ambulance cot mattress. Storing items under the mattress can interfere with the operation of the ambulance cot. The weight of the equipment in the head end storage flat (if equipped) must not exceed 40 pounds (18 kg). Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Do not store items under the ambulance cot mattress. Storing items under the mattress can interfere with the operation of the ambulance cot. • weight of the equipment in the pocketed backrest storage pouch (if equipped) must not exceed 20 pounds (9 kilograms). Return To Table of Contents 6500-309-001 REV A www.stryker.com...
The equipment hook is used to hang additional accessories or equipment such as defibrillators or monitors. CaUTION To avoid damage to the equipment hook, the weight of the accessories or equipment must not exceed 20 pounds (9 kilograms). Return To Table of Contents www.stryker.com 6500-309-001 REV A...
To avoid damage to the I.V. pole, the weight of the I.V. bags or equipment must not exceed 40 pounds (18kg). Turn the lock actuator (B) counterclockwise and slide section (C) into the bottom tube. Lift up and pivot the pole down into the storage position. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
(C) into the bottom tube. Lift up and pivot the pole down into the storage position. figure 37 - 3-Stage I.V. Pole Storage Position figure 38 - 3-Stage I.V. Pole Return To Table of Contents www.stryker.com 6500-309-001 REV A...
• Use the cot only as described in this manual. • Read all labels and instructions on the cot before using the cot. • Use a minimum of two (2) operators to manipulate the cot while a patient is on the cot. If additional assistance is needed, see the reference chart on page 55. • Do not adjust, roll or load the cot into a vehicle without advising the patient. Stay with the patient and control the cot at all times. • The ambulance cot can be transported in any position. Stryker recommends transporting the patient in as low a position as is comfortable for the operators to maneuver the cot. • Only use the wheel lock(s) during patient transfer or without a patient on the ambulance cot. • Always use the restraint straps. • Use properly trained helpers when necessary to control the cot and patient. WaRNING •...
USING The OPTIONaL TRaNSfeR fLaT When transferring large patients, use of the Transfer Flat (Stryker part number 6083-001-200) is recommended. aMbULaNCe COT MOTION Make sure all the restraint straps are securely buckled around the patient (see page 19 & 20 for restraint strap usage instructions).
41 - 2 Operators Loading the Cot both Operators - Push the cot into the patient compartment (Figure 41), until the cot engages the cot fastener (not included). Return To Table of Contents 6500-309-001 REV A www.stryker.com...
(–) button until the undercarriage of the cot retracts fully. Both operators should push the cot into the patient compartment, until the cot engages the cot fastener (not included). Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Disengage the cot from the cot fastener (For more detailed instructions, reference the ambulance cot fastener installation/operation manual - Stryker part number 6370-090-010). Lift the vehicle bumper to the raised position (if possible).
Disengage the cot from the cot fastener (For more detailed in- structions, reference the ambulance cot fastener installation/ operation manual - Stryker part number 6370-090-010). figure 48 - Pull the Cot out of the Vehicle Lift the vehicle bumper to the raised position (if possible).
53 - 2 Operators - One Lifting the base lift with a slow constant motion. figure 54 - Push the Cot into the Vehicle Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Disengage the cot from the cot fastener (For more detailed instructions, reference the ambulance cot fastener installation/operation manual - Stryker part figure 55 - 2 Operators Unloading the Cot number 6370-090-010). Lift the vehicle bumper to the raised position (if equipped).
Disengage the cot from the cot fastener (For more detailed instructions, reference the ambulance cot fastener figure 58 - Pull the Cot out of the Vehicle installation/operation manual - Stryker part number 6370-090-010). Lift the vehicle bumper to the raised position (if possible).
62 - head Section Retracted the patient or operator and/or damage to the product. figure 63 - head Section Release handles Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Avoid contact with a wet battery or battery enclosure. Contact may cause injury to the patient or operator. CaUTION Remove the battery if the cot is not going to be used for an extended period of time (over 24 hours). Return To Table of Contents www.stryker.com 6500-309-001 REV A...
• Remove the battery! The battery and charger are not immersible or power washable. • Follow the cleaning solution manufacturer’s dilution recommendations exactly. • The preferred method Stryker Medical recommends for power washing the Power-PRO™ XT ambulance cot is with the standard hospital surgical cart washer or hand held wand unit.
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY. If the products described above are used to clean Stryker patient care equipment, measures must be taken to insure the cots are wiped with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the cots will leave a corrosive residue on the surface of the cots, possibly causing premature corrosion of critical components.
