2) Inspecting

● The instrument can be easily checked to ascertain that it is perform-
ing measurement correctly, by using the checker supplied with the
Charger unit.
● It is recommended that the instrument be checked once or more each
day it is used.
● Since the configuration of this instrument differs from that of old mod-
els, never use conventional checkers.
● Do not touch the checker's surface with fingers. If the checker gets
dirty, wipe it with a soft cloth dampened with water, then wipe it with a
dry cloth.
1
Set the POWER switch to
ON while holding down the
RESET button.
2
Release the RESET button
immediately
is displayed on the display
window.
◆ If the RESET button is held down for
more than 15 seconds, the unit of
measurement will be switched. If this
occurs, switch the unit back to the pre-
vious one. (For the method of switch-
ing the unit of measurement, refer to
"Selecting the Unit of Measurement"
on page 19.)
after
"CHE"