Calibration

The accuracy of the shock energy measurement needs to be checked on a regular basis (see
Maintenance and Technical Inspection on page 54) and after any replacements of PCBs or update of
the software.
C
AUTION
No more than 16 shocks can be applied without a pause of about 15 minutes afterwards to avoid
overheating of the device!
Check the current calibration first:
Apply two shocks with each 50 J, 100 J and 360 J to a calibrated defibtester (e.g. DNI Nevada Inc.). If
any of the delivered energy values deviates for more than ± 5% from the chosen value calibration is
required.
The calibration procedure requires
Dash Responder Service kit (2006861-001)
PC/Laptop
calibrated defibtester with energy display (e.g. DNI Nevada Inc.)
hard paddles or the adapter cable for the adhesive paddles which are suitable for the used
defibtester.
N
OTE
The charge level of the battery should be sufficient for several shocks.
The following steps describe how to calibrate the Dash Responder
Page 20 of 72
Servicing Manual Dash Responder V1.0
Revision F
2004768-001