Preparation; Unpacking The Product; Inspecting The Device; Assembling The Fabric Tube For The Compressed Air Feed - Storz CALCUSPLIT Instructions For Use Manual

Handpiece with lithotripsy solid probes
Table of Contents

Advertisement

5 Preparation

5.1 Unpacking the product

1.
Carefully remove the product and accessories from the packaging.
2.
Check the delivery for missing items and evidence of shipping damage.
3.
In the case of damage, hidden defects, and short deliveries, document their nature and
extent and contact the manufacturer or supplier immediately.

5.2 Inspecting the device

1.
Inspect the product for visible contamination. Do not use if contaminated.
2.
Inspect the product for the following characteristics:
Good working order
Functionality
Correct assembly of the components
Completeness

5.3 Assembling the fabric tube for the compressed air feed

1.
Insert the fabric tube into the connection nozzle for the compressed air supply.
2.
Insert the other end of the fabric tube into the connector on the rear of the CALCUSPLIT.
3.
Ensure that the compressed air feed is min. 3.5 bar and max. 5 bar.
4.
Ensure that dry and oil-free compressed air filtered to ≤ 5 μm is used.

5.4 Assembling the damping unit

Push the damping unit onto the lithotripsy solid probe from a distal direction until it cannot
go any further.
The damping unit must be replaced after a maximum of five applications.
Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes •