Smart 13
Micro Centrifuge
Date of purchase
Serial No.

Advertisement

Table of Contents
loading

Summary of Contents for Hanil Smart 13

  • Page 1 Operation Manual Smart 13 Micro Centrifuge Date of purchase Serial No.
  • Page 2 Research Use Only Copyrightⓒ 2018 Hanil Scientific Inc. All rights reserved. Contact Us If you have any questions, contactHanil Scientific Inc. or place of purchase. techsupport@ihanil.com / info@ihanil.com Doc.No. : OMM4PDEN1801 2/20 www.ihanil.com...
  • Page 3 Index 1. General Considerations -------------------------------------------------------- 4 1.1 Safety - 4 1.2 Transport & Storage - 5 1.3 Safety label - 5 1.4 Electric safety information - 5 2. Product Description ------------------------------------------------------------- 6 2.1 Appearance - 6 2.2 Delivery Package - 6 2.3 Technical Specifications - 7 3.
  • Page 4: General Considerations

    2. Check the voltage to be used, before connecting the centrifuge to the power source. 3.  Only use rotors, parts, and accessories provided by Hanil Scientific Inc. Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by using parts and accessories not recommended.
  • Page 5: Transport And Storage

    5. Do not block vents. 6. If you have the following emergencies, shut off the power supply and unplug the power cord from outlet and contact your place of purchase or Hanil Scientific Inc. - Unusual noises or smell from the device.
  • Page 6: Product Description

    ⑨ Absorbent rubber : Rubber fix the device on the laboratory bench. 2.2 Delivery Package ① Standard offering ② Optional products [ See. 7. Rotors & Acc.] - Smart 13 main body(incl. :A2.0-12) - PCR tube & strip rotor - Powercable(300/500V,6A) - 0.5 ml tube adaptor - Usermanual - 0.2 ml tube adaptor...
  • Page 7 2. Product Description Technical Specifications Max. RPM 13,500 RPM Max. RCF 12,032 xg Time ≤ 15 min, Pulse Acc. time to max. speed ≤15 sec Dec. time from max. speed ≤8 sec Tube Capacity 1.5/2.0ml x 12 Power supply AC110/220V, 50/60Hz Dimension (W x D x H, mm) 189.7 x 202 x 129 Weight (kg)
  • Page 8: Installation

    3.1 Packing Inspection · Check packing conditions very carefully before unpacking. · Contact Hanil Scientific Inc. immediately if damages found. · Check the delivery for completeness. · You can get contact details on packing box or back of the manual.
  • Page 9: Connect Power

    3. Installation 3.3 Connect Power · Connect the device to voltage sources which correspond to the electrical requirements on the label attached to the device. · Use sockets with a protective earth conductor and provided power cord. 1. Connect the power cable to the power socket on the back side of the device and plug the cord into the power outlet.
  • Page 10 3.5 Loading and Removal of the Rotor · This device is equipped withthe rotor . · Only use rotors, parts, and accessories provided by Hanil Scientific Inc . · Do not use a defective rotor. Follow the steps below to remove or replace the rotor.
  • Page 11 3. Installation 3.6 Loading tube · Only use tubes provided or approved by Hanil Scientific Inc . · Always use the same type of tube . · Tubes should be loaded symmetrically . · Do not exceed the maximum rated speed of the tube .
  • Page 12: Operation

    4. Operation 4.1 Control Panel Button Description ① RPM setting knob Rotate clockwise or counterclockwise to set the rpm value. Rotate clockwise or counterclockwise to set the run time. ② Time setting knob (To use pulse function, set the time to '0' and press the start key.) ③...
  • Page 13: Setting Rpm

    4. Operation 4.2 Setting RPM [A2.0-12 rotor] 1. Rotate the 'RPM setting knob' clockwise or counterclockwise to set the rpm value. ▶ 0~13,500 RPM [A0.2-32 rotor/A5.0-6 rotor] 1. Press the 6K RUN button to start operation at 6,000 rpm ▶ Do not exceed 6,000 rpm. 4.3 Setting Run Time 1.
  • Page 14: Maintenance

    2. Rotate the shaft with your hand to make sure it turns smoothly 3. Use the stopwatch to check that the time setting is correct 4. If you find any damages, do not use the device. Contact Hanil Scientific Inc. 5. Be careful not to get any foreign substances on the sensor.
  • Page 15: Troubleshooting

    6. Troubleshooting 6.1 General Errors Problem Recommended Action Power failure Check the power cord connection. When opening the lid, error sound activated. Error sound doesn't stop Press the 6K Run button to stop the error sound. Lid cannot be opened Press the ‘Lid open button’. Remove the dirt on the lid latch and close the lid again.
  • Page 16: Rotors And Accessories

    7. Rotors and Accessories Tube Tube capacity 0.2 ml 0.5 ml 1.5/2.0 ml Adaptor None Fixed Angle Rotor, A2.0-12 Cat. No. TR0.2 TR0.5 Adaptor bore 12 x 1 . 5 / 2 . 0 ml 6.3 x 21 8 x 37 (Фx L, mm) Radius (mm) 44.04...
  • Page 17 MEMO...