Improper maintenance can cause injury or damage to the product. Maintain the ambulance cot as described in this manual. Use only Stryker approved parts and maintenance procedures. Using unapproved parts and procedures could cause unpredictable operation and/or injury and will void the product warranty.
The Power-PRO™ XT ambulance cot has been designed to operate without the need for lubrication. CaUTION • Do not lubricate the bearings in the X-frame as it will degrade the performance of the Power-PRO™ XT ambulance cot and may void its warranty. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Training Record Training Date Training Method Trainee Name basic Refresher Owner’s Manual, In-Service, Training Update formal Class, etc. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Cot assembly For Reference Only: 6500-001-010 Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Page 68
Cot assembly Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 69
Cot assembly Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Page 70
Cot assembly Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 71
Cot assembly Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Page 72
Cot assembly Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 73
0016-002-000 Nylock Hex Nut 0016-028-000 Nylock Hex Nut 0016-035-000 Nylock Hex Nut 0016-102-000 Nylock Hex Nut 0025-079-000 Dome Head Rivet 0028-181-000 Truarc Ring 0038-574-000 Spring 1010-031-077 Gas Spring 6082-026-010 Siderail Assembly Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Page 74
Item Part No. Part Name qty. 6060-090-002 Serial Number Tag 0025-133-000 Dome Head Rivet 6082-032-045 Fowler Skin 6082-090-043 Stryker Label 6060-090-004 Rugged Label 6100-003-125 Straight T Pivot 6500-001-012 Base Assembly 6500-001-015 Foot End Assembly (pg. 93) 6500-001-017 Slider Magnet 6500-001-018 Fowler Assembly (pg.
Inner Lift Tube, Litter Pivot - 6500-001-034 Patient Right assembly Item Part No. Part Name qty. 0004-634-000 Button Head Cap Screw 0055-100-075 6500-001-355 Inner Lift Tube Weldment 6500-001-125 Dead Stop Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Inner Lift Tube, Litter Pivot - 6500-001-035 Patient Left assembly Item Part No. Part Name qty. 0004-634-000 Button Head Cap Screw 0055-100-075 6500-301-053 Inner Lift Tube Weldment 6500-001-125 Dead Stop Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Outer Rail, Patient Right assembly For Reference Only: 6500-001-032 See Detail A Detail A Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 85
Dead Stop 6500-001-104 Support Bracket 6500-001-106 Support Bracket 6500-001-114 Outer Rail 6500-001-116 Siderail Bracket 6500-001-117 Siderail Clamp 6500-001-243 I.V. Pole Backer Plate 6500-001-244 I.V. Clip Backer Plate 6500-001-245 Sensor Housing Backer Plate Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Outer Rail, Patient Left assembly For Reference Only: 6500-001-033 See Detail A Detail A Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 87
Dead Stop 6500-001-104 Support Bracket 6500-001-106 Support Bracket 6500-001-115 Outer Rail 6500-001-116 Siderail Bracket 6500-001-117 Siderail Clamp 6500-001-243 I.V. Pole Backer Plate 6500-001-244 I.V. Clip Backer Plate 6500-001-245 Sensor Housing Backer Plate Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Sensor assembly For Reference Only: 6500-001-028 Item Part No. Part Name qty. 0004-596-000 Button Head Cap Screw 6500-001-124 Sensor Housing 6500-001-160 Hall Effects Sensor 6500-001-199 Housing Cover Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Sensor housing assembly For Reference Only: 6500-001-029 Item Part No. Part Name qty. 0004-596-000 Button Head Cap Screw 6500-001-124 Sensor Housing 6500-001-199 Housing Cover Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Powerplant assembly For Reference Only: 6500-001-031 Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 91
0015-052-000 0016-002-000 0016-102-000 0055-100-074 6500-001-030 Hydraulics Assembly (pg. 92) 6500-001-105 Cross Tube 6500-001-152 Bracket 6500-001-169 Cylinder Rod End 6500-001-194 Motor Mount 6500-001-212 Cross Bar 6500-001-249 Spacer 6500-001-250 Spacer 6500-001-294 Motor Mount Return To Table of Contents www.stryker.com 6500-309-001 REV A...
