Single Measurement

Make sure that single measurement is selected.
(For the method of setting the number of measurements to be taken, re-
fer to "Setting the Number of Average Measurements" on page 21.)
4
While "n-1" is displayed,
make sure that the READY
lamp is lit.
5
Place the measuring probe
against the point to be mea-
sured, and push it gently
until a click is heard.
◆ The instrument's xenon lamp will flash
momentarily, and the measured value
will appear on the display.
◆ If the instrument is not operated for
more than 60 seconds, the back-light
on the display will go out.
6
Reading the value.
◆If the measured value blinks, the
value is outside the guaranteed mea-
surement accuracy range. If "-0-" is
displayed, the measured value is out-
side the display range.
(For details of the guaranteed mea-