Disassembling the product ........................ 16 8 Maintenance, servicing, repairs, and disposal.................... 17 Repairing the product .......................... 17 Disposing of the product ........................ 17 9 Accessories and spare parts ........................ 18 Accessories............................. 18 Spare parts ............................. 18 10 Subsidiaries.............................. 19 Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
In combination with the following items: Product name Article number Lithotripsy solid probe 27632534 27632535 27632536 27632634 27632635 27632636 27632232 27632234 27634031 27634032 27634033 27634034 27630136 27632231 27632233 Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries. NOTICE ATTENTION Designates a possibly harmful situation. If this is not avoided, the products could be damaged. Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
2 Normal use 2.1 Intended purpose CALCUSPLIT probe CALCUSPLIT probes are intended to guide pneumatic pulse energy for lithotripsy to the calculus. Probes are surgically invasive and meant for short-term use. Lithotripsy handpiece Handpieces are intended to convert as well as transmit energy from the main system to the probe in endoscopic lithotripsy applications.
Products that come into contact with the central nervous system can become contaminated by organic residue containing prions. Prions lead to infection with Creutzfeldt-Jakob disease. If Creutzfeld-Jakob disease has been diagnosed or is suspected: Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
Page 8
Safety Dispose of the product properly and do not continue to use it. Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
27000L, 27001L, 27632635 27002L, 27003L, in conjunc- tion with instrument inlet 27001G and handpiece 27630038 27632636 KARL STORZ uretero-reno- scope 27002L, in conjunction with instrument inlet 27001G and handpiece 27630038 Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
27001G and handpiece 27630038 4.3 Symbols on the packaging Symbol Meaning Manufacturer Date of manufacture Medical device Article no. Batch code Number of products in the product packaging Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
The device must not be modified in any way 4.4 Ambient conditions There are no special transport and storage conditions for this product. Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
Insert the fabric tube into the connection nozzle for the compressed air supply. Insert the other end of the fabric tube into the connector on the rear of the CALCUSPLIT. Ensure that the compressed air feed is min. 3.5 bar and max. 5 bar.
Push the silicone tube onto the connector of the handpiece as far as it will go. Connect the silicone tube with the compressed air inlet. The lithotripsy solid probe must be replaced after a maximum of five applications. Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
Using the product outside the field of vision can cause injury to tissue and damage to accessories. Only activate the product under visual contact. Consult the CALCUSPLIT instructions for use. Only use medical compressed air. Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
Hold the fabric tube while the fabric hose is removed from the connection nozzle for the compressed air feed. The steps for assembly are carried out in the reverse order. Instructions for use • CALCUSPLIT handpiece with lithotripsy solid probes • PQW255_EN_V1.0_03-2021_IFU_CE-MDR...
8 Maintenance, servicing, repairs, and disposal 8.1 Repairing the product Repair work may only be performed by KARL STORZ or by a company authorized by KARL STORZ. The interventions described in this instruction manual are exempt from this rule. Please contact your local KARL STORZ subsidiary or authorized dealer (see the list of subsidiaries).