For Reference Only: 6500-001-015 Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Page 94
Route Hall Sensor Cable around Item (AP) and through this area Route Switch Cables around Item (AP) and through this area Route Manual Release Cable through this area Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 95
Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Switch assembly - 6500-001-016 Item Part No. Part Name qty. 0004-616-000 Socket Head Cap Screw 6500-001-130 Activation Switch 6500-001-143 Housing, Upper, Button, Foot End 1 Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Button Head Cap Screw 0015-050-000 0016-028-000 0021-119-000 Set Screw 0021-138-000 Set Screw 0025-131-000 Rivet 0028-076-000 External Retaining Ring 0946-035-025 Liner 6060-032-038 Yoke 6060-032-040 Pivot 6082-032-050 Fowler Weldment 6082-032-052 Release Handle Weldment Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Pivot, Safety Bar, Right 6500-001-324 Pivot, Safety Bar, Left 6500-001-280 Head Section Guard, Right 6500-001-281 Head Section Guard, Left 6500-001-325 Safety Bar Torsion Spring, Left 6500-001-326 Safety Bar Torsion Spring, Right Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Section Lock assembly - 6500-001-026 Item Part No. Part Name qty. 0038-570-000 Compression Spring 0038-572-000 Compression Spring 6500-001-089 Bearing Tip 6500-001-091 Latch Housing 6500-001-092 Latch Housing 6500-001-094 6500-001-095 Actuation Slide Return To Table of Contents 6500-309-001 REV A www.stryker.com...
In-fastener Shut-Off assembly - 6500-001-027 Item Part No. Part Name qty. 0004-135-000 Button Head Cap Screw 0015-016-000 Square Nut 0025-079-000 Rivet 6500-001-271 Magnet 6500-001-272 Holder Return To Table of Contents www.stryker.com 6500-309-001 REV A...
0004-508-000 Button Head Cap Screw 0011-001-000 Washer 0016-003-000 Nylock Nut 0025-079-000 Dome Head Rivet 6060-140-013 Strap Assembly 6500-001-040 Oxygen Bottle Holder Ass’y (pg. 105) NOTe: Max Load = 40 lbs. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Part No. Part Name qty. 0004-636-000 Button Head Cap Screw 6500-001-119 Bracket 6500-001-231 Label 6500-001-260 Bottle Holder Cover 6500-001-261 Bottle Holder Strap 6500-001-262 0025-079-000 Rivet NOTe: Max Load = 40 lbs. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
(pg. 109) Optional Three-Stage I.V. Pole assembly, Right - 6500-215-000 (for Reference Only) Item Part No. Part Name qty. 6100-115-060 Clip, I.V. Pole 6500-001-043 I.V. Pole Assembly, 3-stage, Right (pg. 110) Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Page 109
Two-Stage I.V. Pole assembly Patient Right - 6500-001-041 Item Part No. Part Name qty. 6070-090-105 Caution Label 6070-210-070 Pole Assembly 6100-115-051 Socket Weldment 6500-001-253 Label 0025-079-000 Dome Head Rivet 6070-110-037 I.V. Pivot Pin Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Three-Stage I.V. Pole assembly Patient Right - 6500-001-043 Item Part No. Part Name qty. 6070-090-105 Caution Label 6070-210-070 Pole Assembly 6100-115-051 Socket Weldment 6500-001-255 Label 6070-110-037 I.V. Pivot Pin 0025-079-000 Dome Head Rivet Return To Table of Contents 6500-309-001 REV A www.stryker.com...
(pg. 112) Optional 3-Stage I.V Pole assembly, Left - 6500-216-000 (for Reference Only) Item Part No. Part Name qty. 6100-115-060 Clip, I.V. Pole 6500-001-044 I.V. Pole Assembly, 3-stage, Left (pg. 113) Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Two-Stage I.V. Pole assembly Patient Left - 6500-001-042 Item Part No. Part Name qty. 6070-090-105 Caution Label 6070-210-070 Pole Assembly 6100-115-051 Socket Weldment 6500-001-254 Label 0025-079-000 Dome Head Rivet 6070-110-037 I.V. Pivot Pin Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Three-Stage I.V. Pole assembly Patient Left - 6500-001-044 Item Part No. Part Name qty. 6070-090-105 Caution Label 6070-210-070 Pole Assembly 6100-115-051 Socket Weldment 6500-001-256 Label 6070-110-037 I.V. Pivot Pin 0025-079-000 Dome Head Rivet Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Optional Three-Stage I.V. Pole assembly, Dual - 6500-217-000 (for Reference Only) Item Part No. Part Name qty. 6100-115-060 Clip, I.V. Pole 6500-001-043 I.V. Pole Assembly, 3-stage, Right (pg. ) 6500-001-044 I.V. Pole Assembly, 3-stage, Left (pg. 113) Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Head End Bracket Weldment, Right 2 6082-170-050 Defib U-Foot Weldment 6100-170-020 Defib Latch Strap Assembly 0004-234-000 Button Head Cap Screw 0011-355-000 Washer 0016-102-000 Nylock Nut 0026-172-000 Slotted Spring Pin 6500-001-298 Label Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Optional head extension - 6100-044-000 Item Part No. Part Name qty. 6100-041-030 Head Extension Pillow 6100-044-012 Head Extension Assembly (pg. 117) Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Rod-Side (Non-Locking) Manual Valve Cap-Side (Locking) Manual Valve Item Connection Connection Item DeTaIL b Orange A-Valve Blue Wiring Schematics B-Valve Main Cable 8-Pin Connector White Green Pressure Switch Brown Black Wireless Expansion Port Yellow Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Top Hall Effects Stop until button is pressed Jog Up Sensor Detected? Ramp in motor (Up) Run motor at max speed Top Hall Effects Ramp out motor and stop Sensor Detected? Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Page 122
Open B-Valve and Ramp in Motor Hold B-Valve open until button released or pressure switch closes Run motor at max speed Bottom Hall Effects Sensor Detected? Ramp out motor and stop, Close B-Valve Return To Table of Contents 6500-309-001 REV A www.stryker.com...
If voltage is present, replace the wire harness. C. If the GREEN light turns on, but does not lower, try the other switch. If the other switch works, replace the bad switch. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Check for 24 VDC on electronics assembly connection (H) (-) lead on black (+) lead on green while pressing the (+) button. If voltage is not present, replace the electronics assembly. If voltage is present, replace the main cable. 6500-309-001 REV A www.stryker.com...
Litter does not extend in the manual Check manual cable adjustment. mode. Change non-locking manual valve. Call Technical Support. High Speed retract does not engage. Check that weight is off the casters. Change pressure switch. Change hall effect cable. Call Technical Support. www.stryker.com 6500-309-001 REV A...
Page 126
Remove the port plug using a 3/16" Allen wrench. Fill the reservoir up to the bottom of the fill port. Replace the plug and run the cot up and down a few times. Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Recycling Passport PRODUCT: 6500-700-006 (REFERENCE ONLY) View of battery Item Recycling/Material Code Important Information Battery Pack NiCd NOTe Reference D WALT ® Manual (Part Number 6500-001-206) for battery recycling information. Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Recycling Passport PRODUCT: 6500-001-016 (REFERENCE ONLY) Exploded view of switch assembly Item Recycling/Material Code Important Information Printed Circuit Board Return To Table of Contents 6500-309-001 REV A www.stryker.com...
Recycling Passport PRODUCT: 6500-001-028 (REFERENCE ONLY) Exploded view of hall sensor assembly Item Recycling/Material Code Important Information Printed Circuit Board Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Exploded view of hydraulics assembly Item Recycling/Material Code Important Information Motor Contains Automatic Transmission Fluid* External Electrical Cable External Electrical Cable External Electrical Cable External Electrical Cable * Mobil Mercon V Synthetic Blend or equivalent Return To Table of Contents www.stryker.com 6500-309-001 REV A...
Recycling Passport PRODUCT: 6500-001-159 (REFERENCE ONLY) Main Cable Assembly Item Recycling/Material Code Important Information External Electrical Cable Return To Table of Contents 6500-309-001 REV A www.stryker.com...
The parts and accessories listed on this page are all currently available for purchase. Some of the parts identified on the assembly drawing parts in this manual may not be individually available for purchase. Please call Stryker Customer Service USA: 1-800-327-0770 (Option 2) for availability and pricing.
Page 136
UNITED STATES Stryker Medical 3800 E. Centre Ave., Portage, Michigan USA 49002 CANADA Stryker Medical Stryker Canada 3800 E. Centre Ave., 45 Innovation Drive Portage, Michigan Hamilton, Ontario Canada 49002 L9H 7L8 european Representative european Representative Stryker France Stryker France...
Need help?
Do you have a question about the Power-PRO XT 6500 and is the answer not in the manual?
Questions and answers
Como inyectarle aceite hidráulico a la camilla? Mi camilla Stryker 6500, al parecer necesita cambio de aceite. Últimamente no sube ni baja presionando + o - . Lo hace muy esporádicamente. De ahí deduZco UE le falta aceite hidráulico.
To inject hydraulic oil into the Stryker Power-PRO XT 6500 stretcher, follow these steps:
1. Raise the cot to the full up position.
2. Ensure the fill port is horizontal and aligned with the hole in the motor mount.
3. Remove the port plug using a 3/16" Allen wrench.
4. Fill the reservoir up to the bottom of the fill port using Mobil Mercon® V Synthetic Blend Oil (6500-001-293).
5. Replace the plug and run the cot up and down a few times to circulate the oil.
This ensures proper hydraulic function.
This answer is automatically generated