Check that the loops of the coil are aligned with respect to each other
(running in parallel) and that the turn-to-turn separation (the pitch)
of neighboring loops lies within the specified tolerance for your
instrument type:

Radial Induction Coil

Figure 8. Induction coil—radial where 1. Torch, 2. Induction coil and 3. Torch
clamp
Pitch of Radial Induction Coil
Figure 9. Pitch measurement—a radial induction coil must have a pitch (↕) of
4.7 ±0.5 mm
Pitch is measured from the center of one coil to the next, as shown above by the
NOTE
arrows between the dotted gray lines, and not the gap between adjacent coils.
Agilent 700 Series ICP Optical Emission Spectrometers User's Guide
Radial induction coil (Agilent 715/725 ICP-OES radial
instruments only)
Axial induction coil (Agilent 710/720 ICP-OES axial instruments
only)
Maintenance and Troubleshooting